| | Class 1 Device Recall Cardinal Health Presource |  |
| Date Initiated by Firm | September 24, 2026 |
| Date Posted | November 02, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0221-2027 |
| Recall Event ID |
100058 |
| Product Classification |
Cesarean section tray - Product Code OHM
|
| Product | Cardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump Module Model 8100, labeled as:
1) C SECTION, ADM, KIT , CATALOG #MA3HD0326B;
2) C SECTION, ADM, KIT , CATALOG #MA3HD03263;
3) C SECTION, ADM, KIT , CATALOG #MA1V125141;
4) C SECTION, ADM, KIT , CATALOG #MA3HD03262;
5) C SECTION, ADM ,KIT , CATALOG #MA3HD03261;
6) KIT,C SECTION ADMIT, , CATALOG #MATJV8300B;
7) C SECTION, ADM, KIT , CATALOG #MA1V12514A;
8) KIT, C SECTION, ADM , CATALOG #MA1V12514A;
9) KIT,C SECTION ADMIT, , CATALOG #MATJV8300A;
10) C SECTION, ADMISSION KIT , CATALOG #PLHDCAHC2;
11) C SECTION, ADMISSION ,KIT , CATALOG #PLHDCAHCO;
12) C SECTION, ADMISSION KIT , CATALOG #PLHDCAHC1;
13) KIT,C SECTION ADMIT, , CATALOG #PZJVADMMX;
14) KIT,C SECTION ADMIT,, CATALOG #PZJVADMMW;
15) C SECTION ADMIT,KIT, CATALOG #PZJVADMMV;
16) C SECTION, ADMISSION KIT , CATALOG #PLHDCAHCP;
17) C SECTION, ADM KIT , CATALOG #MA3HD0326A;
18) C SECTION, ADM, KIT , CATALOG #MA3HD0326A;
19) KIT, C SECTION, ADM , CATALOG #MA3HD0326A. |
| Code Information |
CATALOG # MA3HD0326B, GTIN 10887488229751 (ea) 50887488229759 (cs), Lot # 331735;
CATALOG # MA3HD0326B, GTIN 10887488229751 (ea) 50887488229759 (cs), Lot # 331735;
CATALOG # MA3HD0326B, GTIN 10887488229751 (ea) 50887488229759 (cs), Lot # 339023;
CATALOG # MA3HD0326B, GTIN 10887488229751 (ea) 50887488229759 (cs), Lot # 339023;
CATALOG # MA3HD0326B, GTIN 10887488229751 (ea) 50887488229759 (cs), Lot # 341402;
CATALOG # MA3HD0326B, GTIN 10887488229751 (ea) 50887488229759 (cs), Lot # 341402;
CATALOG # MA3HD0326B, GTIN 10887488229751 (ea) 50887488229759 (cs), Lot # 341402;
CATALOG # MA3HD0326B, GTIN 10887488229751 (ea) 50887488229759 (cs), Lot # 351145;
CATALOG # MA3HD0326B, GTIN 10887488229751 (ea) 50887488229759 (cs), Lot # 351145;
CATALOG # MA3HD0326B, GTIN 10887488229751 (ea) 50887488229759 (cs), Lot # 351145;
CATALOG # MA3HD0326B, GTIN 10887488229751 (ea) 50887488229759 (cs), Lot # 356177;
CATALOG # MA3HD0326B, GTIN 10887488229751 (ea) 50887488229759 (cs), Lot # 356177;
CATALOG # MA3HD0326B, GTIN 10887488229751 (ea) 50887488229759 (cs), Lot # 356177;
CATALOG # MA3HD0326B, GTIN 10887488229751 (ea) 50887488229759 (cs), Lot # 361881;
CATALOG # MA3HD0326B, GTIN 10887488229751 (ea) 50887488229759 (cs), Lot # 361881;
CATALOG # MA3HD0326B, GTIN 10887488229751 (ea) 50887488229759 (cs), Lot # 361881;
CATALOG # MA3HD0326B, GTIN 10887488229751 (ea) 50887488229759 (cs), Lot # 361881;
CATALOG # MA3HD0326B, GTIN 10887488229751 (ea) 50887488229759 (cs), Lot # 369350;
CATALOG # MA3HD0326B, GTIN 10887488229751 (ea) 50887488229759 (cs), Lot # 369350;
CATALOG # MA3HD0326B, GTIN 10887488229751 (ea) 50887488229759 (cs), Lot # 375879;
CATALOG # MA3HD0326B, GTIN 10887488229751 (ea) 50887488229759 (cs), Lot # 375879;
CATALOG # MA3HD0326B, GTIN 10887488229751 (ea) 50887488229759 (cs), Lot # 375879;
