| | Class 2 Device Recall FilmArray GI Control Panel M238 |  |
| Date Initiated by Firm | October 22, 2026 |
| Date Posted | November 20, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0505-2027 |
| Recall Event ID |
100057 |
| 510(K)Number | K251526 |
| Product Classification |
Assayed external control material for microbiology nucleic acid amplification (NAT) assays - Product Code PMN
|
| Product | FilmArray GI Control Panel M238;
Model/Catalog Number: M238;
Product Description: FilmArray GI Control Panel M238, P/N M238, is a quality control panel consisting of 2 single use, ready-to-use, liquid controls, FilmArray GI Control M239, P/N M239, and FilmArray GI Control M240, P/N M240. Each kit of FilmArray GI Control Panel M238 is comprised of 6 tubes of M239 and 6 tubes of M240. Together the 2 controls contain synthetic RNA corresponding to genome segments of all the pathogens detected by the BIOFIRE FILMARRAY Gastrointestinal (GI) Panel and BIOFIRE FILMARRAY GI Panel Mid assays, suspended in a non-infectious solution of buffers, preservatives and stabilizers.
Each liquid control of FilmArray GI Control Panel M238 is processed separately according to the BIOFIRE GI Panel and BIOFIRE GI Panel Mid manufacturer s Instructions for Use for patient samples, stool samples in Cary Blair transport media obtained from individuals with signs and/or symptoms of gastrointestinal infection.
Component: N/A |
| Code Information |
Model/Catalog Number: M238;
UDI: 00852720008275;
Lot Number: B04AUG26; |
| FEI Number |
3005959679
|
Recalling Firm/ Manufacturer |
Maine Molecular Quality Controls, Inc.
23 Mill Brook Rd
Saco
ME
04072-9806
|
| For Additional Information Contact | Joan Gordon 1-207-8851072 Ext. 201 |
Manufacturer Reason for Recall | Some devices are mislabeled with "Research Use Only (ROU). |
FDA Determined Cause 2 | Labeling mix-ups |
| Action | On October 22, 2026 Urgent Medical Device Update letters were sent to customers.
Actions Required by Customers
Please take the following actions as soon as possible:
1. If you have distributed any affected product to other facilities or customers, promptly forward this notification to those parties and follow the instructions below to complete the attached Recall Response Form.
2. Complete the attached Recall Response Form:
" Inspect all FilmArray GI Control Panel M238 kits from Lot B04AUG26, for FilmArray GI
Control M240 tubes labeled RUO.
" If no RUO tubes are identified, enter O for both the number of tubes and number of kits in the Response Form. After the Response Form is returned to MMQCI, no further action is required.
" If RUO tubes are identified, enter the number of RUO tubes and the number of kits in the Response Form.
" If the laboratory would like a replacement kit (s) for any kit that included an RUO tube(s), enter the number of kits needed for replacement on the Response Form.
" If the laboratory would like to use the FilmArray GI Control M240 tube(s) labeled RUO,
indicate your intention to do so on the Response Form.
" If the laboratory is requesting replacements, the RUO tubes(s) should be discarded according to your local, state, and federal regulations.
" Return the Response Form to OARA@mmqci.com
3. Notify all personnel within your organization who use or manage this product of this notification.
Please ensure they are aware that the results previously obtained using the RUO-labeled tubes remain accurate. |
| Quantity in Commerce | 448 |
| Distribution | Worldwide distribution - US Nationwide and the country of Germany. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = PMN
|
|
|
|