| | Class 2 Device Recall Esenta Stomahesive Paste |  |
| Date Initiated by Firm | September 25, 2026 |
| Date Posted | November 25, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0553-2027 |
| Recall Event ID |
100015 |
| Product Classification |
Protector, ostomy - Product Code EXE
|
| Product | Generic Name: ESENTA
Generic Name: Stomahesive Paste
Model/Catalog Number: 183910
Software Version: N/A
Product Description: Stomahesive Paste is a hydrocolloid based protective barrier. It is designed for use in conjunction with Stomahesive and Durahesive skin barrier/ wafers in the protection of the skin around the stoma. It is used as a filler or caulking substance to increase skin barrier wear time by providing a flat pouching surface and
helping prevent leakage under the skin barrier.
Component: N/A |
| Code Information |
Lot Code: US Affected Product Only
Product Name: ESENTA S/HESIVE PASTE 1TUBEX56.7G
Primary UDI: 00768455209255
Product Code: 183910
Lot: 5J04201, 5K00858, 5L00433, 5L02930, 6A03082, 6B00827, 6E01386, 6E03598
|
| FEI Number |
3011987967
|
Recalling Firm/ Manufacturer |
ConvaTec, Inc
7815 National Service Rd Ste 600
Greensboro
NC
27409-9403
|
| For Additional Information Contact | Ashley Sweetland 336-5473730 |
Manufacturer Reason for Recall | Due to quality issue involving presence of foreign material in product. |
FDA Determined Cause 2 | Process control |
| Action | On 10/14/2026, an "URGENT MEDICAL DEVICE RECAL" Letter was emailed to customers informing them that, Convatec is conducting a voluntary field action (removal) of specific lots of ESENTA" Stomahesive Paste due to a quality related issue involving the potential presence of foreign matter within the product.
Distributors and Healthcare facilities are instructed to:
"Examine inventory and identify all affected product lots.
"Immediately quarantine and discontinue distribution, sale, transfer, and use of all affected product to prevent further distribution or use.
"Inform all appropriate personnel within your organization, including those responsible for inventory management, purchasing, distribution, and product use, of this recall action.
"If you have distributed any affected product further, immediately identify all customers who received the affected lots and notify them of this corrective action. A copy of this communication can be provided accompanied with a cover letter from your organization. Maintain records of all customer notifications and provide confirmation of completion upon request.
"If affected product is identified within your inventory, Convatec will provide a Certificate of Destruction (COD) for completion.
"Following destruction, complete and return the signed Certificate of Destruction (COD) to Convatec as evidence to support compliance with this action.
Users/Healthcare Professionals:
"Stop using the product immediately.
"Set the product aside and do not use it.
"Contact your healthcare provider, supplier, pharmacy, or Convatec representative (me+ Team) for further instructions.
"Retain the product and lot number information until you receive further guidance.
"Report any adverse events or complaints associated with the affected device to Convatec and applicable regulatory authorities, as required.
For questions - contact customer service at cic@convatec.com or 1-800-422-8811
|
| Quantity in Commerce | 193261 |
| Distribution | Worldwide - U.S. Nationwide distribution including in the states of Alabama, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, Florida, Georgia, Hawaii, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Nebraska, Nevada, New Jersey, New Mexico, New York, North Carolina, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Washington, West Virginia, and Wisconsin. The countries of Argentina, Bahrain, Belgium, Bosnia-Herz, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Estonia, Finland, Germany, Greece, Hong Kong, Hungary, Iceland, India, Iran, Ireland, Israel, Italy, Kenya, Lebanon, Lithuania, Luxembourg, Macedonia, Malta, Mexico, Netherlands, New Zealand, Norway, Poland, Portugal, Qatar, Romania, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Kingdom, and UAE |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|