| | Class 2 Device Recall Direxion HIFLO |  |
| Date Initiated by Firm | September 16, 2026 |
| Date Posted | October 22, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0243-2027 |
| Recall Event ID |
99906 |
| 510(K)Number | K163701 |
| Product Classification |
Catheter, continuous flush - Product Code KRA
|
| Product | Direxion Torqueable Microcatheter, NGMC/SINGLE/027/BERN/1RO/130;
Material Number: M001195440 |
| Code Information |
Material Number: M001195440;
UDI-DI: 08714729839514;
Batch Numbers: 38181961, 38198832, 38224582, 38259483, 38324070, 38460469, 38560455, 38560456, 38595654, 38649857, 38745194, 38745195, 38810586, 38810591, 38904805, 39092267, 39092272, 39143222, 40470259, 40478976, 40478977, 40492318, 40497671, 40497858, 40502038, 40502039; |
| FEI Number |
3002913962
|
Recalling Firm/ Manufacturer |
Boston Scientific Cork, Ltd.
Boston Scientific Limited, Business And Technology Park, Model Farm Road
Cork
Ireland
|
| For Additional Information Contact | Clara Johnson 763-494-1133 |
Manufacturer Reason for Recall | The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. |
FDA Determined Cause 2 | Process control |
| Action | On September 16, 2026, Urgent Medical Device Removal - Immediate Action Required letters were sent to customers.
Actions to be taken:
1. Immediately stop further use or distribution and segregate affected product (see Attachment 1 for identification of the affected devices).
2. Forward this notice to relevant personnel and any facilities to which the product was transferred.
3. Continue to follow the standard of practice when monitoring patients treated with the product.
4. Complete and return the enclosed Reply Verification Tracking Form per the included instructions.
5. Return product to Boston Scientific in accordance with the enclosed instructions.
|
| Quantity in Commerce | 620 |
| Distribution | Worldwide distribution US Nationwide and territory of Puerto Rico. The countries of Albania, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Lithuania, Mexico, Moldova, Netherlands, New Zealand, Norway, Panama, PERU, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, T¿rkiye, UAE, United Kingdom. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = KRA
|
|
|
|