| | Class 2 Device Recall Azurion |  |
| Date Initiated by Firm | October 15, 2026 |
| Date Posted | November 18, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0482-2027 |
| Recall Event ID |
100086 |
| 510(K)Number | K200917 K172822 K181830 |
| Product Classification |
Interventional fluoroscopic x-ray system - Product Code OWB
|
| Product | Philips Azurion systems. Containing the below descriptions and corresponding Model Numbers.
1. Azurion 3 M12, Model Numbers: 722063, 722221.
2. Azurion 3 M15, Model Numbers: 722064, 722222.
3. Azurion 5 M12, Model Number: 722227.
4. Azurion 5 M20, Model Number: 722228.
5. Azurion 7 B12, Model Numbers: 722067, 722225.
6. Azurion 7 B20, Model Numbers: 722068, 722226.
7. Azurion 7 M12, Model Numbers: 722078, 722223.
8. Azurion 7 M20, Model Numbers: 722079, 722224. |
| Code Information |
1. Azurion 3 M12, Model Number (UDI) Serial Numbers: 722063 ((01)00884838085275(21)122, (01)00884838085275(21)131, (01)00884838085275(21)132, (01)00884838085275(21)96, (01)00884838085275(21)108, (01)00884838085275(21)145, (01)00884838085275(21)146, (01)00884838085275(21)43, (01)00884838085275(21)149, (01)00884838085275(21)144, (01)00884838085275(21)120, (01)00884838085275(21)110, (01)00884838085275(21)97, (01)00884838085275(21)127, (01)00884838085275(21)101, (01)00884838085275(21)143, (01)00884838085275(21)116, (01)00884838085275(21)151, (01)00884838085275(21)141, (01)00884838085275(21)137, (01)00884838085275(21)150, (01)00884838085275(21)128, (01)00884838085275(21)105, (01)00884838085275(21)129, (01)00884838085275(21)139, (01)00884838085275(21)130, (01)00884838085275(21)100, (01)00884838085275(21)135, (01)00884838085275(21)123, (01)00884838085275(21)107, (01)00884838085275(21)86, (01)00884838085275(21)133, (01)00884838085275(21)140, (01)00884838085275(21)148) 122, 131, 132, 96, 108, 145, 146, 43, 149, 144, 120, 110, 97, 127, 101, 143, 116, 151, 141, 137, 150, 128, 105, 129, 139, 130, 100, 135, 123, 107, 86, 133, 140, 148. 722221 ((01)00884838099203(21)3, (01)00884838099203(21)7, (01)00884838099203(21)6, (01)00884838099203(21)9, (01)00884838099203(21)19, (01)00884838099203(21)45, (01)00884838099203(21)51, (01)00884838099203(21)36, (01)00884838099203(21)4, (01)00884838099203(21)42, (01)00884838099203(21)32, (01)00884838099203(21)60, (01)00884838099203(21)55, (01)00884838099203(21)44, (01)00884838099203(21)13, (01)00884838099203(21)34, (01)00884838099203(21)39, (01)00884838099203(21)59, (01)00884838099203(21)41, (01)00884838099203(21)38, (01)00884838099203(21)48, (01)00884838099203(21)31, (01)00884838099203(21)23, (01)00884838099203(21)26, (01)00884838099203(21)10, (01)00884838099203(21)30, (01)00884838099203(21)50, (01)00884838099203(21)29, (01)00884838099203(21)15, (01)00884838099203(21)57, (01)00884838099203(21)27, (01)00884838099203(21)2, (01)00884838099203(21)47, (01)00884838099203(21)16, (01)00884838099203(21)35, (01)00884838099203(21)5, (01)00884838099203(21)22, (01)00884838099203(21)58, (01)00884838099203(21)28, (01)00884838099203(21)46, (01)00884838099203(21)53, (01)00884838099203(21)8, (01)00884838099203(21)56, (01)00884838099203(21)43, (01)00884838099203(21)1, (01)00884838099203(21)11) 3, 7, 6, 9, 19, 45, 51, 36, 4, 42, 32, 60, 55, 44, 13, 34, 39, 59, 41, 38, 48, 31, 23, 26, 10, 30, 50, 29, 15, 57, 27, 2, 47, 16, 35, 5, 22, 58, 28, 46, 53, 8, 56, 43, 1, 11.
2. Azurion 3 M15, Model Number (UDI) Serial Numbers: 722064 ((01)00884838085282(21)238, (01)00884838085282(21)258, (01)00884838085282(21)260, (01)00884838085282(21)261, (01)00884838085282(21)314, (01)00884838085282(21)291, (01)00884838085282(21)270, (01)00884838085282(21)267, (01)00884838085282(21)335, (01)00884838085282(21)297, (01)00884838085282(21)479, (01)00884838085282(21)268, (01)00884838085282(21)359, (01)00884838085282(21)375, (01)00884838085282(21)256, (01)00884838085282(21)290, (01)00884838085282(21)425, (01)00884838085282(21)303, (01)00884838085282(21)254, (01)00884838085282(21)257, (01)00884838085282(21)280, (01)00884838085282(21)373, (01)00884838085282(21)417, (01)00884838085282(21)277, (01)00884838085282(21)467, (01)00884838085282(21)461, (01)00884838085282(21)421, (01)00884838085282(21)345, (01)00884838085282(21)271, (01)00884838085282(21)246, (01)00884838085282(21)282, (01)00884838085282(21)354, (01)00884838085282(21)374, (01)00884838085282(21)274, (01)00884838085282(21)451, (01)00884838085282(21)276, (01)00884838085282(21)200, (01)00884838085282(21)249, (01)00884838085282(21)230, (01)00884838085282(21)231, (01)00884838085282(21)233, (01)00884838085282(21)287, (01)00884838085282(21)288, (01)00884838085282(21)289, (01)00884838085282(21)448, (01)00884838085282(21)323, (01)00884838085282(21)278, (01)00884838085282(21)94, (01)00884838085282(21)422, (01)00884838085282(21)439, (01)00884838085282(21)446, (01)00884838085282(21)292, (01)00884838085282(21)293, (01)00884838085282(21)321, (01)00884838085282(21)379, (01)00884838085282(21)462, (01)00884838085282(21)342, (01)00884838085282(21)391, (01)00884838085282(21)441, (01)00884838085282(21)320, (01)00884838085282(21)443, (01)00884838085282(21)311, (01)00884838085282(21)442, (01)00884838085282(21)485, (01)00884838085282(21)384, (01)00884838085282(21)228, (01)00884838085282(21)309, (01)00884838085282(21)251, (01)00884838085282(21)480, (01)00884838085282(21)387, (01)00884838085282(21)269, (01)00884838085282(21)454, (01)00884838085282(21)262, (01)00884838085282(21)491, (01)00884838085282(21)476, (01)00884838085282(21)337, (01)00884838085282(21)275, (01)00884838085282(21)468, (01)00884838085282(21)412, (01)00884838085282(21)414, (01)00884838085282(21)405, (01)00884838085282(21)396, (01)00884838085282(21)247, (01)00884838085282(21)406, (01)00884838085282(21)394, (01)00884838085282(21)432, (01)00884838085282(21)504, (01)00884838085282(21)457, (01)00884838085282(21)426, (01)00884838085282(21)469, (01)00884838085282(21)340, (01)00884838085282(21)393, (01)00884838085282(21)343, (01)00884838085282(21)322, (01)00884838085282(21)444, (01)00884838085282(21)395, (01)00884838085282(21)385, (01)00884838085282(21)472, (01)00884838085282(21)333, (01)00884838085282(21)236, (01)00884838085282(21)401, (01)00884838085282(21)234, (01)00884838085282(21)407, (01)00884838085282(21)498, (01)00884838085282(21)413, (01)00884838085282(21)440, (01)00884838085282(21)463, (01)00884838085282(21)330, (01)00884838085282(21)473, (01)00884838085282(21)501, (01)00884838085282(21)308, (01)00884838085282(21)474, (01)00884838085282(21)344, (01)00884838085282(21)389, (01)00884838085282(21)324, (01)00884838085282(21)265, (01)00884838085282(21)427, (01)00884838085282(21)341, (01)00884838085282(21)388, (01)00884838085282(21)177, (01)00884838085282(21)272, (01)00884838085282(21)250, (01)00884838085282(21)452, (01)00884838085282(21)447, (01)00884838085282(21)302, (01)00884838085282(21)317, (01)00884838085282(21)378, (01)00884838085282(21)312, (01)00884838085282(21)403, (01)00884838085282(21)386, (01)00884838085282(21)408, (01)00884838085282(21)381, (01)00884838085282(21)404, (01)00884838085282(21)285, (01)00884838085282(21)449, (01)00884838085282(21)338, (01)00884838085282(21)215, (01)00884838085282(21)164, (01)00884838085282(21)434, (01)00884838085282(21)144, (01)00884838085282(21)327, (01)00884838085282(21)295, (01)00884838085282(21)294, (01)00884838085282(21)319, (01)00884838085282(21)241, (01)00884838085282(21)284, (01)00884838085282(21)419, (01)00884838085282(21)429, (01)00884838085282(21)296, (01)00884838085282(21)283, (01)00884838085282(21)357, (01)00884838085282(21)306, (01)00884838085282(21)437, (01)00884838085282(21)315, (01)00884838085282(21)255, (01)00884838085282(21)189, (01)00884838085282(21)313, (01)00884838085282(21)245) 238, 258, 260, 261, 314, 291, 270, 267, 335, 297, 479, 268, 359, 375, 256, 290, 425, 303, 254, 257, 280, 373, 417, 277, 467, 461, 421, 345, 271, 246, 282, 354, 374, 274, 451, 276, 200, 249, 230, 231, 233, 287, 288, 289, 448, 323, 278, 94, 422, 439, 446, 292, 293, 321, 379, 462, 342, 391, 441, 320, 443, 311, 442, 485, 384, 228, 309, 251, 480, 387, 269, 454, 262, 491, 476, 337, 275, 468, 412, 414, 405, 396, 247, 406, 394, 432, 504, 457, 426, 469, 340, 393, 343, 322, 444, 395, 385, 472, 333, 236, 401, 234, 407, 498, 413, 440, 463, 330, 473, 501, 308, 474, 344, 389, 324, 265, 427, 341, 388, 177, 272, 250, 452, 447, 302, 317, 378, 312, 403, 386, 408, 381, 404, 285, 449, 338, 215, 164, 434, 144, 327, 295, 294, 319, 241, 284, 419, 429, 296, 283, 357, 306, 437, 315, 255, 189, 313, 245. 722222 ((01)00884838099210(21)12, (01)00884838099210(21)15, (01)00884838099210(21)19, (01)00884838099210(21)23, (01)00884838099210(21)17, (01)00884838099210(21)55, (01)00884838099210(21)44, (01)00884838099210(21)37, (01)00884838099210(21)27, (01)00884838099210(21)56, (01)00884838099210(21)35, (01)00884838099210(21)88, (01)00884838099210(21)89, (01)00884838099210(21)102, (01)00884838099210(21)118, (01)00884838099210(21)48, (01)00884838099210(21)53, (01)00884838099210(21)4, (01)00884838099210(21)80, (01)00884838099210(21)83, (01)00884838099210(21)47, (01)00884838099210(21)21, (01)00884838099210(21)64, (01)00884838099210(21)22, (01)00884838099210(21)18, (01)00884838099210(21)24, (01)00884838099210(21)11, (01)00884838099210(21)72, (01)00884838099210(21)9, (01)00884838099210(21)66, (01)00884838099210(21)65, (01)00884838099210(21)82, (01)00884838099210(21)51, (01)00884838099210(21)5, (01)00884838099210(21)76, (01)00884838099210(21)106, (01)00884838099210(21)92, (01)00884838099210(21)16, (01)00884838099210(21)105, (01)00884838099210(21)54, (01)00884838099210(21)14, (01)00884838099210(21)38, (01)00884838099210(21)10, (01)00884838099210(21)90, (01)00884838099210(21)13, (01)00884838099210(21)107, (01)00884838099210(21)73, (01)00884838099210(21)87) 12, 15, 19, 23, 17, 55, 44, 37, 27, 56, 35, 88, 89, 102, 118, 48, 53, 4, 80, 83, 47, 21, 64, 22, 18, 24, 11, 72, 9, 66, 65, 82, 51, 5, 76, 106, 92, 16, 105, 54, 14, 38, 10, 90, 13, 107, 73, 87.
