• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Smartmix CTS Vacuum Mixing Bowl

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Smartmix CTS Vacuum Mixing Bowlsee related information
Date Initiated by FirmSeptember 02, 2026
Date PostedNovember 17, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0471-2027
Recall Event ID 100050
Product Classification Mixer, cement, for clinical use - Product Code JDZ
ProductVacuum mixing bowl for PMMA bone cement
Code Information UDI-DI: 15060238200070. Lots: 35080, 35081. Expiration: 06/09/2029
FEI Number 1000345108
Recalling Firm/
Manufacturer
Summit Medical Ltd.
Bourton Industrial Park
Bourton on the Water
Gloucestershire United Kingdom
For Additional Information Contact
+44 1451 821311
Manufacturer Reason
for Recall
Vacuum mixing bowl for bone cement has scraper components that, due to a processing error, may lead finished device endotoxin levels to be out of specification.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionOn 9/2/2026, the recall notice was emailed to the customer who was asked to do the following: 1) Destruction of the listed product must be carried out in such a way as to prevent future use of the affected product. 2) Confirm the destruction by completing and returning the Acknowledgement and Receipt Form via email to cs@summit-medical.co.uk Should you have any concerns or questions, please contact the firm's customer care team via email at cs@summit-medical.co.uk
Quantity in Commerce2850
DistributionUS Nationwide distribution in the state of IN.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-