| | Class 2 Device Recall Medline |  |
| Date Initiated by Firm | August 31, 2026 |
| Date Posted | October 15, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0182-2027 |
| Recall Event ID |
99888 |
| Product Classification |
General surgery tray - Product Code LRO
|
| Product | Convenience Kits containing CONMED AirSeal Access Ports
GENERAL DAVINCI
DYNJG901004C
LAP CHOLE
DYNJ88071
ROBOTIC PACK
DYNJ50474F |
| Code Information |
DYNJG901004C
UDI-DI 10198459535772
Lots 25LLA307
26BLA900
DYNJ88071
UDI-DI 10198459098512
Lots 24KMD416
24LMG268
25AMH987
25BMA076
25GMK411
DYNJ50474F
UDI-DI 10195327559007
Lots 24DDA107
24DDA469
24EDA512
24GDB884
24IDA882
|
| FEI Number |
1417592
|
Recalling Firm/ Manufacturer |
Medline Industries, LP
3 Lakes Dr
Northfield
IL
60093-2753
|
| For Additional Information Contact | Haley Barclay 800-633-5463 |
Manufacturer Reason for Recall | Kits contain CONMED AirSeal Access Ports. The ports were recalled by manufacturer due to potential for erroneous overpressure alerts and system venting during use, which may result in delays or ending procedures. |
FDA Determined Cause 2 | Process change control |
| Action | The firm began notifying customers on August 31, 2026.
Customers were provided with CONMED's original recall notification, which instructs users to have a replacement AirSeal disposable be available prior to the procedure in order to minimize potential delays.
Users should ensure tubing is properly connected between the tube set and access ports prior to initiating insufflation. If the connection cannot be securely tightened and the "overpressure" alert cannot be resolved, users should replace the affected device.
Users were also provided with instructions on how to connect the AirSeal tubeset and access ports, which include new figures as an update to the prior Instructions for Use.
The corrected Instructions for Use are available electronically at eifu.conmed.com/eifu. |
| Quantity in Commerce | 3,551 kits |
| Distribution | US Nationwide distribution in the states of AZ, CA, IN, MD, MO, NJ, OH, and WI. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|