 Class 2 Device Recall Xactsee related information
Date Initiated by FirmSeptember 22, 2026
Date PostedOctober 22, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0268-2027
Recall Event ID 99956
Product Classification System, x-ray, film marking, radiographic - Product Code JAC
ProductXact mole marker for mammography Model/Catalog Number: M15RM
Code Information part #: M15RM; UDI-DI: 00854623006068; UPC-A; lot #: M15RM26A;
FEI Number 1221946
Recalling Firm/
Manufacturer
Solstice Corporation
85 E St
South Portland ME 04106-2870
For Additional Information ContactDanforth DeSena
1-800-8747922
Manufacturer Reason
for Recall
The radiopaque material may detach from the marker. In some cases, artifact was noted and attributed to the detached material.
FDA Determined
Cause 2
Process control
ActionSeptember 22, 2026, URGENT: Medical Device Removal letters were sent to customers. Action to be Taken: Please discontinue use of lot M15RM26A immediately and place any remaining inventory in quarantine. Product should be returned to the address below for free replacement or refund using Solstice UPS Account #A5255W. If you do not have UPS access, please email info@solsticecorp.com and request a prepaid label. Solstice Corporation Solstice Corporation has identified the cause of this defect and corrective action has been taken to ensure it will not recur. Any adverse events or other problems should be reported to www.FDA.gov/MedWatch. This product is now backordered with availability expected around October 15, 2026.
Quantity in Commerce736
DistributionWorldwide - US Nationwide distribution in the states and territory of CA, FL, IL, IN, MN, MO, NY, OH, PA, TX, WV and Puerto Rico. The country of Canada.
Total Product Life CycleTPLC Device Report