 Class 2 Device Recall Medlinesee related information
Date Initiated by FirmAugust 31, 2026
Date PostedOctober 15, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0182-2027
Recall Event ID 99888
Product Classification General surgery tray - Product Code LRO
ProductConvenience Kits containing CONMED AirSeal Access Ports GENERAL DAVINCI DYNJG901004C LAP CHOLE DYNJ88071 ROBOTIC PACK DYNJ50474F
Code Information DYNJG901004C UDI-DI 10198459535772 Lots 25LLA307 26BLA900 DYNJ88071 UDI-DI 10198459098512 Lots 24KMD416 24LMG268 25AMH987 25BMA076 25GMK411 DYNJ50474F UDI-DI 10195327559007 Lots 24DDA107 24DDA469 24EDA512 24GDB884 24IDA882
FEI Number 1417592
Recalling Firm/
Manufacturer
Medline Industries, LP
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information ContactHaley Barclay
800-633-5463
Manufacturer Reason
for Recall
Kits contain CONMED AirSeal Access Ports. The ports were recalled by manufacturer due to potential for erroneous overpressure alerts and system venting during use, which may result in delays or ending procedures.
FDA Determined
Cause 2
Process change control
ActionThe firm began notifying customers on August 31, 2026. Customers were provided with CONMED's original recall notification, which instructs users to have a replacement AirSeal disposable be available prior to the procedure in order to minimize potential delays. Users should ensure tubing is properly connected between the tube set and access ports prior to initiating insufflation. If the connection cannot be securely tightened and the "overpressure" alert cannot be resolved, users should replace the affected device. Users were also provided with instructions on how to connect the AirSeal tubeset and access ports, which include new figures as an update to the prior Instructions for Use. The corrected Instructions for Use are available electronically at eifu.conmed.com/eifu.
Quantity in Commerce3,551 kits
DistributionUS Nationwide distribution in the states of AZ, CA, IN, MD, MO, NJ, OH, and WI.
Total Product Life CycleTPLC Device Report