Class 2 Device Recall Roche cobas C 303 analytical unit

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Roche cobas C 303 analytical unitsee related information
Date Initiated by FirmSeptember 23, 2026
Date PostedNovember 18, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0481-2027
Recall Event ID 99980
510(K)NumberK220134 
Product Classification Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
ProductRoche cobas c 303 analytical unit, Material Number 09793194001, clinical chemistry analyzer
Code Information UDI/DI 07613336220153, All units up to and including SN (Serial Number) 7696-10
FEI Number 1823260
Recalling Firm/
Manufacturer
Roche Diagnostics Operations, Inc.
9115 Hague Rd
Indianapolis IN 46256-1025
For Additional Information ContactRoche Support Network Customer Support Center
800-428-2336
Manufacturer Reason
for Recall
Premature deterioration of the vacuum tubing in the cell rinse unit due to excess bending stress, which may cause pinholes, cracks or tubing rupture, and fluid to drop into random reaction cells without any system alarm, increasing the risk of erroneous patient and QC results.
FDA Determined
Cause 2
Component design/selection
ActionRoche issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 9/23/2026 via UPS Ground. The notice explained the issue, clinical significance, and provided the following guidance: "Customer Actions You can continue to use your cobas c 303 analytical unit. Until the affected tubing has been replaced at the next site visit by a Roche service representative, or in the event of a suspected rupture, perform the following actions: " Perform daily quality control (QC) measurement with QC panel of available tests on the analytical unit before releasing patient results. If you identify a systematic deviation of multiple QC results, immediately stop using the analytical unit and contact the Roche Support Network Customer Support Center at 1-800-428-2336 to be scheduled for a tubing replacement. " Continue to follow your standard laboratory operating procedures." For questions: contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336.
Quantity in Commerce1002 units
DistributionUS Nationwide distribution and territory of PR.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.