• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Illumina DRAGEN Server

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Illumina DRAGEN Serversee related information
Date Initiated by FirmMay 08, 2024
Date PostedOctober 30, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0323-2027
Recall Event ID 99957
Product Classification High throughput DNA sequence analyzer - Product Code PFF
ProductIllumina DRAGEN Server for NextSeq 550Dx REF: 20086130,
Code Information All Lots/UDI: N/A
FEI Number 3003218906
Recalling Firm/
Manufacturer
Illumina, Inc.
5200 Illumina Way
San Diego CA 92122-4616
For Additional Information ContactChristine Douglass
858-202-4500
Manufacturer Reason
for Recall
(Retrospectively reported) Cybersecurity vulnerability identified in communication protocol between the customer-provided infrastructure (network storage, servers, or computers) and in vitro diagnostic devices which may result in compromised data transmission.
FDA Determined
Cause 2
Labeling design
ActionOn May 8, 2024, Illumina, Inc. issued a Urgent Medical Device Recall via email. Illumina ask consignees to take the following 1. Download & read this Cybersecurity Guidance document. The full hyperlink is below. https://support.illumina.com/support-content/cyberguidance.html 2. For Each Product Connected to the Customer Network: work with your internal IT department to determine which secure protocol works best in your environment and to implement the appropriate configuration change. Note: a change to the network protocol affects both Dx and RUO modes simultaneously therefore there is no need to boot into each mode separately to take the necessary action. 3. Complete and return the Verification Form after carrying out all the steps in the instructions provided on your specific product(s) identified as affected in Table 1.
Quantity in Commerce32 units
DistributionWorldwide - US Nationwide distribution including in the states and territory of AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and the countries of AE, AR, AT, AU, BE, BG, CA, CH, CL, CN, CY, CZ, DE, DK, DZ, EG, ES, FI, FR, GB, Ghana, GR, GT, HK, HU, IE, IL, IS, IT, JP, KR, KW, LT, LU, LV, MA, MT, NG, NL, NO, NZ, PE, PH, PL, PT, QA, RO, RU, SA, SE, SG, SI, TH, TR, UA, VN, ZA.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-