3. Azurion 5 M12, Model Number (UDI) Serial Numbers: 722227 ((01)00884838099227(21)23, (01)00884838099227(21)54, (01)00884838099227(21)56, (01)00884838099227(21)31, (01)00884838099227(21)81, (01)00884838099227(21)24, (01)00884838099227(21)37, (01)00884838099227(21)53, (01)00884838099227(21)84, (01)00884838099227(21)38, (01)00884838099227(21)28, (01)00884838099227(21)29, (01)00884838099227(21)25, (01)00884838099227(21)59, (01)00884838099227(21)19, (01)00884838099227(21)61, (01)00884838099227(21)55, (01)00884838099227(21)76, (01)00884838099227(21)73, (01)00884838099227(21)41, (01)00884838099227(21)85, (01)00884838099227(21)20, (01)00884838099227(21)40, (01)00884838099227(21)5, (01)00884838099227(21)6, (01)00884838099227(21)7, (01)00884838099227(21)36, (01)00884838099227(21)43, (01)00884838099227(21)18, (01)00884838099227(21)14, (01)00884838099227(21)15, (01)00884838099227(21)17, (01)00884838099227(21)21, (01)00884838099227(21)64, (01)00884838099227(21)49, (01)00884838099227(21)91, (01)00884838099227(21)22, (01)00884838099227(21)69, (01)00884838099227(21)72, (01)00884838099227(21)77, (01)00884838099227(21)2, (01)00884838099227(21)1, (01)00884838099227(21)12, (01)00884838099227(21)16, (01)00884838099227(21)67, (01)00884838099227(21)52, (01)00884838099227(21)26, (01)00884838099227(21)44, (01)00884838099227(21)58, (01)00884838099227(21)48, (01)00884838099227(21)47, (01)00884838099227(21)46, (01)00884838099227(21)57, (01)00884838099227(21)3, (01)00884838099227(21)32, (01)00884838099227(21)79, (01)00884838099227(21)60, (01)00884838099227(21)74, (01)00884838099227(21)9, (01)00884838099227(21)8, (01)00884838099227(21)35, (01)00884838099227(21)11, (01)00884838099227(21)13, (01)00884838099227(21)4) 23, 54, 56, 31, 81, 24, 37, 53, 84, 38, 28, 29, 25, 59, 19, 61, 55, 76, 73, 41, 85, 20, 40, 5, 6, 7, 36, 43, 18, 14, 15, 17, 21, 64, 49, 91, 22, 69, 72, 77, 2, 1, 12, 16, 67, 52, 26, 44, 58, 48, 47, 46, 57, 3, 32, 79, 60, 74, 9, 8, 35, 11, 13, 4.
4. Azurion 5 M20, Model Number (UDI) Serial Numbers: 722228 ((01)00884838099234(21)49, (01)00884838099234(21)10, (01)00884838099234(21)36, (01)00884838099234(21)64, (01)00884838099234(21)45, (01)00884838099234(21)18, (01)00884838099234(21)180, (01)00884838099234(21)61, (01)00884838099234(21)76, (01)00884838099234(21)17, (01)00884838099234(21)38, (01)00884838099234(21)32, (01)00884838099234(21)23, (01)00884838099234(21)25, (01)00884838099234(21)62, (01)00884838099234(21)48, (01)00884838099234(21)24, (01)00884838099234(21)26, (01)00884838099234(21)29, (01)00884838099234(21)6, (01)00884838099234(21)37, (01)00884838099234(21)3, (01)00884838099234(21)44, (01)00884838099234(21)102, (01)00884838099234(21)103, (01)00884838099234(21)16, (01)00884838099234(21)71, (01)00884838099234(21)21, (01)00884838099234(21)22, (01)00884838099234(21)75, (01)00884838099234(21)78, (01)00884838099234(21)60, (01)00884838099234(21)4, (01)00884838099234(21)9, (01)00884838099234(21)95, (01)00884838099234(21)84, (01)00884838099234(21)39, (01)00884838099234(21)52, (01)00884838099234(21)53, (01)00884838099234(21)31, (01)00884838099234(21)90, (01)00884838099234(21)46, (01)00884838099234(21)34, (01)00884838099234(21)70, (01)00884838099234(21)13, (01)00884838099234(21)12, (01)00884838099234(21)56, (01)00884838099234(21)57, (01)00884838099234(21)86, (01)00884838099234(21)63, (01)00884838099234(21)35, (01)00884838099234(21)14, (01)00884838099234(21)7, (01)00884838099234(21)94, (01)00884838099234(21)93, (01)00884838099234(21)85, (01)00884838099234(21)5, (01)00884838099234(21)43, (01)00884838099234(21)66, (01)00884838099234(21)33, (01)00884838099234(21)82, (01)00884838099234(21)164, (01)00884838099234(21)81, (01)00884838099234(21)161, (01)00884838099234(21)65, (01)00884838099234(21)30, (01)00884838099234(21)59, (01)00884838099234(21)11, (01)00884838099234(21)28, (01)00884838099234(21)68, (01)00884838099234(21)55, (01)00884838099234(21)27) 49, 10, 36, 64, 45, 18, 180, 61, 76, 17, 38, 32, 23, 25, 62, 48, 24, 26, 29, 6, 37, 3, 44, 102, 103, 16, 71, 21, 22, 75, 78, 60, 4, 9, 95, 84, 39, 52, 53, 31, 90, 46, 34, 70, 13, 12, 56, 57, 86, 63, 35, 14, 7, 94, 93, 85, 5, 43, 66, 33, 82, 164, 81, 161, 65, 30, 59, 11, 28, 68, 55, 27.
5. Azurion 7 B12, Model Number (UDI) Serial Numbers: 722067 ((01)00884838085350(21)172, (01)00884838085350(21)250, (01)00884838085350(21)225, (01)00884838085350(21)221, (01)00884838085350(21)278, (01)00884838085350(21)277, (01)00884838085350(21)244, (01)00884838085350(21)234, (01)00884838085350(21)188, (01)00884838085350(21)233, (01)00884838085350(21)255, (01)00884838085350(21)256, (01)00884838085350(21)270, (01)00884838085350(21)238, (01)00884838085350(21)157, (01)00884838085350(21)251, (01)00884838085350(21)237, (01)00884838085350(21)222, (01)00884838085350(21)262, (01)00884838085350(21)209, (01)00884838085350(21)260, (01)00884838085350(21)191, (01)00884838085350(21)95, (01)00884838085350(21)274, (01)00884838085350(21)218, (01)00884838085350(21)231, (01)00884838085350(21)275, (01)00884838085350(21)30, (01)00884838085350(21)206, (01)00884838085350(21)265, (01)00884838085350(21)268, (01)00884838085350(21)254, (01)00884838085350(21)266, (01)00884838085350(21)97, (01)00884838085350(21)235, (01)00884838085350(21)272, (01)00884838085350(21)273, (01)00884838085350(21)276, (01)00884838085350(21)232, (01)00884838085350(21)271, (01)00884838085350(21)202, (01)00884838085350(21)243, (01)00884838085350(21)207, (01)00884838085350(21)264, (01)00884838085350(21)84, (01)00884838085350(21)239, (01)00884838085350(21)245, (01)00884838085350(21)252, (01)00884838085350(21)195, (01)00884838085350(21)196, (01)00884838085350(21)197, (01)00884838085350(21)171, (01)00884838085350(21)200, (01)00884838085350(21)193, (01)00884838085350(21)229, (01)00884838085350(21)227, (01)00884838085350(21)189, (01)00884838085350(21)198, (01)00884838085350(21)199, (01)00884838085350(21)194) 172, 250, 225, 221, 278, 277, 244, 234, 188, 233, 255, 256, 270, 238, 157, 251, 237, 222, 262, 209, 260, 191, 95, 274, 218, 231, 275, 30, 206, 265, 268, 254, 266, 97, 235, 272, 273, 276, 232, 271, 202, 243, 207, 264, 84, 239, 245, 252, 195, 196, 197, 171, 200, 193, 229, 227, 189, 198, 199, 194. 722225 ((01)00884838099265(21)46, (01)00884838099265(21)28, (01)00884838099265(21)41, (01)00884838099265(21)112, (01)00884838099265(21)29, (01)00884838099265(21)71, (01)00884838099265(21)37, (01)00884838099265(21)53, (01)00884838099265(21)55, (01)00884838099265(21)84, (01)00884838099265(21)111, (01)00884838099265(21)56, (01)00884838099265(21)128, (01)00884838099265(21)124, (01)00884838099265(21)140, (01)00884838099265(21)43, (01)00884838099265(21)8, (01)00884838099265(21)5, (01)00884838099265(21)38, (01)00884838099265(21)14, (01)00884838099265(21)243, (01)00884838099265(21)4, (01)00884838099265(21)30, (01)00884838099265(21)79, (01)00884838099265(21)103, (01)00884838099265(21)107, (01)00884838099265(21)83, (01)00884838099265(21)87, (01)00884838099265(21)123, (01)00884838099265(21)7, (01)00884838099265(21)17, (01)00884838099265(21)31, (01)00884838099265(21)20, (01)00884838099265(21)32, (01)00884838099265(21)155, (01)00884838099265(21)121, (01)00884838099265(21)117, (01)00884838099265(21)82, (01)00884838099265(21)85, (01)00884838099265(21)131, (01)00884838099265(21)132, (01)00884838099265(21)40, (01)00884838099265(21)106, (01)00884838099265(21)44, (01)00884838099265(21)13, (01)00884838099265(21)54, (01)00884838099265(21)69, (01)00884838099265(21)64, (01)00884838099265(21)18, (01)00884838099265(21)19, (01)00884838099265(21)77, (01)00884838099265(21)76, (01)00884838099265(21)10, (01)00884838099265(21)150, (01)00884838099265(21)119, (01)00884838099265(21)88, (01)00884838099265(21)75, (01)00884838099265(21)39, (01)00884838099265(21)127, (01)00884838099265(21)60, (01)00884838099265(21)86, (01)00884838099265(21)61, (01)00884838099265(21)115, (01)00884838099265(21)35, (01)00884838099265(21)59, (01)00884838099265(21)36, (01)00884838099265(21)154, (01)00884838099265(21)15, (01)00884838099265(21)62, (01)00884838099265(21)57, (01)00884838099265(21)94, (01)00884838099265(21)95, (01)00884838099265(21)139, (01)00884838099265(21)11, (01)00884838099265(21)120, (01)00884838099265(21)58, (01)00884838099265(21)34, (01)00884838099265(21)21, (01)00884838099265(21)67, (01)00884838099265(21)89, (01)00884838099265(21)90, (01)00884838099265(21)92, (01)00884838099265(21)91, (01)00884838099265(21)73, (01)00884838099265(21)108, (01)00884838099265(21)143, (01)00884838099265(21)138, (01)00884838099265(21)100, (01)00884838099265(21)16, (01)00884838099265(21)27, (01)00884838099265(21)80, (01)00884838099265(21)42, (01)00884838099265(21)105, (01)00884838099265(21)74, (01)00884838099265(21)63, (01)00884838099265(21)70, (01)00884838099265(21)12, (01)00884838099265(21)9) 46, 28, 41, 112, 29, 71, 37, 53, 55, 84, 111, 56, 128, 124, 140, 43, 8, 5, 38, 14, 243, 4, 30, 79, 103, 107, 83, 87, 123, 7, 17, 31, 20, 32, 155, 121, 117, 82, 85, 131, 132, 40, 106, 44, 13, 54, 69, 64, 18, 19, 77, 76, 10, 150, 119, 88, 75, 39, 127, 60, 86, 61, 115, 35, 59, 36, 154, 15, 62, 57, 94, 95, 139, 11, 120, 58, 34, 21, 67, 89, 90, 92, 91, 73, 108, 143, 138, 100, 16, 27, 80, 42, 105, 74, 63, 70, 12, 9.
