Class 2 Device Recall TCD Electrodes

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall TCD Electrodessee related information
Date Initiated by FirmOctober 05, 2026
Date PostedNovember 06, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0368-2027
Recall Event ID 100005
510(K)NumberK082012 
Product Classification Probe, radiofrequency lesion - Product Code GXI
ProductTCD Electrodes (Description/REF(Inner/Outer)): TCD Electrode, Multi-Pack, 3M Cable, 10cm x 28ga (0.36mm)/(TCD-10-3M/TCD-10-3M-P); TCD Electrode, Multi-Pack, 10cm x 28ga (0.36mm)/(TCD-10/TCD-10-P); TCD Electrode, Multi-Pack, 3M Cable, 15cm x 28ga (0.36mm)/(TCD-15-3M/TCD-15-3M-P); TCD Electrode, Multi-Pack, 15cm x 28ga (0.36mm)/(TCD-15/TCD-15-P); TCD Electrode, Multi-Pack, 5cm x 28ga (0.36mm)/(TCD-5/TCD-5-P);
Code Information Inner REF/UDI-DI/Lot(Expiration): TCD-10-3M/00813250014396/39749333(7/6/2030), 85009822(7/12/2030); TCD-10/00813250010978/39741131(7/5/2030), 39750861(7/6/2030), 39757187(7/7/2030), 39766608(7/8/2030), 39768774(7/8/2030), 39772208(7/9/2030), 39778427(7/9/2030), 39780344(7/9/2030), 39781367(7/10/2030), 85006486(7/12/2030), 85011925(7/13/2030); TCD-15-3M/00813250014402/39770714(7/8/2030); TCD-15/00813250010985/39741129(7/5/2030), 39746187(7/6/2030), 39780349(7/10/2030), 39783782(7/10/2030); TCD-5/00813250010961/39751579(7/7/2030), 39783599(7/10/2030)
FEI Number 3006630150
Recalling Firm/
Manufacturer
Boston Scientific Neuromodulation Corporation
25155 Rye Canyon Loop
Valencia CA 91355-5004
For Additional Information ContactRenee Archie
508-683-4523
Manufacturer Reason
for Recall
Radiofrequency probes may have small cut marks on the Tyvek side of the sterile pouch, and if the sterile barrier is compromised and devices are used, this could cause microbial or other foreign material contamination to be introduced into patient tissue, which may lead to infection, pain, inflammation, allergic reaction, toxicity, the need for additional medical and/or surgical intervention.
FDA Determined
Cause 2
Packaging change control
ActionOn 10/5/2026, removal notices were mailed to Materials Managers and Healthcare Professionals who were asked to do the following: 1) Stop further use or distribution and segregate affected product. Because the cut marks associated with this removal may be small and difficult to identify, do not rely on visual inspection to determine whether products are safe to use. 2) No additional follow-up is recommended for patients previously treated with affected product. 3) Return product to Boston Scientific in accordance with the notice instructions. 4) Forward this notice to relevant personnel and any facilities to which the product was transferred. 5) Complete and return the Reply Verification Tracking Form via email to BSCFieldActionCenter@bsci.com Report any adverse events or quality concerns associated with use of these devices to the firm via email at BSCComplaints@bsci.com
Quantity in Commerce13320
DistributionUS Nationwide distribution including in the states of OH, ID, KY, NJ, GA, NY, MO, SC, MN, IA, MI, MA, WI, TX, ME, IN, AR, CO, IL, VT, VA, MS, WA, OR, NV, AZ, UT, FL, WY, CA, OK, NC, TN, MD, DE, LA, HI, CT, WV, AL, NH, ND.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.