| | Class 2 Device Recall ALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM |  |
| Date Initiated by Firm | October 12, 2026 |
| Date Posted | October 27, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0290-2027 |
| Recall Event ID |
99889 |
| 510(K)Number | K182415 |
| Product Classification |
Interventional fluoroscopic x-ray system - Product Code OWB
|
| Product | ALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM
REF: INFX-8000F |
| Code Information |
DI-04987670100208/ |
| FEI Number |
2020563
|
Recalling Firm/ Manufacturer |
Canon Medical System, USA, INC.
2441 Michelle Dr
Tustin
CA
92780-7047
|
| For Additional Information Contact | Canon InTouch 800-521-1968 |
Manufacturer Reason for Recall | Due to software issue, X-ray interventional system may experience internal error if F-Store or imagine acquisition is performed while clinical analysis function is use. |
FDA Determined Cause 2 | Software design |
| Action | On October 12, 2026, Canon Medical Systems USA, Inc. issued a Urgent: Medical Device Correction to affected consignees. Canon provided the following to consignees:
1. A representative of Canon Medical Systems USA will contact you to schedule a time to conduct installation of modified software to prevent internal errors from occuring if F-STORE or image acquisition is performed while CAAS clinical analysis function is used.
2. If you have any questions regarding this matter, please contact your service representative or contact our InTouch team at (800) 521-1968 or by email intouchdispatch@us.medical.canon.
3. In order to comply with the FDA record keeping requirement, Canon respectfully requests that the attached form be completed and returned via email to fieldcorrections@us.medical.canon |
| Quantity in Commerce | 3 units |
| Distribution | US Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, HI, IL, LA, MA, MI, MN, MS, NC, NV, OH, OR, TX. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = OWB
|
|
|
|