| | Class 2 Device Recall iSTAT hsTnI cartridge |  |
| Date Initiated by Firm | October 12, 2026 |
| Date Posted | November 20, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0504-2027 |
| Recall Event ID |
100056 |
| 510(K)Number | K240984 |
| Product Classification |
Immunoassay method, troponin subunit - Product Code MMI
|
| Product | i-STAT hs-TnI cartridge;
List Number: 09P81-25; |
| Code Information |
List Number: 09P81-25;
UDI-DI: 00054749003051;
Lot Numbers: All lots beginning with: A24XXX, A25XXX, B25XXX, B26XXX, C26XXX, D26XXX; |
| FEI Number |
2245578
|
Recalling Firm/ Manufacturer |
Abbott Point Of Care Inc.
400 College Rd E
Princeton
NJ
08540-6607
|
| For Additional Information Contact | Amber Dutra 224-416-2807 |
Manufacturer Reason for Recall | Variability associated with a component used during reagent manufacturing which may result in falsely
elevated cardiac troponin I results which may lead to unnecessary clinical intervention (e.g.
cardiac catheterization without true evidence of myocardial injury). |
FDA Determined Cause 2 | Process control |
| Action | On October 12, 2026, URGENT MEDICAL DEVICE CORRECTION letters were sent to customers.
Actions to be taken:
1. Clinicians should be advised that serial cardiac troponin I results may not be helpful to distinguish true elevation of cTnI results versus elevated results due to interference; therefore the clinician should interpret results in conjunction with the patient's signs, symptoms, risk factors, medical history, physical examination findings, and results of other diagnostic testing. If results are inconsistent with the
patient's clinical presentation, the possibility of analytical interference should be considered.
2. If analytical interference is suspected based on the patient's clinical presentation, retest the sample using an i-STAT hs-TnI cartridge lot beginning with A26 (when available) or test the sample using an alternative validated laboratory method.
3. Complete and return the Business Reply Card included with this notification to confirm receipt and understanding. If you have forwarded any of the i-STAT hs-TnI cartridges to another facility, we request that you please provide a copy of this letter to them. |
| Quantity in Commerce | 3431000 (1996950 US, 1434050 OUS) |
| Distribution | Worldwide distribution - US Nationwide and the countries of Australia, Germany, Philippines, Austria, Greece, Poland, Bahrain, Hong Kong, Portugal, Belgium, IRELAND, Saudi Arabia, Canada, ITALY, Slovenia, Carribean countries, Kuwait, South Africa, Chile, Lebanon, Sweden, Colombia, Luxembourg, Switzerland, Denmark, Netherlands, UAE, Finland, New Zealand, United Kingdom, France, Norway, Uruguay. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = MMI
|
|
|
|