 Class 2 Device Recall eNATsee related information
Date Initiated by FirmNovember 03, 2026
Date PostedNovember 25, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0552-2027
Recall Event ID 99999
510(K)NumberK201849 
Product Classification Microbial nucleic acid storage and stabilization device - Product Code QBD
ProducteNAT REF: REF: 6U073S01 and 6U074S01 UDI-DI code: 38053326009694, and 38053326009700 tube filled with non-nutritive transport medium and a flocked swab for sample collection (items 6U073S01 and 6U074S01), REF: 6U072S UDI-DI code: 38053326009687 tube filled with non-nutritive transport medium only (item 6U072S).
Code Information REF: 6U073S01 and 6U074S01 UDI-DI code: 38053326009694, and 38053326009700 Lot Numbers: 2106443 2211172 2215581 2315284 2403732 2408629 2409778 2410421 2414859 2419949 2503345 2512451 2604125 2605194 REF: 6U072S UDI-DI code: 38053326009687 Lot Numbers: 210472401 210734100 220358100 220937400 221096000 221532900 221532200 231407100 231475200 231528500 231584000 240604900 240726900 241323400 241365400 241412100 241750600 241412101 241486000 241791400 250090300 250334600 250549200 250632100 251373900 251493000 250741900 251605500 251693400 251868500 260118600 260554100 260359700 260658700 260788800 210644100 210572300 221095700 221532201 230218800 231582900 231407101 231528501 231848600 240107400 240726901 241042200 240977901 241697300 241486001 241946800 250151600 250090301 250334601 250297100 250423700 250514900 250632101 250833300 251001600 251373901 251245200 251054101 251125500 251166600
FEI Number 1000387145
Recalling Firm/
Manufacturer
Copan Italia Srl
Via Cesare Battisti 43/45
Bovezzo Italy
For Additional Information ContactMrs. Nichole Tunget
800-216-4016 Ext. 222
Manufacturer Reason
for Recall
Lack of Pre-Market Clearance. Several intended use statements in the product labeling have not been reviewed, approved/cleared by FDA.
FDA Determined
Cause 2
No Marketing Application
ActionOn 11/03/2026, an "URGENT: MEDICAL DEVICE CORRECTION" Letter was emailed to customers informing them that, Up to August 7, 2026 (date of www.copanusa.com website correction), the eNAT product pages web content included intended uses that have not undergone premarket review or received 510(k) clearance; therefore, the safety and effectiveness of these claims have not been evaluated by FDA. Specifically, web content published on www.copanusa.com described pathogens (including SARS-CoV-2, STIs, HPV, Chlamydia, Protozoa, and Mycoplasma), specimen collection sites (including nasal, urethral, nasopharyngeal, throat, vaginal, groin, axillary, rectal, wound, buccal, fecal, urine, and cervical sites), applications (including gastrointestinal diseases, genetics and microbiome analysis, and pharmacogenomics), and device characteristics (including complete inactivation of microbial viability within minutes and protection of polynucleotides against nuclease exposure) that are not encompassed by Copan's existing FDA clearance for the eNAT device. Customers are instructed to: -Confirm you have read and understood the information provided in this letter. -ATTACHMENT #1 shall be provided by email, within 7 working days from receipt, to: Customer Service Supervisor at Copan Diagnostics, Inc. Mob. 951.816.5679 - Toll Free: 800.216.4016 x222 Nichole.Tunget@copangroup.com 26055 Jefferson Ave Murrieta, CA 92562 -This notice should be passed on to all those who need to be aware within your organization. -If you have further distributed the devices, provide those Customers with a copy of the recall letter and related attachments. -Review your current procedures and processes to align them with Copan's cleared claim and intended use of eNAT devices. -Discard any content containing the non-cleared information, described above and included in eNAT product pages published on www.copanusa.com website until August 7, 2026.
Quantity in Commerce851400
DistributionU.S. Nationwide distribution in the states of AL, CA, CT, DC, FL, GA, HI, IA, IL, MA, MD, NC, NH, NJ, NM, NV, NY, OH, OR, PA, RI, TX, UT, VA, WA, and WI.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = QBD
-
-