Class 2 Device Recall TruSystem 7000 U Operating Table (dV)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall TruSystem 7000 U Operating Table (dV)see related information
Date Initiated by FirmSeptember 14, 2026
Date PostedNovember 03, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0333-2027
Recall Event ID 99826
Product Classification Table, operating-room, electrical - Product Code GDC
ProductMobile column TruSystem 7500 U, Product Code 1730720
Code Information UDI-DI: 00887761974388; Serial number: 110183858
FEI Number 1416980
Recalling Firm/
Manufacturer
Baxter Healthcare Corporation
25212 W Il Route 120
Round Lake IL 60073-9799
For Additional Information ContactCenter for One Baxter
1-847-9484770
Manufacturer Reason
for Recall
The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others.
FDA Determined
Cause 2
Equipment maintenance
ActionCustomers were notified by telephone beginning 9/14/26. Baxter Healthcare Corporation is issuing a voluntary correction for the TruSystem Surgical Tables after identifying that four units underwent spindle replacement without the required belt tension verification. The affected units were released without confirmation that belt tension met the specified requirements, which could lead to unintended downward movement of the table column. Unintended downward movement of the surgical table could occur, potentially resulting in disruption of patient positioning, reduced perfusion or oxygenation, musculoskeletal injury and/or interruption of a surgical procedure. To date, Baxter has not received any reports of patient injury related to this issue. Baxter will call to schedule a time to come in and perform the belt tension verification. Customers can continue to use the tables.
Quantity in Commerce1 unit
DistributionUS Nationwide distribution in the states of IL, MO, TX.