| | Class 2 Device Recall Roche cobas C 303 analytical unit |  |
| Date Initiated by Firm | September 23, 2026 |
| Date Posted | November 18, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0481-2027 |
| Recall Event ID |
99980 |
| 510(K)Number | K220134 |
| Product Classification |
Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
|
| Product | Roche cobas c 303 analytical unit, Material Number 09793194001, clinical chemistry analyzer |
| Code Information |
UDI/DI 07613336220153, All units up to and including SN (Serial Number) 7696-10 |
| FEI Number |
1823260
|
Recalling Firm/ Manufacturer |
Roche Diagnostics Operations, Inc.
9115 Hague Rd
Indianapolis
IN
46256-1025
|
| For Additional Information Contact | Roche Support Network Customer Support Center 800-428-2336 |
Manufacturer Reason for Recall | Premature deterioration of the vacuum tubing in the cell rinse unit due to excess bending stress, which may cause pinholes, cracks or tubing rupture, and fluid to drop into random reaction cells without any system alarm, increasing the risk of erroneous patient and QC results. |
FDA Determined Cause 2 | Component design/selection |
| Action | Roche issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 9/23/2026 via UPS Ground. The notice explained the issue, clinical significance, and provided the following guidance:
"Customer Actions
You can continue to use your cobas c 303 analytical unit. Until the affected tubing has been replaced at the next site visit by a Roche service representative, or in the event of a suspected rupture, perform the following actions:
" Perform daily quality control (QC) measurement with QC panel of available tests on the analytical unit before releasing patient results. If you identify a systematic deviation of multiple QC results, immediately stop using the analytical unit and contact the Roche Support Network Customer Support Center at 1-800-428-2336 to be scheduled for a tubing replacement.
" Continue to follow your standard laboratory operating procedures."
For questions: contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336. |
| Quantity in Commerce | 1002 units |
| Distribution | US Nationwide distribution and territory of PR. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = JJE
|
|
|
|