6. Azurion 7 B20, Model Number (UDI) Serial Numbers: 722068 ((01)00884838085367(21)547, (01)00884838085367(21)467, (01)00884838085367(21)494, (01)00884838085367(21)564, (01)00884838085367(21)418, (01)00884838085367(21)440, (01)00884838085367(21)42, (01)00884838085367(21)324, (01)00884838085367(21)407, (01)00884838085367(21)505, (01)00884838085367(21)443, (01)00884838085367(21)322, (01)00884838085367(21)373, (01)00884838085367(21)442, (01)00884838085367(21)524, (01)00884838085367(21)400, (01)00884838085367(21)510, (01)00884838085367(21)446, (01)00884838085367(21)528, (01)00884838085367(21)472, (01)00884838085367(21)457, (01)00884838085367(21)470, (01)00884838085367(21)520, (01)00884838085367(21)490, (01)00884838085367(21)251, (01)00884838085367(21)469, (01)00884838085367(21)431, (01)00884838085367(21)374, (01)00884838085367(21)511, (01)00884838085367(21)477, (01)00884838085367(21)413, (01)00884838085367(21)485, (01)00884838085367(21)459, (01)00884838085367(21)506, (01)00884838085367(21)473, (01)00884838085367(21)417, (01)00884838085367(21)314, (01)00884838085367(21)468, (01)00884838085367(21)422, (01)00884838085367(21)454, (01)00884838085367(21)410, (01)00884838085367(21)466, (01)00884838085367(21)495, (01)00884838085367(21)445, (01)00884838085367(21)537, (01)00884838085367(21)438, (01)00884838085367(21)536, (01)00884838085367(21)313, (01)00884838085367(21)481, (01)00884838085367(21)577, (01)00884838085367(21)480, (01)00884838085367(21)364, (01)00884838085367(21)491, (01)00884838085367(21)423, (01)00884838085367(21)509, (01)00884838085367(21)376, (01)00884838085367(21)539, (01)00884838085367(21)508, (01)00884838085367(21)301, (01)00884838085367(21)550, (01)00884838085367(21)460, (01)00884838085367(21)478, (01)00884838085367(21)522, (01)00884838085367(21)568, (01)00884838085367(21)448, (01)00884838085367(21)544, (01)00884838085367(21)428, (01)00884838085367(21)432, (01)00884838085367(21)447, (01)00884838085367(21)339, (01)00884838085367(21)555, (01)00884838085367(21)570, (01)00884838085367(21)387, (01)00884838085367(21)500, (01)00884838085367(21)347, (01)00884838085367(21)560, (01)00884838085367(21)518, (01)00884838085367(21)521, (01)00884838085367(21)529, (01)00884838085367(21)453, (01)00884838085367(21)427, (01)00884838085367(21)534, (01)00884838085367(21)497, (01)00884838085367(21)527, (01)00884838085367(21)168, (01)00884838085367(21)396, (01)00884838085367(21)515, (01)00884838085367(21)486, (01)00884838085367(21)543, (01)00884838085367(21)483, (01)00884838085367(21)484, (01)00884838085367(21)365, (01)00884838085367(21)471, (01)00884838085367(21)540, (01)00884838085367(21)584, (01)00884838085367(21)553, (01)00884838085367(21)240, (01)00884838085367(21)589, (01)00884838085367(21)526, (01)00884838085367(21)559, (01)00884838085367(21)450, (01)00884838085367(21)556, (01)00884838085367(21)597, (01)00884838085367(21)516, (01)00884838085367(21)563, (01)00884838085367(21)590, (01)00884838085367(21)548, (01)00884838085367(21)420, (01)00884838085367(21)397, (01)00884838085367(21)475, (01)00884838085367(21)180, (01)00884838085367(21)414, (01)00884838085367(21)562, (01)00884838085367(21)519, (01)00884838085367(21)234, (01)00884838085367(21)533, (01)00884838085367(21)398, (01)00884838085367(21)549, (01)00884838085367(21)416, (01)00884838085367(21)531, (01)00884838085367(21)426, (01)00884838085367(21)326, (01)00884838085367(21)331, (01)00884838085367(21)465, (01)00884838085367(21)281, (01)00884838085367(21)501, (01)00884838085367(21)393, (01)00884838085367(21)503, (01)00884838085367(21)439, (01)00884838085367(21)263, (01)00884838085367(21)391, (01)00884838085367(21)379, (01)00884838085367(21)464, (01)00884838085367(21)502, (01)00884838085367(21)498, (01)00884838085367(21)430, (01)00884838085367(21)436, (01)00884838085367(21)409, (01)00884838085367(21)507, (01)00884838085367(21)451) 547, 467, 494, 564, 418, 440, 42, 324, 407, 505, 443, 322, 373, 442, 524, 400, 510, 446, 528, 472, 457, 470, 520, 490, 251, 469, 431, 374, 511, 477, 413, 485, 459, 506, 473, 417, 314, 468, 422, 454, 410, 466, 495, 445, 537, 438, 536, 313, 481, 577, 480, 364, 491, 423, 509, 376, 539, 508, 301, 550, 460, 478, 522, 568, 448, 544, 428, 432, 447, 339, 555, 570, 387, 500, 347, 560, 518, 521, 529, 453, 427, 534, 497, 527, 168, 396, 515, 486, 543, 483, 484, 365, 471, 540, 584, 553, 240, 589, 526, 559, 450, 556, 597, 516, 563, 590, 548, 420, 397, 475, 180, 414, 562, 519, 234, 533, 398, 549, 416, 531, 426, 326, 331, 465, 281, 501, 393, 503, 439, 263, 391, 379, 464, 502, 498, 430, 436, 409, 507, 451. 722226 ((01)00884838099272(21)122, (01)00884838099272(21)123, (01)00884838099272(21)125, (01)00884838099272(21)126, (01)00884838099272(21)120, (01)00884838099272(21)124, (01)00884838099272(21)127, (01)00884838099272(21)153, (01)00884838099272(21)209, (01)00884838099272(21)112, (01)00884838099272(21)113, (01)00884838099272(21)121, (01)00884838099272(21)239, (01)00884838099272(21)105, (01)00884838099272(21)114, (01)00884838099272(21)108, (01)00884838099272(21)197, (01)00884838099272(21)182, (01)00884838099272(21)100, (01)00884838099272(21)107, (01)00884838099272(21)144, (01)00884838099272(21)98, (01)00884838099272(21)117, (01)00884838099272(21)244, (01)00884838099272(21)78, (01)00884838099272(21)11, (01)00884838099272(21)129, (01)00884838099272(21)220, (01)00884838099272(21)159, (01)00884838099272(21)175, (01)00884838099272(21)143, (01)00884838099272(21)173, (01)00884838099272(21)147, (01)00884838099272(21)204, (01)00884838099272(21)419, (01)00884838099272(21)174, (01)00884838099272(21)181, (01)00884838099272(21)205, (01)00884838099272(21)59, (01)00884838099272(21)75, (01)00884838099272(21)19, (01)00884838099272(21)214, (01)00884838099272(21)213, (01)00884838099272(21)215, (01)00884838099272(21)296, (01)00884838099272(21)149, (01)00884838099272(21)141, (01)00884838099272(21)21, (01)00884838099272(21)225, (01)00884838099272(21)201, (01)00884838099272(21)202, (01)00884838099272(21)184, (01)00884838099272(21)243, (01)00884838099272(21)178, (01)00884838099272(21)179, (01)00884838099272(21)250, (01)00884838099272(21)368, (01)00884838099272(21)176, (01)00884838099272(21)60, (01)00884838099272(21)232, (01)00884838099272(21)331, (01)00884838099272(21)282, (01)00884838099272(21)222, (01)00884838099272(21)221, (01)00884838099272(21)61, (01)00884838099272(21)306, (01)00884838099272(21)83, (01)00884838099272(21)199, (01)00884838099272(21)203, (01)00884838099272(21)119, (01)00884838099272(21)230, (01)00884838099272(21)327, (01)00884838099272(21)298, (01)00884838099272(21)10, (01)00884838099272(21)227, (01)00884838099272(21)259, (01)00884838099272(21)140, (01)00884838099272(21)142, (01)00884838099272(21)263, (01)00884838099272(21)321, (01)00884838099272(21)17, (01)00884838099272(21)247, (01)00884838099272(21)99, (01)00884838099272(21)36, (01)00884838099272(21)88, (01)00884838099272(21)245, (01)00884838099272(21)5, (01)00884838099272(21)188, (01)00884838099272(21)14, (01)00884838099272(21)45, (01)00884838099272(21)97, (01)00884838099272(21)228, (01)00884838099272(21)66, (01)00884838099272(21)366, (01)00884838099272(21)18, (01)00884838099272(21)217, (01)00884838099272(21)47, (01)00884838099272(21)148, (01)00884838099272(21)238, (01)00884838099272(21)283, (01)00884838099272(21)284, (01)00884838099272(21)84, (01)00884838099272(21)118, (01)00884838099272(21)224, (01)00884838099272(21)233, (01)00884838099272(21)1, (01)00884838099272(21)320, (01)00884838099272(21)307, (01)00884838099272(21)235, (01)00884838099272(21)25, (01)00884838099272(21)275, (01)00884838099272(21)87, (01)00884838099272(21)251, (01)00884838099272(21)29, (01)00884838099272(21)131, (01)00884838099272(21)22, (01)00884838099272(21)50, (01)00884838099272(21)168, (01)00884838099272(21)166, (01)00884838099272(21)16, (01)00884838099272(21)207, (01)00884838099272(21)54, (01)00884838099272(21)82, (01)00884838099272(21)330, (01)00884838099272(21)277, (01)00884838099272(21)278, (01)00884838099272(21)152, (01)00884838099272(21)305, (01)00884838099272(21)231, (01)00884838099272(21)6, (01)00884838099272(21)24, (01)00884838099272(21)86, (01)00884838099272(21)7, (01)00884838099272(21)52, (01)00884838099272(21)53, (01)00884838099272(21)63, (01)00884838099272(21)26, (01)00884838099272(21)171, (01)00884838099272(21)69, (01)00884838099272(21)74, (01)00884838099272(21)286, (01)00884838099272(21)42, (01)00884838099272(21)115, (01)00884838099272(21)229, (01)00884838099272(21)285, (01)00884838099272(21)156, (01)00884838099272(21)161, (01)00884838099272(21)162, (01)00884838099272(21)73, (01)00884838099272(21)46, (01)00884838099272(21)312, (01)00884838099272(21)160, (01)00884838099272(21)216, (01)00884838099272(21)95, (01)00884838099272(21)70, (01)00884838099272(21)128, (01)00884838099272(21)287, (01)00884838099272(21)40, (01)00884838099272(21)89, (01)00884838099272(21)265, (01)00884838099272(21)237, (01)00884838099272(21)106, (01)00884838099272(21)304, (01)00884838099272(21)137, (01)00884838099272(21)110, (01)00884838099272(21)257, (01)00884838099272(21)68, (01)00884838099272(21)234, (01)00884838099272(21)64, (01)00884838099272(21)192, (01)00884838099272(21)193, (01)00884838099272(21)32, (01)00884838099272(21)292, (01)00884838099272(21)206, (01)00884838099272(21)49, (01)00884838099272(21)258, (01)00884838099272(21)172, (01)00884838099272(21)38, (01)00884838099272(21)33, (01)00884838099272(21)30, (01)00884838099272(21)92, (01)00884838099272(21)269, (01)00884838099272(21)240, (01)00884838099272(21)163, (01)00884838099272(21)62, (01)00884838099272(21)210, (01)00884838099272(21)187, (01)00884838099272(21)264, (01)00884838099272(21)90, (01)00884838099272(21)154, (01)00884838099272(21)273, (01)00884838099272(21)186, (01)00884838099272(21)34, (01)00884838099272(21)44, (01)00884838099272(21)65, (01)00884838099272(21)71, (01)00884838099272(21)196, (01)00884838099272(21)91, (01)00884838099272(21)35, (01)00884838099272(21)31, (01)00884838099272(21)80, (01)00884838099272(21)41, (01)00884838099272(21)56, (01)00884838099272(21)155, (01)00884838099272(21)96, (01)00884838099272(21)28, (01)00884838099272(21)324, (01)00884838099272(21)157, (01)00884838099272(21)79, (01)00884838099272(21)77, (01)00884838099272(21)236, (01)00884838099272(21)146, (01)00884838099272(21)165, (01)00884838099272(21)241) 122, 123, 125, 126, 120, 124, 127, 153, 209, 112, 113, 121, 239, 105, 114, 108, 197, 182, 100, 107, 144, 98, 117, 244, 78, 11, 129, 220, 159, 175, 143, 173, 147, 204, 419, 174, 181, 205, 59, 75, 19, 214, 213, 215, 296, 149, 141, 21, 225, 201, 202, 184, 243, 178, 179, 250, 368, 176, 60, 232, 331, 282, 222, 221, 61, 306, 83, 199, 203, 119, 230, 327, 298, 10, 227, 259, 140, 142, 263, 321, 17, 247, 99, 36, 88, 245, 5, 188, 14, 45, 97, 228, 66, 366, 18, 217, 47, 148, 238, 283, 284, 84, 118, 224, 233, 1, 320, 307, 235, 25, 275, 87, 251, 29, 131, 22, 50, 168, 166, 16, 207, 54, 82, 330, 277, 278, 152, 305, 231, 6, 24, 86, 7, 52, 53, 63, 26, 171, 69, 74, 286, 42, 115, 229, 285, 156, 161, 162, 73, 46, 312, 160, 216, 95, 70, 128, 287, 40, 89, 265, 237, 106, 304, 137, 110, 257, 68, 234, 64, 192, 193, 32, 292, 206, 49, 258, 172, 38, 33, 30, 92, 269, 240, 163, 62, 210, 187, 264, 90, 154, 273, 186, 34, 44, 65, 71, 196, 91, 35, 31, 80, 41, 56, 155, 96, 28, 324, 157, 79, 77, 236, 146, 165, 241.
7. Azurion 7 M12, Model Number (UDI) Serial Numbers: 722078 ((01)00884838085251(21)753, (01)00884838085251(21)840, (01)00884838085251(21)870, (01)00884838085251(21)679, (01)00884838085251(21)829, (01)00884838085251(21)833, (01)00884838085251(21)921, (01)00884838085251(21)700, (01)00884838085251(21)862, (01)00884838085251(21)1007, (01)00884838085251(21)786, (01)00884838085251(21)803, (01)00884838085251(21)890, (01)00884838085251(21)684, (01)00884838085251(21)824, (01)00884838085251(21)902, (01)00884838085251(21)709, (01)00884838085251(21)906, (01)00884838085251(21)788, (01)00884838085251(21)644, (01)00884838085251(21)876, (01)00884838085251(21)844, (01)00884838085251(21)748, (01)00884838085251(21)483, (01)00884838085251(21)797, (01)00884838085251(21)733, (01)00884838085251(21)888, (01)00884838085251(21)923, (01)00884838085251(21)986, (01)00884838085251(21)977, (01)00884838085251(21)738, (01)00884838085251(21)690, (01)00884838085251(21)943, (01)00884838085251(21)898, (01)00884838085251(21)764, (01)00884838085251(21)770, (01)00884838085251(21)771, (01)00884838085251(21)772, (01)00884838085251(21)773, (01)00884838085251(21)439, (01)00884838085251(21)938, (01)00884838085251(21)980, (01)00884838085251(21)981, (01)00884838085251(21)982, (01)00884838085251(21)1003, (01)00884838085251(21)897, (01)00884838085251(21)845, (01)00884838085251(21)960, (01)00884838085251(21)1036, (01)00884838085251(21)967, (01)00884838085251(21)564, (01)00884838085251(21)934, (01)00884838085251(21)835, (01)00884838085251(21)836, (01)00884838085251(21)929, (01)00884838085251(21)975, (01)00884838085251(21)893, (01)00884838085251(21)604, (01)00884838085251(21)849, (01)00884838085251(21)346, (01)00884838085251(21)347, (01)00884838085251(21)958, (01)00884838085251(21)751, (01)00884838085251(21)881, (01)00884838085251(21)846, (01)00884838085251(21)800, (01)00884838085251(21)838, (01)00884838085251(21)762, (01)00884838085251(21)727, (01)00884838085251(21)726, (01)00884838085251(21)1039, (01)00884838085251(21)847, (01)00884838085251(21)959, (01)00884838085251(21)962, (01)00884838085251(21)865, (01)00884838085251(21)769, (01)00884838085251(21)1031, (01)00884838085251(21)792, (01)00884838085251(21)698, (01)00884838085251(21)702, (01)00884838085251(21)1025, (01)00884838085251(21)722, (01)00884838085251(21)912, (01)00884838085251(21)699, (01)00884838085251(21)815, (01)00884838085251(21)743, (01)00884838085251(21)757, (01)00884838085251(21)703, (01)00884838085251(21)818, (01)00884838085251(21)779, (01)00884838085251(21)855, (01)00884838085251(21)918, (01)00884838085251(21)919, (01)00884838085251(21)976, (01)00884838085251(21)817, (01)00884838085251(21)795, (01)00884838085251(21)631, (01)00884838085251(21)632, (01)00884838085251(21)796, (01)00884838085251(21)718, (01)00884838085251(21)462, (01)00884838085251(21)464, (01)00884838085251(21)891, (01)00884838085251(21)768, (01)00884838085251(21)944, (01)00884838085251(21)920, (01)00884838085251(21)761, (01)00884838085251(21)1037, (01)00884838085251(21)799, (01)00884838085251(21)839, (01)00884838085251(21)957, (01)00884838085251(21)854, (01)00884838085251(21)785, (01)00884838085251(21)789, (01)00884838085251(21)877, (01)00884838085251(21)801, (01)00884838085251(21)802, (01)00884838085251(21)878, (01)00884838085251(21)952, (01)00884838085251(21)988, (01)00884838085251(21)563, (01)00884838085251(21)953, (01)00884838085251(21)806, (01)00884838085251(21)950, (01)00884838085251(21)1005, (01)00884838085251(21)979, (01)00884838085251(21)831, (01)00884838085251(21)841, (01)00884838085251(21)842, (01)00884838085251(21)905, (01)00884838085251(21)843, (01)00884838085251(21)830, (01)00884838085251(21)673, (01)00884838085251(21)674, (01)00884838085251(21)675, (01)00884838085251(21)676, (01)00884838085251(21)609, (01)00884838085251(21)749, (01)00884838085251(21)935, (01)00884838085251(21)968, (01)00884838085251(21)750, (01)00884838085251(21)579, (01)00884838085251(21)926, (01)00884838085251(21)899, (01)00884838085251(21)322, (01)00884838085251(21)978, (01)00884838085251(21)670, (01)00884838085251(21)909, (01)00884838085251(21)1033, (01)00884838085251(21)1035, (01)00884838085251(21)1009, (01)00884838085251(21)1047, (01)00884838085251(21)984, (01)00884838085251(21)989, (01)00884838085251(21)907, (01)00884838085251(21)1004, (01)00884838085251(21)1048, (01)00884838085251(21)1059, (01)00884838085251(21)875, (01)00884838085251(21)1026, (01)00884838085251(21)1045, (01)00884838085251(21)942, (01)00884838085251(21)945, (01)00884838085251(21)946, (01)00884838085251(21)947, (01)00884838085251(21)948, (01)00884838085251(21)940, (01)00884838085251(21)1052, (01)00884838085251(21)971, (01)00884838085251(21)1062, (01)00884838085251(21)1063, (01)00884838085251(21)701, (01)00884838085251(21)993, (01)00884838085251(21)1046, (01)00884838085251(21)970, (01)00884838085251(21)985, (01)00884838085251(21)798, (01)00884838085251(21)1011, (01)00884838085251(21)1000, (01)00884838085251(21)964, (01)00884838085251(21)965, (01)00884838085251(21)1043, (01)00884838085251(21)735, (01)00884838085251(21)856, (01)00884838085251(21)774, (01)00884838085251(21)864, (01)00884838085251(21)837, (01)00884838085251(21)784, (01)00884838085251(21)991, (01)00884838085251(21)466, (01)00884838085251(21)778, (01)00884838085251(21)879, (01)00884838085251(21)896, (01)00884838085251(21)782, (01)00884838085251(21)961, (01)00884838085251(21)857, (01)00884838085251(21)954, (01)00884838085251(21)816, (01)00884838085251(21)892, (01)00884838085251(21)853, (01)00884838085251(21)972, (01)00884838085251(21)1032, (01)00884838085251(21)1051, (01)00884838085251(21)850, (01)00884838085251(21)575, (01)00884838085251(21)939, (01)00884838085251(21)869, (01)00884838085251(21)874, (01)00884838085251(21)871, (01)00884838085251(21)1010, (01)00884838085251(21)1013, (01)00884838085251(21)917, (01)00884838085251(21)913, (01)00884838085251(21)908, (01)00884838085251(21)901) 753, 840, 870, 679, 829, 833, 921, 700, 862, 1007, 786, 803, 890, 684, 824, 902, 709, 906, 788, 644, 876, 844, 748, 483, 797, 733, 888, 923, 986, 977, 738, 690, 943, 898, 764, 770, 771, 772, 773, 439, 938, 980, 981, 982, 1003, 897, 845, 960, 1036, 967, 564, 934, 835, 836, 929, 975, 893, 604, 849, 346, 347, 958, 751, 881, 846, 800, 838, 762, 727, 726, 1039, 847, 959, 962, 865, 769, 1031, 792, 698, 702, 1025, 722, 912, 699, 815, 743, 757, 703, 818, 779, 855, 918, 919, 976, 817, 795, 631, 632, 796, 718, 462, 464, 891, 768, 944, 920, 761, 1037, 799, 839, 957, 854, 785, 789, 877, 801, 802, 878, 952, 988, 563, 953, 806, 950, 1005, 979, 831, 841, 842, 905, 843, 830, 673, 674, 675, 676, 609, 749, 935, 968, 750, 579, 926, 899, 322, 978, 670, 909, 1033, 1035, 1009, 1047, 984, 989, 907, 1004, 1048, 1059, 875, 1026, 1045, 942, 945, 946, 947, 948, 940, 1052, 971, 1062, 1063, 701, 993, 1046, 970, 985, 798, 1011, 1000, 964, 965, 1043, 735, 856, 774, 864, 837, 784, 991, 466, 778, 879, 896, 782, 961, 857, 954, 816, 892, 853, 972, 1032, 1051, 850, 575, 939, 869, 874, 871, 1010, 1013, 917, 913, 908, 901. 