| Date Initiated by Firm | August 31, 2026 |
| Date Posted | October 15, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0181-2027 |
| Recall Event ID |
99888 |
| Product Classification |
General surgery tray - Product Code LRO
|
| Product | Convenience Kits containing CONMED AirSeal Access Ports
HERNANDEZ DAVINCI ACC BUNDLE M
DYNJ911258
KIT ROBOTICS UROLOGY PROSTATE
DYKMBNDL116I
ROBOTIC CHEST
DYNJ910997
DYNJ910997A
RR-DAVINCI PROSTATE ACCESSORY
DYNJ63392D
UROLOGY LAPAROSCOPIC - 51638
DYNJ908720B
UROLOGY ROBOT ADD ON PACK
DYNJ85189 |
| Code Information |
DYNJ911258
UDI-DI 10198459306723
Lots 25DMC798
25EMD961
25FMB704
25FMB704Z
25JMG936
25JMI469
DYKMBNDL116I
UDI-DI 10198459644764
Lots 26ABL226
26BBD710
26EBC888
26EBH673
DYNJ910997
UDI-DI 10198459241987
Lots 25DMB772
25GMA914
25GMB737
25GMC286
25GMJ509
25IMB332
25IME774
25KMB190
25KMF850
25LMB486
DYNJ910997A
UDI-DI 10198459582103
Lots 25LLA437
26BLA786
26CLB107
26DLA500
26DLB085
26FLA287
26GLA058
26GLA968
26GLB068
DYNJ63392D
UDI-DI 10198459164507
Lots 25ALA298
25BLA967
DYNJ908720B
UDI-DI 10198459642517
Lots 26EBE658
DYNJ85189
UDI-DI 10195327473754
Lots 24AMA883
|
| FEI Number |
1417592
|
Recalling Firm/ Manufacturer |
Medline Industries, LP
3 Lakes Dr
Northfield
IL
60093-2753
|
| For Additional Information Contact | Haley Barclay 800-633-5463 |
Manufacturer Reason for Recall | Kits contain CONMED AirSeal Access Ports. The ports were recalled by manufacturer due to potential for erroneous overpressure alerts and system venting during use, which may result in delays or ending procedures. |
FDA Determined Cause 2 | Process change control |
| Action | The firm began notifying customers on August 31, 2026.
Customers were provided with CONMED's original recall notification, which instructs users to have a replacement AirSeal disposable be available prior to the procedure in order to minimize potential delays.
Users should ensure tubing is properly connected between the tube set and access ports prior to initiating insufflation. If the connection cannot be securely tightened and the "overpressure" alert cannot be resolved, users should replace the affected device.
Users were also provided with instructions on how to connect the AirSeal tubeset and access ports, which include new figures as an update to the prior Instructions for Use.
The corrected Instructions for Use are available electronically at eifu.conmed.com/eifu. |
| Quantity in Commerce | 3,551 kits |
| Distribution | US Nationwide distribution in the states of AZ, CA, IN, MD, MO, NJ, OH, and WI. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|