Class 2 Device Recall Medline

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Medlinesee related information
Date Initiated by FirmAugust 31, 2026
Date PostedOctober 15, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0181-2027
Recall Event ID 99888
Product Classification General surgery tray - Product Code LRO
ProductConvenience Kits containing CONMED AirSeal Access Ports HERNANDEZ DAVINCI ACC BUNDLE M DYNJ911258 KIT ROBOTICS UROLOGY PROSTATE DYKMBNDL116I ROBOTIC CHEST DYNJ910997 DYNJ910997A RR-DAVINCI PROSTATE ACCESSORY DYNJ63392D UROLOGY LAPAROSCOPIC - 51638 DYNJ908720B UROLOGY ROBOT ADD ON PACK DYNJ85189
Code Information DYNJ911258 UDI-DI 10198459306723 Lots 25DMC798 25EMD961 25FMB704 25FMB704Z 25JMG936 25JMI469 DYKMBNDL116I UDI-DI 10198459644764 Lots 26ABL226 26BBD710 26EBC888 26EBH673 DYNJ910997 UDI-DI 10198459241987 Lots 25DMB772 25GMA914 25GMB737 25GMC286 25GMJ509 25IMB332 25IME774 25KMB190 25KMF850 25LMB486 DYNJ910997A UDI-DI 10198459582103 Lots 25LLA437 26BLA786 26CLB107 26DLA500 26DLB085 26FLA287 26GLA058 26GLA968 26GLB068 DYNJ63392D UDI-DI 10198459164507 Lots 25ALA298 25BLA967 DYNJ908720B UDI-DI 10198459642517 Lots 26EBE658 DYNJ85189 UDI-DI 10195327473754 Lots 24AMA883
FEI Number 1417592
Recalling Firm/
Manufacturer
Medline Industries, LP
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information ContactHaley Barclay
800-633-5463
Manufacturer Reason
for Recall
Kits contain CONMED AirSeal Access Ports. The ports were recalled by manufacturer due to potential for erroneous overpressure alerts and system venting during use, which may result in delays or ending procedures.
FDA Determined
Cause 2
Process change control
ActionThe firm began notifying customers on August 31, 2026. Customers were provided with CONMED's original recall notification, which instructs users to have a replacement AirSeal disposable be available prior to the procedure in order to minimize potential delays. Users should ensure tubing is properly connected between the tube set and access ports prior to initiating insufflation. If the connection cannot be securely tightened and the "overpressure" alert cannot be resolved, users should replace the affected device. Users were also provided with instructions on how to connect the AirSeal tubeset and access ports, which include new figures as an update to the prior Instructions for Use. The corrected Instructions for Use are available electronically at eifu.conmed.com/eifu.
Quantity in Commerce3,551 kits
DistributionUS Nationwide distribution in the states of AZ, CA, IN, MD, MO, NJ, OH, and WI.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-