722223 ((01)00884838099241(21)189, (01)00884838099241(21)47, (01)00884838099241(21)77, (01)00884838099241(21)171, (01)00884838099241(21)126, (01)00884838099241(21)36, (01)00884838099241(21)252, (01)00884838099241(21)125, (01)00884838099241(21)207, (01)00884838099241(21)205, (01)00884838099241(21)192, (01)00884838099241(21)197, (01)00884838099241(21)177, (01)00884838099241(21)37, (01)00884838099241(21)340, (01)00884838099241(21)282, (01)00884838099241(21)214, (01)00884838099241(21)44, (01)00884838099241(21)210, (01)00884838099241(21)206, (01)00884838099241(21)186, (01)00884838099241(21)187, (01)00884838099241(21)87, (01)00884838099241(21)69, (01)00884838099241(21)256, (01)00884838099241(21)67, (01)00884838099241(21)131, (01)00884838099241(21)318, (01)00884838099241(21)229, (01)00884838099241(21)262, (01)00884838099241(21)308, (01)00884838099241(21)124, (01)00884838099241(21)198, (01)00884838099241(21)433, (01)00884838099241(21)188, (01)00884838099241(21)52, (01)00884838099241(21)127, (01)00884838099241(21)104, (01)00884838099241(21)334, (01)00884838099241(21)96, (01)00884838099241(21)277, (01)00884838099241(21)223, (01)00884838099241(21)161, (01)00884838099241(21)180, (01)00884838099241(21)196, (01)00884838099241(21)130, (01)00884838099241(21)164, (01)00884838099241(21)139, (01)00884838099241(21)89, (01)00884838099241(21)119, (01)00884838099241(21)116, (01)00884838099241(21)154, (01)00884838099241(21)99, (01)00884838099241(21)100, (01)00884838099241(21)136, (01)00884838099241(21)155, (01)00884838099241(21)17, (01)00884838099241(21)18, (01)00884838099241(21)249, (01)00884838099241(21)167, (01)00884838099241(21)227, (01)00884838099241(21)46, (01)00884838099241(21)237, (01)00884838099241(21)348, (01)00884838099241(21)287, (01)00884838099241(21)135, (01)00884838099241(21)132, (01)00884838099241(21)303, (01)00884838099241(21)322, (01)00884838099241(21)10, (01)00884838099241(21)134, (01)00884838099241(21)247, (01)00884838099241(21)7, (01)00884838099241(21)128, (01)00884838099241(21)393, (01)00884838099241(21)434, (01)00884838099241(21)353, (01)00884838099241(21)398, (01)00884838099241(21)213, (01)00884838099241(21)142, (01)00884838099241(21)302, (01)00884838099241(21)137, (01)00884838099241(21)25, (01)00884838099241(21)191, (01)00884838099241(21)106, (01)00884838099241(21)232, (01)00884838099241(21)311, (01)00884838099241(21)113, (01)00884838099241(21)81, (01)00884838099241(21)272, (01)00884838099241(21)140, (01)00884838099241(21)63, (01)00884838099241(21)109, (01)00884838099241(21)316, (01)00884838099241(21)317, (01)00884838099241(21)78, (01)00884838099241(21)202, (01)00884838099241(21)108, (01)00884838099241(21)185, (01)00884838099241(21)760, (01)00884838099241(21)239, (01)00884838099241(21)238, (01)00884838099241(21)299, (01)00884838099241(21)40, (01)00884838099241(21)354, (01)00884838099241(21)280, (01)00884838099241(21)281, (01)00884838099241(21)279, (01)00884838099241(21)306, (01)00884838099241(21)225, (01)00884838099241(21)226, (01)00884838099241(21)173, (01)00884838099241(21)212, (01)00884838099241(21)8, (01)00884838099241(21)9, (01)00884838099241(21)157, (01)00884838099241(21)165, (01)00884838099241(21)344, (01)00884838099241(21)431, (01)00884838099241(21)274, (01)00884838099241(21)217, (01)00884838099241(21)395, (01)00884838099241(21)111, (01)00884838099241(21)35, (01)00884838099241(21)331, (01)00884838099241(21)79, (01)00884838099241(21)65, (01)00884838099241(21)295, (01)00884838099241(21)270, (01)00884838099241(21)468, (01)00884838099241(21)49, (01)00884838099241(21)68, (01)00884838099241(21)465, (01)00884838099241(21)20, (01)00884838099241(21)245, (01)00884838099241(21)246, (01)00884838099241(21)222, (01)00884838099241(21)374, (01)00884838099241(21)61, (01)00884838099241(21)273, (01)00884838099241(21)294, (01)00884838099241(21)4, (01)00884838099241(21)118, (01)00884838099241(21)264, (01)00884838099241(21)64, (01)00884838099241(21)43, (01)00884838099241(21)179, (01)00884838099241(21)215, (01)00884838099241(21)21, (01)00884838099241(21)76, (01)00884838099241(21)257, (01)00884838099241(21)194, (01)00884838099241(21)174, (01)00884838099241(21)48, (01)00884838099241(21)12, (01)00884838099241(21)243, (01)00884838099241(21)283, (01)00884838099241(21)30, (01)00884838099241(21)22, (01)00884838099241(21)169, (01)00884838099241(21)26, (01)00884838099241(21)176, (01)00884838099241(21)91, (01)00884838099241(21)304, (01)00884838099241(21)443, (01)00884838099241(21)115, (01)00884838099241(21)259, (01)00884838099241(21)159, (01)00884838099241(21)160, (01)00884838099241(21)170, (01)00884838099241(21)82, (01)00884838099241(21)199, (01)00884838099241(21)271, (01)00884838099241(21)278, (01)00884838099241(21)219, (01)00884838099241(21)66, (01)00884838099241(21)298, (01)00884838099241(21)285, (01)00884838099241(21)92, (01)00884838099241(21)218, (01)00884838099241(21)255, (01)00884838099241(21)254, (01)00884838099241(21)41, (01)00884838099241(21)72, (01)00884838099241(21)74, (01)00884838099241(21)80, (01)00884838099241(21)291, (01)00884838099241(21)241, (01)00884838099241(21)24, (01)00884838099241(21)110, (01)00884838099241(21)309, (01)00884838099241(21)102, (01)00884838099241(21)315, (01)00884838099241(21)38, (01)00884838099241(21)123, (01)00884838099241(21)250, (01)00884838099241(21)172, (01)00884838099241(21)175, (01)00884838099241(21)305, (01)00884838099241(21)103, (01)00884838099241(21)27, (01)00884838099241(21)267, (01)00884838099241(21)268, (01)00884838099241(21)293, (01)00884838099241(21)321, (01)00884838099241(21)182, (01)00884838099241(21)156, (01)00884838099241(21)158, (01)00884838099241(21)224, (01)00884838099241(21)163, (01)00884838099241(21)389, (01)00884838099241(21)83, (01)00884838099241(21)86, (01)00884838099241(21)33, (01)00884838099241(21)366, (01)00884838099241(21)345, (01)00884838099241(21)94, (01)00884838099241(21)95, (01)00884838099241(21)13, (01)00884838099241(21)428, (01)00884838099241(21)190, (01)00884838099241(21)296, (01)00884838099241(21)297, (01)00884838099241(21)263, (01)00884838099241(21)319, (01)00884838099241(21)320, (01)00884838099241(21)300, (01)00884838099241(21)276, (01)00884838099241(21)292, (01)00884838099241(21)184, (01)00884838099241(21)338, (01)00884838099241(21)107, (01)00884838099241(21)166, (01)00884838099241(21)200, (01)00884838099241(21)327) 189, 47, 77, 171, 126, 36, 252, 125, 207, 205, 192, 197, 177, 37, 340, 282, 214, 44, 210, 206, 186, 187, 87, 69, 256, 67, 131, 318, 229, 262, 308, 124, 198, 433, 188, 52, 127, 104, 334, 96, 277, 223, 161, 180, 196, 130, 164, 139, 89, 119, 116, 154, 99, 100, 136, 155, 17, 18, 249, 167, 227, 46, 237, 348, 287, 135, 132, 303, 322, 10, 134, 247, 7, 128, 393, 434, 353, 398, 213, 142, 302, 137, 25, 191, 106, 232, 311, 113, 81, 272, 140, 63, 109, 316, 317, 78, 202, 108, 185, 760, 239, 238, 299, 40, 354, 280, 281, 279, 306, 225, 226, 173, 212, 8, 9, 157, 165, 344, 431, 274, 217, 395, 111, 35, 331, 79, 65, 295, 270, 468, 49, 68, 465, 20, 245, 246, 222, 374, 61, 273, 294, 4, 118, 264, 64, 43, 179, 215, 21, 76, 257, 194, 174, 48, 12, 243, 283, 30, 22, 169, 26, 176, 91, 304, 443, 115, 259, 159, 160, 170, 82, 199, 271, 278, 219, 66, 298, 285, 92, 218, 255, 254, 41, 72, 74, 80, 291, 241, 24, 110, 309, 102, 315, 38, 123, 250, 172, 175, 305, 103, 27, 267, 268, 293, 321, 182, 156, 158, 224, 163, 389, 83, 86, 33, 366, 345, 94, 95, 13, 428, 190, 296, 297, 263, 319, 320, 300, 276, 292, 184, 338, 107, 166, 200, 327.
