| Date Initiated by Firm | May 15, 2026 |
| Date Posted | November 06, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0380-2027 |
| Recall Event ID |
100063 |
| PMA Number | P180051 |
| Product Classification |
Organ care system (OCS) heart system - Product Code QIK
|
| Product | OCS Heart Perfusion Set
Model/Catalog Number: 1200
Software Version: rev 3
Product Description: OCS Heart Perfusion Module,
included in the Heart Perfusion Set used for organ transport. |
| Code Information |
1. Perfusion Module SN: 4822-8-22505;
Description: OCS Heart Perfusion Set ;
EXP Date: 4/9/29;
UDI: *+B00948225/$$+74822-8-22505/14D20290409C*;
2. Perfusion Module SN: 4822-8-23601;
Description: OCS Heart Perfusion Set;
EXP Date: 7/27/29;
UDI: *+B00948225/$$+74822-8-23601/14D20290727D*;
3. Perfusion Module SN: 4822-8-22496;
Description: OCS Heart Perfusion Set;
EXP Date: 4/9/29;
UDI: *+B00948225/$$+74822-8-22496/14D20290409L*;
4. Perfusion Module SN: 4822-8-23497;
Description: OCS Heart Perfusion Set;
EXP Date: 4/9/29;
UDI: *+B00948225/$$+74822-8-23497/14D20290409N*;
5. Perfusion Module SN: 4822-8-22479;
Description: OCS Heart Perfusion Set;
EXP Date: 4/9/29;
UDI: *+B00948225/$$+74822-8-22479/14D20290409M*;
6. Perfusion Module SN: 4822-8-22504;
Description: OCS Heart Perfusion Set;
EXP Date: 4/9/29;
UDI: *+B00948225/$$+74822-8-22504/14D20290409B*;
7. Perfusion Module SN: 4822-8-23604;
Description: OCS Heart Perfusion Set;
EXP Date: 7/13/29;
UDI: *+B00948225/$$+74822-8-23604/14D20290713B*;
8. Perfusion Module SN: 4822-8-23600;
Description: OCS Heart Perfusion Set;
EXP Date: 7/13/29;
UDI: *+B00948225/$$+74822-8-23600/14D202907137*;
9. Perfusion Module SN: 4822-8-23499;
Description: OCS Heart Perfusion Set;
EXP Date: 7/13/29;
UDI: *+B00948225/$$+74822-8-23499/14D20290713N*;
10. Perfusion Module SN: 4822-8-23602;
Description: OCS Heart Perfusion Set;
EXP Date: 7/13/29;
UDI: *+B00948225/$$+74822-8-23602/14D202907139*;
11. Perfusion Module SN: 4822-8-23674;
Description: OCS Heart Perfusion Set;
EXP Date: 7/13/29;
UDI: *+B00948225/$$+74822-8-23674/14D20290713I*;
12. Perfusion Module SN: 4822-8-22460;
Description: OCS Heart Perfusion Set;
EXP Date: 4/9/29;
UDI: *+B00948225/$$+74822-8-22460/14D20290409C*;
13. Perfusion Module SN: 4822-8-23675;
Description: OCS Heart Perfusion Set;
EXP Date: 7/13/29;
UDI: *+B00948225/$$+74822-8-23675/14D20290713J*;
14. Perfusion Module SN: 4822-8-22481;
Description: OCS Heart Perfusion Set;
EXP Date: 4/9/29;
UDI: *+B00948225/$$+74822-8-22481/14D20290409F*;
15. Perfusion Module SN: 4822-9-22492;
Description: OCS Heart Perfusion Set;
EXP Date: 4/9/29;
UDI: *+B00948225/$$+74822-9-22492/14D20290409I*;
|
| FEI Number |
3003152463
|
Recalling Firm/ Manufacturer |
TransMedics, Inc.
200 Minuteman Rd Ste 302
Andover
MA
01810-1046
|
| For Additional Information Contact | Christy Foreman 978-8098795 |
Manufacturer Reason for Recall | The Perfusion Module may have an unprogrammed Front End Board (FEB) which will not allow the system to proceed beyond the initial startup process. The OCS system will display an error message saying Unsupported Perfusion Module Present and will prevent further system use. |
FDA Determined Cause 2 | Process control |
| Action | On or about May 15, 2026 and June 29, 2026 OCS Heart Perfusion Set Device Recall
Direct Customer Notifications, OCS Liver Perfusion Set Device Recall
Direct Customer Notifications, and OCS Lung Mission Critical Perfusion Set Device Recall
Direct Customer Notifications, were sent to customers.
Actions to be taken:
TransMedics requests that you return any unused affected units to us at your convenience.
Actions Taken by TransMedics
We plan to ship replacement units to you to replace the affected product.
|
| Quantity in Commerce | 6 |
| Distribution | Worldwide - US Nationwide distribution in the states of CA, FL, IL, MA, MN, NC, NE, NY, TX and the country of Serbia;. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| PMA Database | PMAs with Product Code = QIK
|