Class 2 Device Recall Atellica CH Total Protein_2 (TP_2)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Atellica CH Total Protein_2 (TP_2)see related information
Date Initiated by FirmSeptember 22, 2026
Date PostedNovember 06, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0365-2027
Recall Event ID 100035
Product Classification Biuret (colorimetric), total protein - Product Code CEK
ProductAtellica CH Total Protein_2 (TP_2); Siemens Material Number (SMN): 1. 11645755 (3700 tests), 2. 11537228 (7400 tests);
Code Information 1. Siemens Material Number (SMN): 11645755 (3700 tests); UDI-DI: 00630414654010; All lots; 2. Siemens Material Number (SMN): 11537228 (7400 tests); UDI-DI: 00630414611082; All lots;
FEI Number 2432235
Recalling Firm/
Manufacturer
Siemens Healthcare Diagnostics, Inc.
511 Benedict Ave
Tarrytown NY 10591-5005
For Additional Information ContactJennifer Sacks
443-793-7288
Manufacturer Reason
for Recall
Reagent probe carryover on the Atellica CH and Atellica CI Analyzers may cause falsely elevated Hemoglobin A1c (A1c_E/A1c_H) results due to HDLC carryover (biases of 9 29% in NGSP units), and falsely depressed Direct Bilirubin 2 (DBil_2) results due to TP_2 carryover (biases of 0% to -25%), which may affect patient samples, quality control, and/or calibrator results.
FDA Determined
Cause 2
Software Design Change
ActionOn September 29, 2026, Urgent Medical Device Correction letters were sent to customers. Customer Actions: The Clean 2 probe wash is incorporated in the following Test Definitions Update Packages, which will be available soon: o Atellica Solution CH Test Definition Update Package Version 1.2.1 o Atellica CI Analyzer CH Test Definition Update Package Version 1.2.1 Note: Customers whose systems are connected to Siemens Remote Services (SRS) may self-install the applicable version. - To receive the updated Test Definitions, confirm that your systems are running the following commercially available system software versions: o Atellica CH Analyzer Software Version 1.31.1 o Atellica CI Analyzer Software Version 1.31.51 Note: The updated Test Definitions can be installed only after the system has been updated to the applicable system software version listed above. - Once the applicable system software version has been installed and Test Definition Update Package Version 1.2.1 becomes available on your system, follow the installation instructions in the accompanying Release Notes. - Until the applicable system software versions and Test Definitions have been installed, follow the interim instructions in Appendix B or Appendix C, depending on your current software version. - If both the applicable system software version and updated Test Definitions are already installed, no further system action is required. For instructions to verify if the Test Definition Update Package Version 1.2.1 has been installed, see Appendix D. - Complete and return the Field Correction Effectiveness Check Form attached to this letter within (30) days. - Please retain this letter with your laboratory records and forward this letter to those who may have received this product.
Quantity in Commerce25210 units (8087 US, 17123 OUS)
DistributionWorldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Belgium, Bolivia, Brazil, Bulgaria, Burkina Faso, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cura¿ao, Cura¿ao, St Eus, Czech Republic, Democratic Republic Congo, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Ireland, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Maldives, Mexico, Morocco, Nepal, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Turkmenistan, U.A.E., Uganda, United Kingdom , Uruguay, Uzbekistan, Vatikancity, Vietnam.