CATALOG # MA3HD0326B, GTIN 10887488229751 (ea) 50887488229759 (cs), Lot # 381100;
CATALOG # MA3HD0326B, GTIN 10887488229751 (ea) 50887488229759 (cs), Lot # 381117;
CATALOG # MA3HD0326B, GTIN 10887488229751 (ea) 50887488229759 (cs), Lot # 381117;
CATALOG # MA3HD0326B, GTIN 10887488229751 (ea) 50887488229759 (cs), Lot # 381117;
CATALOG # MA3HD03263, GTIN 10887488224343 (ea) 50887488224341 (cs), Lot # 326029;
CATALOG # MA3HD03263, GTIN 10887488224343 (ea) 50887488224341 (cs), Lot # 326029;
CATALOG # MA3HD03263, GTIN 10887488224343 (ea) 50887488224341 (cs), Lot # 326029;
CATALOG # MA1V125141, GTIN 10887488212074 (ea) 50887488212072 (cs), Lot # 323763;
CATALOG # MA3HD03262, GTIN 10887488181394 (ea) 50887488181392 (cs), Lot # 317367;
CATALOG # MA3HD03262, GTIN 10887488181394 (ea) 50887488181392 (cs), Lot # 317367;
CATALOG # MA3HD03261, GTIN 10887488178172 (ea) 50887488178170 (cs), Lot # 315984;
CATALOG # MATJV8300B, GTIN 10887488138411 (ea) 50887488138419 (cs), Lot # 263376;
CATALOG # MATJV8300B, GTIN 10887488138411 (ea) 50887488138419 (cs), Lot # 295952;
CATALOG # MATJV8300B, GTIN 10887488138411 (ea) 50887488138419 (cs), Lot # 316164;
CATALOG # MATJV8300B, GTIN 10887488138411 (ea) 50887488138419 (cs), Lot # 316164;
CATALOG # MATJV8300B, GTIN 10887488138411 (ea) 50887488138419 (cs), Lot # 332772;
CATALOG # MATJV8300B, GTIN 10887488138411 (ea) 50887488138419 (cs), Lot # 332772;
CATALOG # MATJV8300B, GTIN 10887488138411 (ea) 50887488138419 (cs), Lot # 345148;
CATALOG # MATJV8300B, GTIN 10887488138411 (ea) 50887488138419 (cs), Lot # 345148;
CATALOG # MATJV8300B, GTIN 10887488138411 (ea) 50887488138419 (cs), Lot # 345148;
CATALOG # MATJV8300B, GTIN 10887488138411 (ea) 50887488138419 (cs), Lot # 365913;
CATALOG # MATJV8300B, GTIN 10887488138411 (ea) 50887488138419 (cs), Lot # 376327;
CATALOG # MATJV8300B, GTIN 10887488138411 (ea) 50887488138419 (cs), Lot # 376327;
CATALOG # MATJV8300B, GTIN 10887488138411 (ea) 50887488138419 (cs), Lot # 398577;
CATALOG # MATJV8300B, GTIN 10887488138411 (ea) 50887488138419 (cs), Lot # 398577;
CATALOG # MATJV8300B, GTIN 10887488138411 (ea) 50887488138419 (cs), Lot # 398577;
CATALOG # MATJV8300B, GTIN 10887488138411 (ea) 50887488138419 (cs), Lot # 418270;
CATALOG # MATJV8300B, GTIN 10887488138411 (ea) 50887488138419 (cs), Lot # 418270;
CATALOG # MATJV8300B, GTIN 10887488138411 (ea) 50887488138419 (cs), Lot # 442158;
CATALOG # MATJV8300B, GTIN 10887488138411 (ea) 50887488138419 (cs), Lot # 445625;
CATALOG # MATJV8300B, GTIN 10887488138411 (ea) 50887488138419 (cs), Lot # 445625;
CATALOG # MATJV8300B, GTIN 10887488138411 (ea) 50887488138419 (cs), Lot # 445625;
CATALOG # MATJV8300B, GTIN 10887488138411 (ea) 50887488138419 (cs), Lot # 487330;
CATALOG # MATJV8300B, GTIN 10887488138411 (ea) 50887488138419 (cs), Lot # 487330;
CATALOG # MATJV8300B, GTIN 10887488138411 (ea) 50887488138419 (cs), Lot # 479198;
CATALOG # MA1V12514A, GTIN 10887488134222 (ea) 50887488134220 (cs), Lot # 322269;
CATALOG # MA1V12514A, GTIN 10887488134222 (ea) 50887488134220 (cs), Lot # 310120;