8. Azurion 7 M20, Model Number (UDI) Serial Numbers: 722079 ((01)00884838085268(21)1387, (01)00884838085268(21)703307.(01)00884838085268(21)1155, (01)00884838085268(21)1395, (01)00884838085268(21)703559, (01)00884838085268(21)703391, (01)00884838085268(21)1310, (01)00884838085268(21)1211, (01)00884838085268(21)1213, (01)00884838085268(21)1365, (01)00884838085268(21)1408, (01)00884838085268(21)703540, (01)00884838085268(21)703558, (01)00884838085268(21)703336, (01)00884838085268(21)703493, (01)00884838085268(21)1086, (01)00884838085268(21)703350, (01)00884838085268(21)703395, (01)00884838085268(21)703323, (01)00884838085268(21)703775, (01)00884838085268(21)1216, (01)00884838085268(21)1392, (01)00884838085268(21)703408, (01)00884838085268(21)1363, (01)00884838085268(21)1355, (01)00884838085268(21)1414, (01)00884838085268(21)703387, (01)00884838085268(21)703404, (01)00884838085268(21)528, (01)00884838085268(21)703465, (01)00884838085268(21)1380, (01)00884838085268(21)703861, (01)00884838085268(21)1410, (01)00884838085268(21)703329, (01)00884838085268(21)703413, (01)00884838085268(21)1432, (01)00884838085268(21)1312, (01)00884838085268(21)1424, (01)00884838085268(21)499, (01)00884838085268(21)1230, (01)00884838085268(21)703508, (01)00884838085268(21)1401, (01)00884838085268(21)1370, (01)00884838085268(21)703324, (01)00884838085268(21)703306, (01)00884838085268(21)703615, (01)00884838085268(21)1297, (01)00884838085268(21)1399, (01)00884838085268(21)1276, (01)00884838085268(21)703478, (01)00884838085268(21)919, (01)00884838085268(21)703528, (01)00884838085268(21)703419, (01)00884838085268(21)703435, (01)00884838085268(21)703310, (01)00884838085268(21)703521, (01)00884838085268(21)273, (01)00884838085268(21)703305, (01)00884838085268(21)703495, (01)00884838085268(21)1285, (01)00884838085268(21)904, (01)00884838085268(21)1005, (01)00884838085268(21)703497, (01)00884838085268(21)1400, (01)00884838085268(21)954, (01)00884838085268(21)703332, (01)00884838085268(21)1411, (01)00884838085268(21)1430, (01)00884838085268(21)703426, (01)00884838085268(21)272, (01)00884838085268(21)1289, (01)00884838085268(21)828, (01)00884838085268(21)857, (01)00884838085268(21)703416, (01)00884838085268(21)703327, (01)00884838085268(21)1222, (01)00884838085268(21)1268, (01)00884838085268(21)1063, (01)00884838085268(21)703335, (01)00884838085268(21)1311, (01)00884838085268(21)703396, (01)00884838085268(21)703420, (01)00884838085268(21)703407, (01)00884838085268(21)1219, (01)00884838085268(21)1042, (01)00884838085268(21)703425, (01)00884838085268(21)703532, (01)00884838085268(21)1436, (01)00884838085268(21)703344, (01)00884838085268(21)1358, (01)00884838085268(21)1359, (01)00884838085268(21)1360, (01)00884838085268(21)1391, (01)00884838085268(21)703389, (01)00884838085268(21)703309, (01)00884838085268(21)703383, (01)00884838085268(21)703510, (01)00884838085268(21)1245, (01)00884838085268(21)1369, (01)00884838085268(21)1323, (01)00884838085268(21)1385, (01)00884838085268(21)1386, (01)00884838085268(21)1252, (01)00884838085268(21)703379, (01)00884838085268(21)703402, (01)00884838085268(21)703311, (01)00884838085268(21)1402, (01)00884838085268(21)703418, (01)00884838085268(21)1242, (01)00884838085268(21)703501, (01)00884838085268(21)1271, (01)00884838085268(21)18, (01)00884838085268(21)1423, (01)00884838085268(21)1270, (01)00884838085268(21)1250, (01)00884838085268(21)1200, (01)00884838085268(21)703314, (01)00884838085268(21)1226, (01)00884838085268(21)1236, (01)00884838085268(21)1237, (01)00884838085268(21)1232, (01)00884838085268(21)703450, (01)00884838085268(21)537, (01)00884838085268(21)703317, (01)00884838085268(21)703334, (01)00884838085268(21)703368, (01)00884838085268(21)703562, (01)00884838085268(21)1231, (01)00884838085268(21)703360, (01)00884838085268(21)1431, (01)00884838085268(21)703308, (01)00884838085268(21)703575, (01)00884838085268(21)1316, (01)00884838085268(21)1393, (01)00884838085268(21)703409, (01)00884838085268(21)731, (01)00884838085268(21)703390, (01)00884838085268(21)703584, (01)00884838085268(21)671, (01)00884838085268(21)1381, (01)00884838085268(21)1382, (01)00884838085268(21)703333, (01)00884838085268(21)862, (01)00884838085268(21)1379, (01)00884838085268(21)1235, (01)00884838085268(21)1164, (01)00884838085268(21)703345, (01)00884838085268(21)990, (01)00884838085268(21)1100, (01)00884838085268(21)703486, (01)00884838085268(21)703515, (01)00884838085268(21)703457, (01)00884838085268(21)703455, (01)00884838085268(21)703456, (01)00884838085268(21)1247, (01)00884838085268(21)703388, (01)00884838085268(21)1415, (01)00884838085268(21)1000, (01)00884838085268(21)703356, (01)00884838085268(21)703380, (01)00884838085268(21)703381, (01)00884838085268(21)1039, (01)00884838085268(21)1374, (01)00884838085268(21)703550, (01)00884838085268(21)703685, (01)00884838085268(21)703551, (01)00884838085268(21)704199, (01)00884838085268(21)703614, (01)00884838085268(21)703813, (01)00884838085268(21)704114, (01)00884838085268(21)703565, (01)00884838085268(21)703338, (01)00884838085268(21)703552, (01)00884838085268(21)703613, (01)00884838085268(21)703838, (01)00884838085268(21)703823, (01)00884838085268(21)703355, (01)00884838085268(21)703449, (01)00884838085268(21)703654, (01)00884838085268(21)703699, (01)00884838085268(21)703870, (01)00884838085268(21)703612, (01)00884838085268(21)1248, (01)00884838085268(21)822, (01)00884838085268(21)1241, (01)00884838085268(21)703375, (01)00884838085268(21)947, (01)00884838085268(21)1319, (01)00884838085268(21)703712, (01)00884838085268(21)1397, (01)00884838085268(21)1398, (01)00884838085268(21)703326, (01)00884838085268(21)703655, (01)00884838085268(21)703702, (01)00884838085268(21)703802, (01)00884838085268(21)1389, (01)00884838085268(21)1053, (01)00884838085268(21)703600, (01)00884838085268(21)703769, (01)00884838085268(21)962, (01)00884838085268(21)703374, (01)00884838085268(21)703710, (01)00884838085268(21)704026, (01)00884838085268(21)703592, (01)00884838085268(21)703384, (01)00884838085268(21)703392, (01)00884838085268(21)703753, (01)00884838085268(21)703967, (01)00884838085268(21)704112, (01)00884838085268(21)703992, (01)00884838085268(21)703807, (01)00884838085268(21)703463, (01)00884838085268(21)703529, (01)00884838085268(21)704037, (01)00884838085268(21)1429, (01)00884838085268(21)704209, (01)00884838085268(21)703619, (01)00884838085268(21)703545, (01)00884838085268(21)703358, (01)00884838085268(21)1095, (01)00884838085268(21)703487, (01)00884838085268(21)1239, (01)00884838085268(21)703572, (01)00884838085268(21)703462, (01)00884838085268(21)703855, (01)00884838085268(21)1205, (01)00884838085268(21)1206, (01)00884838085268(21)1214, (01)00884838085268(21)703348, (01)00884838085268(21)703872, (01)00884838085268(21)1338, (01)00884838085268(21)1168, (01)00884838085268(21)703489, (01)00884838085268(21)703959, (01)00884838085268(21)703966, (01)00884838085268(21)1337, (01)00884838085268(21)1182, (01)00884838085268(21)1341, (01)00884838085268(21)760, (01)00884838085268(21)961, (01)00884838085268(21)703477, (01)00884838085268(21)703982, (01)00884838085268(21)703438, (01)00884838085268(21)703783, (01)00884838085268(21)703812, (01)00884838085268(21)703423, (01)00884838085268(21)1264, (01)00884838085268(21)703313, (01)00884838085268(21)703544, (01)00884838085268(21)703393, (01)00884838085268(21)703428, (01)00884838085268(21)703749, (01)00884838085268(21)1372, (01)00884838085268(21)703815, (01)00884838085268(21)703330, (01)00884838085268(21)1347, (01)00884838085268(21)703472, (01)00884838085268(21)1254, (01)00884838085268(21)703470, (01)00884838085268(21)782, (01)00884838085268(21)703965, (01)00884838085268(21)703542, (01)00884838085268(21)703445, (01)00884838085268(21)703526, (01)00884838085268(21)740, (01)00884838085268(21)704085, (01)00884838085268(21)703485, (01)00884838085268(21)703500, (01)00884838085268(21)703935, (01)00884838085268(21)703436, (01)00884838085268(21)703405, (01)00884838085268(21)596, (01)00884838085268(21)703806, (01)00884838085268(21)703804, (01)00884838085268(21)703906, (01)00884838085268(21)703596, (01)00884838085268(21)703937, (01)00884838085268(21)1288, (01)00884838085268(21)703954, (01)00884838085268(21)1183, (01)00884838085268(21)1217, (01)00884838085268(21)1314, (01)00884838085268(21)703556, (01)00884838085268(21)703754, (01)00884838085268(21)722079_703569, (01)00884838085268(21)704081, (01)00884838085268(21)704103, (01)00884838085268(21)704019, (01)00884838085268(21)1257, (01)00884838085268(21)839, (01)00884838085268(21)703424, (01)00884838085268(21)704023, (01)00884838085268(21)748, (01)00884838085268(21)1351, (01)00884838085268(21)703347, (01)00884838085268(21)704030, (01)00884838085268(21)704126, (01)00884838085268(21)704036, (01)00884838085268(21)703690, (01)00884838085268(21)703597, (01)00884838085268(21)703611, (01)00884838085268(21)1080, (01)00884838085268(21)703648, (01)00884838085268(21)703709, (01)00884838085268(21)703728, (01)00884838085268(21)703826, (01)00884838085268(21)630, (01)00884838085268(21)1318, (01)00884838085268(21)703794, (01)00884838085268(21)703607, (01)00884838085268(21)703574, (01)00884838085268(21)703743, (01)00884838085268(21)703632, (01)00