CATALOG # MATJV8300A, GTIN 10887488061665 (ea) 50887488061663 (cs), Lot # 221039;
CATALOG # MATJV8300A, GTIN 10887488061665 (ea) 50887488061663 (cs), Lot # 221039;
CATALOG # MATJV8300A, GTIN 10887488061665 (ea) 50887488061663 (cs), Lot # 221039;
CATALOG # MATJV8300A, GTIN 10887488061665 (ea) 50887488061663 (cs), Lot # 230037;
CATALOG # MATJV8300A, GTIN 10887488061665 (ea) 50887488061663 (cs), Lot # 230037;
CATALOG # MATJV8300A, GTIN 10887488061665 (ea) 50887488061663 (cs), Lot # 240709;
CATALOG # MATJV8300A, GTIN 10887488061665 (ea) 50887488061663 (cs), Lot # 240709;
CATALOG # PLHDCAHC2, GTIN 8854251476976 (ea) 58854251476971 (cs), Lot # 117230;
CATALOG # PLHDCAHCO, GTIN 8854251476297 (ea) 58854251476292 (cs), Lot # 126959;
CATALOG # PLHDCAHCO, GTIN 8854251476297 (ea) 58854251476292 (cs), Lot # 126959;
CATALOG # PLHDCAHCO, GTIN 8854251476297 (ea) 58854251476292 (cs), Lot # 131445;
CATALOG # PLHDCAHCO, GTIN 8854251476297 (ea) 58854251476292 (cs), Lot # 131445;
CATALOG # PLHDCAHCO, GTIN 8854251476297 (ea) 58854251476292 (cs), Lot # 131445;
CATALOG # PLHDCAHCO, GTIN 8854251476297 (ea) 58854251476292 (cs), Lot # 131445;
CATALOG # PLHDCAHCO, GTIN 8854251476297 (ea) 58854251476292 (cs), Lot # 131445;
CATALOG # PLHDCAHC1, GTIN 8854251374395 (ea) 58854251374390 (cs), Lot # 87291;
CATALOG # PZJVADMMX, GTIN 885425888354 (ea) 50885425888359 (cs), Lot # 70960;
CATALOG # PZJVADMMX, GTIN 885425888354 (ea) 50885425888359 (cs), Lot # 85735;
CATALOG # PZJVADMMX, GTIN 885425888354 (ea) 50885425888359 (cs), Lot # 100525;
CATALOG # PZJVADMMX, GTIN 885425888354 (ea) 50885425888359 (cs), Lot # 118988;
CATALOG # PZJVADMMX, GTIN 885425888354 (ea) 50885425888359 (cs), Lot # 143949;
CATALOG # PZJVADMMX, GTIN 885425888354 (ea) 50885425888359 (cs), Lot # 153492;
CATALOG # PZJVADMMX, GTIN 885425888354 (ea) 50885425888359 (cs), Lot # 153492;
CATALOG # PZJVADMMX, GTIN 885425888354 (ea) 50885425888359 (cs), Lot # 196575;
CATALOG # PZJVADMMX, GTIN 885425888354 (ea) 50885425888359 (cs), Lot # 196575;
CATALOG # PZJVADMMW, GTIN 885425805870 (ea) 50885425805875 (cs), Lot # 24192;
CATALOG # PZJVADMMW, GTIN 885425805870 (ea) 50885425805875 (cs), Lot # 27725;
CATALOG # PZJVADMMW, GTIN 885425805870 (ea) 50885425805875 (cs), Lot # 38222;
CATALOG # PZJVADMMV, GTIN 885425236384 (ea) 50885425236389 (cs), Lot # 85448;
CATALOG # PZJVADMMV, GTIN 885425236384 (ea) 50885425236389 (cs), Lot # 99679;
CATALOG # PZJVADMMV, GTIN 885425236384 (ea) 50885425236389 (cs), Lot # 121945;
CATALOG # PZJVADMMV, GTIN 885425236384 (ea) 50885425236389 (cs), Lot # 131165;
CATALOG # PZJVADMMV, GTIN 885425236384 (ea) 50885425236389 (cs), Lot # 131165;
CATALOG # PZJVADMMV, GTIN 885425236384 (ea) 50885425236389 (cs), Lot # 131165;
CATALOG # PZJVADMMV, GTIN 885425236384 (ea) 50885425236389 (cs), Lot # 131165;
CATALOG # PZJVADMMV, GTIN 885425236384 (ea) 50885425236389 (cs), Lot # 140773;
CATALOG # PZJVADMMV, GTIN 885425236384 (ea) 50885425236389 (cs), Lot # 150991;
CATALOG # PZJVADMMV, GTIN 885425236384 (ea) 50885425236389 (cs), Lot # 150991;
CATALOG # PZJVADMMV, GTIN 885425236384 (ea) 50885425236389 (cs), Lot # 158662;