884838085268(21)703913, (01)00884838085268(21)703889, (01)00884838085268(21)704054, (01)00884838085268(21)703606, (01)00884838085268(21)703473, (01)00884838085268(21)703346, (01)00884838085268(21)703750, (01)00884838085268(21)703638, (01)00884838085268(21)703834, (01)00884838085268(21)703727, (01)00884838085268(21)703875, (01)00884838085268(21)703482, (01)00884838085268(21)703693, (01)00884838085268(21)703533, (01)00884838085268(21)703560, (01)00884838085268(21)704067, (01)00884838085268(21)1135, (01)00884838085268(21)703602, (01)00884838085268(21)833, (01)00884838085268(21)834, (01)00884838085268(21)703506, (01)00884838085268(21)1223, (01)00884838085268(21)1077, (01)00884838085268(21)703349, (01)00884838085268(21)703841, (01)00884838085268(21)703412, (01)00884838085268(21)1272, (01)00884838085268(21)703973, (01)00884838085268(21)703701, (01)00884838085268(21)703628, (01)00884838085268(21)703948, (01)00884838085268(21)703955, (01)00884838085268(21)703665, (01)00884838085268(21)1367, (01)00884838085268(21)703593, (01)00884838085268(21)703828, (01)00884838085268(21)703641, (01)00884838085268(21)703711, (01)00884838085268(21)704173, (01)00884838085268(21)703651, (01)00884838085268(21)703842, (01)00884838085268(21)704186, (01)00884838085268(21)704074, (01)00884838085268(21)704058, (01)00884838085268(21)703586, (01)00884838085268(21)704029, (01)00884838085268(21)703601, (01)00884838085268(21)703979, (01)00884838085268(21)703777, (01)00884838085268(21)704041, (01)00884838085268(21)703778, (01)00884838085268(21)703670, (01)00884838085268(21)703650, (01)00884838085268(21)703646, (01)00884838085268(21)703846, (01)00884838085268(21)703945, (01)00884838085268(21)704021, (01)00884838085268(21)703817, (01)00884838085268(21)704031, (01)00884838085268(21)703726, (01)00884838085268(21)703618, (01)00884838085268(21)703756, (01)00884838085268(21)703724, (01)00884838085268(21)704120, (01)00884838085268(21)704017, (01)00884838085268(21)703921, (01)00884838085268(21)703816, (01)00884838085268(21)703573, (01)00884838085268(21)703800, (01)00884838085268(21)703997, (01)00884838085268(21)703674, (01)00884838085268(21)703673, (01)00884838085268(21)703591, (01)00884838085268(21)703735, (01)00884838085268(21)703757, (01)00884838085268(21)703751, (01)00884838085268(21)703883, (01)00884838085268(21)704139, (01)00884838085268(21)703912, (01)00884838085268(21)703911, (01)00884838085268(21)703605, (01)00884838085268(21)703969, (01)00884838085268(21)704022, (01)00884838085268(21)704039, (01)00884838085268(21)703471, (01)00884838085268(21)703768, (01)00884838085268(21)703663, (01)00884838085268(21)703844, (01)00884838085268(21)703830, (01)00884838085268(21)703707, (01)00884838085268(21)703734, (01)00884838085268(21)704008, (01)00884838085268(21)703799, (01)00884838085268(21)703576, (01)00884838085268(21)704009, (01)00884838085268(21)704069, (01)00884838085268(21)703840, (01)00884838085268(21)703708, (01)00884838085268(21)703998, (01)00884838085268(21)704040, (01)00884838085268(21)703603, (01)00884838085268(21)703786, (01)00884838085268(21)704131, (01)00884838085268(21)704077, (01)00884838085268(21)703630, (01)00884838085268(21)704066, (01)00884838085268(21)703983, (01)00884838085268(21)703507, (01)00884838085268(21)703645, (01)00884838085268(21)703919, (01)00884838085268(21)704088, (01)00884838085268(21)703616, (01)00884838085268(21)703796, (01)00884838085268(21)703857, (01)00884838085268(21)703858, (01)00884838085268(21)703859, (01)00884838085268(21)703860, (01)00884838085268(21)703863, (01)00884838085268(21)703780, (01)00884838085268(21)703640, (01)00884838085268(21)703642, (01)00884838085268(21)703680, (01)00884838085268(21)703878, (01)00884838085268(21)703901, (01)00884838085268(21)704020, (01)00884838085268(21)703958, (01)00884838085268(21)704091, (01)00884838085268(21)703975, (01)00884838085268(21)703918, (01)00884838085268(21)703688, (01)00884838085268(21)704113, (01)00884838085268(21)704090, (01)00884838085268(21)703571, (01)00884838085268(21)703657, (01)00884838085268(21)703547, (01)00884838085268(21)703771, (01)00884838085268(21)703703, (01)00884838085268(21)703781, (01)00884838085268(21)703923, (01)00884838085268(21)704111, (01)00884838085268(21)703805, (01)00884838085268(21)704124, (01)00884838085268(21)703585, (01)00884838085268(21)703759, (01)00884838085268(21)703639, (01)00884838085268(21)703647, (01)00884838085268(21)704064, (01)00884838085268(21)704050, (01)00884838085268(21)703803, (01)00884838085268(21)703697, (01)00884838085268(21)703891, (01)00884838085268(21)703894, (01)00884838085268(21)703895, (01)00884838085268(21)704007, (01)00884838085268(21)703789, (01)00884838085268(21)704075, (01)00884838085268(21)703827, (01)00884838085268(21)703876, (01)00884838085268(21)703989, (01)00884838085268(21)704046, (01)00884838085268(21)703950, (01)00884838085268(21)703679, (01)00884838085268(21)703924, (01)00884838085268(21)704223, (01)00884838085268(21)703934, (01)00884838085268(21)704216, (01)00884838085268(21)704177, (01)00884838085268(21)703825, (01)00884838085268(21)704044, (01)00884838085268(21)703809, (01)00884838085268(21)703836, (01)00884838085268(21)703932, (01)00884838085268(21)703661, (01)00884838085268(21)703676, (01)00884838085268(21)704132, (01)00884838085268(21)703548, (01)00884838085268(21)704200, (01)00884838085268(21)703666, (01)00884838085268(21)703880, (01)00884838085268(21)703730, (01)00884838085268(21)703837, (01)00884838085268(21)703818, (01)00884838085268(21)704096, (01)00884838085268(21)703850, (01)00884838085268(21)703725, (01)00884838085268(21)704161, (01)00884838085268(21)704187, (01)00884838085268(21)704014, (01)00884838085268(21)704082, (01)00884838085268(21)703888, (01)00884838085268(21)703928, (01)00884838085268(21)703929, (01)00884838085268(21)703930, (01)00884838085268(21)703931, (01)00884838085268(21)703974, (01)00884838085268(21)703881, (01)00884838085268(21)703941, (01)00884838085268(21)703996, (01)00884838085268(21)703917, (01)00884838085268(21)703667, (01)00884838085268(21)703668, (01)00884838085268(21)704153, (01)00884838085268(21)703617, (01)00884838085268(21)704169, (01)00884838085268(21)703625, (01)00884838085268(21)703660, (01)00884838085268(21)703949, (01)00884838085268(21)704121, (01)00884838085268(21)704104, (01)00884838085268(21)704128, (01)00884838085268(21)704116, (01)00884838085268(21)703636, (01)00884838085268(21)703843, (01)00884838085268(21)703691, (01)00884838085268(21)703692, (01)00884838085268(21)703664, (01)00884838085268(21)703555, (01)00884838085268(21)703633, (01)00884838085268(21)1224, (01)00884838085268(21)1225, (01)00884838085268(21)704125, (01)00884838085268(21)703684, (01)00884838085268(21)703681, (01)00884838085268(21)703635, (01)00884838085268(21)703946, (01)00884838085268(21)703539, (01)00884838085268(21)703541, (01)00884838085268(21)703678, (01)00884838085268(21)703893, (01)00884838085268(21)703761, (01)00884838085268(21)703845, (01)00884838085268(21)703385, (01)00884838085268(21)703653, (01)00884838085268(21)703351, (01)00884838085268(21)1255, (01)00884838085268(21)703441, (01)00884838085268(21)703715, (01)00884838085268(21)1427, (01)00884838085268(21)703414, (01)00884838085268(21)1366, (01)00884838085268(21)703505, (01)00884838085268(21)1199, (01)00884838085268(21)703658, (01)00884838085268(21)704135, (01)00884838085268(21)703362, (01)00884838085268(21)1332, (01)00884838085268(21)703873, (01)00884838085268(21)703321, (01)00884838085268(21)704010, (01)00884838085268(21)703369, (01)00884838085268(21)704057, (01)00884838085268(21)703459, (01)00884838085268(21)703944, (01)00884838085268(21)1299, (01)00884838085268(21)1388, (01)00884838085268(21)952, (01)00884838085268(21)703720, (01)00884838085268(21)703985, (01)00884838085268(21)906, (01)00884838085268(21)1349, (01)00884838085268(21)703579, (01)00884838085268(21)703810, (01)00884838085268(21)704172, (01)00884838085268(21)703694, (01)00884838085268(21)704015, (01)00884838085268(21)703675, (01)00884838085268(21)703877, (01)00884838085268(21)703956, (01)00884838085268(21)704084, (01)00884838085268(21)703439, (01)00884838085268(21)704035, (01)00884838085268(21)703814, (01)00884838085268(21)1306, (01)00884838085268(21)703717, (01)00884838085268(21)1136, (01)00884838085268(21)703920, (01)00884838085268(21)703922, (01)00884838085268(21)703925, (01)00884838085268(21)704016, (01)00884838085268(21)703729, (01)00884838085268(21)703479, (01)00884838085268(21)703461, (01)00884838085268(21)1229, (01)00884838085268(21)1426, (01)00884838085268(21)703372, (01)00884838085268(21)703622, (01)00884838085268(21)703538, (01)00884838085268(21)703537, (01)00884838085268(21)703517, (01)00884838085268(21)1418, (01)00884838085268(21)1364, (01)00884838085268(21)703440, (01)00884838085268(21)703524, (01)00884838085268(21)703523, (01)00884838085268(21)703502, (01)00884838085268(21)703376, (01)00884838085268(21)703377, (01)00884838085268(21)703503, (01)00884838085268(21)703829, (01)00884838085268(21)1330, (01)00884838085268(21)703952, (01)00884838085268(21)703957, (01)00884838085268(21)703475, (01)00884838085268(21)703947, (01)00884838085268(21)703890, (01)00884838085268(21)703589, (01)00884838085268(21)703536, (01)00884838085268(21)703849