CATALOG # PZJVADMMV, GTIN 885425236384 (ea) 50885425236389 (cs), Lot # 158662;
CATALOG # PZJVADMMV, GTIN 885425236384 (ea) 50885425236389 (cs), Lot # 166757;
CATALOG # PZJVADMMV, GTIN 885425236384 (ea) 50885425236389 (cs), Lot # 175359;
CATALOG # PZJVADMMV, GTIN 885425236384 (ea) 50885425236389 (cs), Lot # 175359;
CATALOG # PZJVADMMV, GTIN 885425236384 (ea) 50885425236389 (cs), Lot # 186278;
CATALOG # PZJVADMMV, GTIN 885425236384 (ea) 50885425236389 (cs), Lot # 186278;
CATALOG # PZJVADMMV, GTIN 885425236384 (ea) 50885425236389 (cs), Lot # 186278;
CATALOG # PZJVADMMV, GTIN 885425236384 (ea) 50885425236389 (cs), Lot # 190944;
CATALOG # PZJVADMMV, GTIN 885425236384 (ea) 50885425236389 (cs), Lot # 190944;
CATALOG # PZJVADMMV, GTIN 885425236384 (ea) 50885425236389 (cs), Lot # 190944;
CATALOG # PZJVADMMV, GTIN 885425236384 (ea) 50885425236389 (cs), Lot # 196571;
CATALOG # PZJVADMMV, GTIN 885425236384 (ea) 50885425236389 (cs), Lot # 196571;
CATALOG # PZJVADMMV, GTIN 885425236384 (ea) 50885425236389 (cs), Lot # 204012;
CATALOG # PZJVADMMV, GTIN 885425236384 (ea) 50885425236389 (cs), Lot # 204012;
CATALOG # PZJVADMMV, GTIN 885425236384 (ea) 50885425236389 (cs), Lot # 204012;
CATALOG # PZJVADMMV, GTIN 885425236384 (ea) 50885425236389 (cs), Lot # 204012;
CATALOG # PZJVADMMV, GTIN 885425236384 (ea) 50885425236389 (cs), Lot # 204012;
CATALOG # PZJVADMMV, GTIN 885425236384 (ea) 50885425236389 (cs), Lot # 204012;
CATALOG # PZJVADMMV, GTIN 885425236384 (ea) 50885425236389 (cs), Lot # 214000;
CATALOG # PZJVADMMV, GTIN 885425236384 (ea) 50885425236389 (cs), Lot # 214000;
CATALOG # PZJVADMMV, GTIN 885425236384 (ea) 50885425236389 (cs), Lot # 222855;
CATALOG # PZJVADMMV, GTIN 885425236384 (ea) 50885425236389 (cs), Lot # 222855;
CATALOG # PZJVADMMV, GTIN 885425236384 (ea) 50885425236389 (cs), Lot # 222855;
CATALOG # PLHDCAHCP, GTIN 885425223544 (ea) 50885425223549 (cs), Lot # 141935;
CATALOG # PLHDCAHCP, GTIN 885425223544 (ea) 50885425223549 (cs), Lot # 141935;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 89795;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 117696;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 117696;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 117696;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 117696;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 132065;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 132065;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 138323;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 142458;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 147779;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 152213;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 156746;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 156746;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 162582;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 165381;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 168556;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 170469;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 175132;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 175132;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 175132;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 175132;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 175132;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 262725;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 262725;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 267881;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 267881;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 274948;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 279734;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 286196;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 286196;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 294184;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 294184;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 294184;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 298007;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 298007;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 298007;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 306961;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 309779;
CATALOG # MA3HD0326A, GTIN 885425110646 (ea) 50885425110641 (cs), Lot # 309779.
|
| FEI Number |
3001236905
|
Recalling Firm/ Manufacturer |
Cardinal Health 200, LLC
3651 Birchwood Dr
Waukegan
IL
60085-8337
|
| For Additional Information Contact | Cardinal Health Quality Systems 800-292-9332 |
Manufacturer Reason for Recall | Kits contain infusion sets which were recalled for performance issues. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | Cardinal Health issued an URGENT: MEDICAL DEVICE CORRECTION notice to its consignees on 9/24/2026 via US Mail. The notice explained the issue and requested the discontinuation of the use of the affected BD Alaris Pump Infusion Sets as per the enclosed notice from Becton Dickinson.
The notice instructed the following:
If you locate affected product within your facility, please do the following:
1. QUARANTINE affected kits.
2. ALERT your clinicians of this action.
3. AFFIX a WARNING LABEL to the front of each Presource kit so that it is clearly visible to
clinicians, instructing them to remove and discard the recalled component(s).
4. RETURN a copy of the enclosed Acknowledgment Form confirming your receipt of this
action and the completion of labeling product via fax to 614.652.9648.
Additional labels will be sent to your facility upon request. Please send an email to:
gmb-FieldCorrectiveAction@cardinalhealth.com
For component credit or replacement, please contact the following:
Hospitals Cardinal Customer Service at 866.551.0533
Distributors Distributor Management at 800.635.6021
Federal Government facilities Government Customer Service at 800.444.1166
Distributors were directed to notify their customers of the recall.
For questions contact: Cardinal Health Quality Systems at 800.292.9332. |
| Quantity in Commerce | 15247 kits |
| Distribution | US and Canada |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|