, (01)00884838085268(21)703659, (01)00884838085268(21)704025, (01)00884838085268(21)703926, (01)00884838085268(21)703991, (01)00884838085268(21)703987, (01)00884838085268(21)703672, (01)00884838085268(21)703467, (01)00884838085268(21)703468, (01)00884838085268(21)703386, (01)00884838085268(21)703652, (01)00884838085268(21)703714, (01)00884838085268(21)703481, (01)00884838085268(21)1339, (01)00884838085268(21)703631, (01)00884838085268(21)703549, (01)00884838085268(21)703443, (01)00884838085268(21)703352, (01)00884838085268(21)1091, (01)00884838085268(21)703649, (01)00884838085268(21)703531, (01)00884838085268(21)703762, (01)00884838085268(21)703763, (01)00884838085268(21)703554, (01)00884838085268(21)703444, (01)00884838085268(21)703353, (01)00884838085268(21)703787, (01)00884838085268(21)703431, (01)00884838085268(21)703785, (01)00884838085268(21)703410, (01)00884838085268(21)1069, (01)00884838085268(21)703522, (01)00884838085268(21)703354, (01)00884838085268(21)703590, (01)00884838085268(21)703795, (01)00884838085268(21)703570, (01)00884838085268(21)703454, (01)00884838085268(21)703422, (01)00884838085268(21)703518, (01)00884838085268(21)1404, (01)00884838085268(21)1437, (01)00884838085268(21)703484, (01)00884838085268(21)703325, (01)00884838085268(21)703865, (01)00884838085268(21)703403, (01)00884838085268(21)703519, (01)00884838085268(21)703437, (01)00884838085268(21)703907, (01)00884838085268(21)703451, (01)00884838085268(21)1307, (01)00884838085268(21)703373, (01)00884838085268(21)1079, (01)00884838085268(21)703365, (01)00884838085268(21)1407, (01)00884838085268(21)1302, (01)00884838085268(21)1303, (01)00884838085268(21)1304, (01)00884838085268(21)1383, (01)00884838085268(21)794, (01)00884838085268(21)703357, (01)00884838085268(21)703363, (01)00884838085268(21)703452, (01)00884838085268(21)1145) 1387, 703307, 1155, 1395, 703559, 703391, 1310, 1211, 1213, 1365, 1408, 703540, 703558, 703336, 703493, 1086, 703350, 703395, 703323, 703775, 1216, 1392, 703408, 1363, 1355, 1414, 703387, 703404, 528, 703465, 1380, 703861, 1410, 703329, 703413, 1432, 1312, 1424, 499, 1230, 703508, 1401, 1370, 703324, 703306, 703615, 1297, 1399, 1276, 703478, 919, 703528, 703419, 703435, 703310, 703521, 273, 703305, 703495, 1285, 904, 1005, 703497, 1400, 954, 703332, 1411, 1430, 703426, 272, 1289, 828, 857, 703416, 703327, 1222, 1268, 1063, 703335, 1311, 703396, 703420, 703407, 1219, 1042, 703425, 703532, 1436, 703344, 1358, 1359, 1360, 1391, 703389, 703309, 703383, 703510, 1245, 1369, 1323, 1385, 1386, 1252, 703379, 703402, 703311, 1402, 703418, 1242, 703501, 1271, 18, 1423, 1270, 1250, 1200, 703314, 1226, 1236, 1237, 1232, 703450, 537, 703317, 703334, 703368, 703562, 1231, 703360, 1431, 703308, 703575, 1316, 1393, 703409, 731, 703390, 703584, 671, 1381, 1382, 703333, 862, 1379, 1235, 1164, 703345, 990, 1100, 703486, 703515, 703457, 703455, 703456, 1247, 703388, 1415, 1000, 703356, 703380, 703381, 1039, 1374, 703550, 703685, 703551, 704199, 703614, 703813, 704114, 703565, 703338, 703552, 703613, 703838, 703823, 703355, 703449, 703654, 703699, 703870, 703612, 1248, 822, 1241, 703375, 947, 1319, 703712, 1397, 1398, 703326, 703655, 703702, 703802, 1389, 1053, 703600, 703769, 962, 703374, 703710, 704026, 703592, 703384, 703392, 703753, 703967, 704112, 703992, 703807, 703463, 703529, 704037, 1429, 704209, 703619, 703545, 703358, 1095, 703487, 1239, 703572, 703462, 703855, 1205, 1206, 1214, 703348, 703872, 1338, 1168, 703489, 703959, 703966, 1337, 1182, 1341, 760, 961, 703477, 703982, 703438, 703783, 703812, 703423, 1264, 703313, 703544, 703393, 703428, 703749, 1372, 703815, 703330, 1347, 703472, 1254, 703470, 782, 703965, 703542, 703445, 703526, 740, 704085, 703485, 703500, 703935, 703436, 703405, 596, 703806, 703804, 703906, 703596, 703937, 1288, 703954, 1183, 1217, 1314, 703556, 703754, 722079_703569, 704081, 704103, 704019, 1257, 839, 703424, 704023, 748, 1351, 703347, 704030, 704126, 704036, 703690, 703597, 703611, 1080, 703648, 703709, 703728, 703826, 630, 1318, 703794, 703607, 703574, 703743, 703632, 703913, 703889, 704054, 703606, 703473, 703346, 703750, 703638, 703834, 703727, 703875, 703482, 703693, 703533, 703560, 704067, 1135, 703602, 833, 834, 703506, 1223, 1077, 703349, 703841, 703412, 1272, 703973, 703701, 703628, 703948, 703955, 703665, 1367, 703593, 703828, 703641, 703711, 704173, 703651, 703842, 704186, 704074, 704058, 703586, 704029, 703601, 703979, 703777, 704041, 703778, 703670, 703650, 703646, 703846, 703945, 704021, 703817, 704031, 703726, 703618, 703756, 703724, 704120, 704017, 703921, 703816, 703573, 703800, 703997, 703674, 703673, 703591, 703735, 703757, 703751, 703883, 704139, 703912, 703911, 703605, 703969, 704022, 704039, 703471, 703768, 703663, 703844, 703830, 703707, 703734, 704008, 703799, 703576, 704009, 704069, 703840, 703708, 703998, 704040, 703603, 703786, 704131, 704077, 703630, 704066, 703983, 703507, 703645, 703919, 704088, 703616, 703796, 703857, 703858, 703859, 703860, 703863, 703780, 703640, 703642, 703680, 703878, 703901, 704020, 703958, 704091, 703975, 703918, 703688, 704113, 704090, 703571, 703657, 703547, 703771, 703703, 703781, 703923, 704111, 703805, 704124, 703585, 703759, 703639, 703647, 704064, 704050, 703803, 703697, 703891, 703894, 703895, 704007, 703789, 704075, 703827, 703876, 703989, 704046, 703950, 703679, 703924, 704223, 703934, 704216, 704177, 703825, 704044, 703809, 703836, 703932, 703661, 703676, 704132, 703548, 704200, 703666, 703880, 703730, 703837, 703818, 704096, 703850, 703725, 704161, 704187, 704014, 704082, 703888, 703928, 703929, 703930, 703931, 703974, 703881, 703941, 703996, 703917, 703667, 703668, 704153, 703617, 704169, 703625, 703660, 703949, 704121, 704104, 704128, 704116, 703636, 703843, 703691, 703692, 703664, 703555, 703633, 1224, 1225, 704125, 703684, 703681, 703635, 703946, 703539, 703541, 703678, 703893, 703761, 703845, 703385, 703653, 703351, 1255, 703441, 703715, 1427, 703414, 1366, 703505, 1199, 703658, 704135, 703362, 1332, 703873, 703321, 704010, 703369, 704057, 703459, 703944, 1299, 1388, 952, 703720, 703985, 906, 1349, 703579, 703810, 704172, 703694, 704015, 703675, 703877, 703956, 704084, 703439, 704035, 703814, 1306, 703717, 1136, 703920, 703922, 703925, 704016, 703729, 703479, 703461, 1229, 1426, 703372, 703622, 703538, 703537, 703517, 1418, 1364, 703440, 703524, 703523, 703502, 703376, 703377, 703503, 703829, 1330, 703952, 703957, 703475, 703947, 703890, 703589, 703536, 703849, 703659, 704025, 703926, 703991, 703987, 703672, 703467, 703468, 703386, 703652, 703714, 703481, 1339, 703631, 703549, 703443, 703352, 1091, 703649, 703531, 703762, 703763, 703554, 703444, 703353, 703787, 703431, 703785, 703410, 1069, 703522, 703354, 703590, 703795, 703570, 703454, 703422, 703518, 1404, 1437, 703484, 703325, 703865, 703403, 703519, 703437, 703907, 703451, 1307, 703373, 1079, 703365, 1407, 1302, 1303, 1304, 1383, 794, 703357, 703363, 703452, 1145. 722224 ((01)00884838099258(21)450, (01)00884838099258(21)409, (01)00884838099258(21)261, (01)00884838099258(21)262, (01)00884838099258(21)275, (01)00884838099258(21)277, (01)00884838099258(21)295, (01)00884838099258(21)272, (01)00884838099258(21)630, (01)00884838099258(21)381, (01)00884838099258(21)232, (01)00884838099258(21)260, (01)00884838099258(21)246, (01)00884838099258(21)475, (01)00884838099258(21)526, (01)00884838099258(21)539, (01)00884838099258(21)421, (01)00884838099258(21)288, (01)00884838099258(21)90, (01)00884838099258(21)50, (01)00884838099258(21)239, (01)00884838099258(21)589, (01)00884838099258(21)91, (01)00884838099258(21)43, (01)00884838099258(21)362, (01)00884838099258(21)210, (01)00884838099258(21)287, (01)00884838099258(21)403, (01)00884838099258(21)723, (01)00884838099258(21)226, (01)00884838099258(21)365, (01)00884838099258(21)266, (01)00884838099258(21)512, (01)00884838099258(21)301, (01)00884838099258(21)553, (01)00884838099258(21)514, (01)00884838099258(21)82, (01)00884838099258(21)40, (01)00884838099258(21)369, (01)00884838099258(21)527, (01)00884838099258(21)297, (01)00884838099258(21)110, (01)00884838099258(21)584, (01)00884838099258(21)425, (01)00884838099258(21)611, (01)00884838099258(21)826, (01)00884838099258(21)565, (01)00884838099258(21)566, (01)00884838099258(21)742, (01)00884838099258(21)434, (01)00884838099258(21)164, (01)00884838099258(21)224, (01)00884838099258(21)190, (01)00884838099258(21)153, (01)00884838099258(21)79, (01)00884838099258(21)650, (01)00884838099258(21)609, (01)00884838099258(21)667, (01)00884838099258(21)395, (01)00884838099258(21)182, (01)00884838099258(21)75, (01)00884838099258(21)380, (01)00884838099258(21)286, (01)00884838099258(21)289, (01)00884838099258(21)363, (01)00884838099258(21)580, (01)00884838099258(21)273, (01)00884838099258(21)333, (01)00884838099258(21)306, (01)00884838099258(21)218, (01)00884838099258(21)391, (01)00884838099258(21)216, (01)00884838099258(21)191, (01)00884838099258(21)454, (01)00884838099258(21)867, (01)00884838099258(21)244, (01)00884838099258(21)360, (01)00884838099258(21)229, (01)00884838099258(21)476, (01)00884838099258(21)238, (01)00884838099258(21)42, (01)00884838099258(21)166, (01)00884838099258(21)56, (01)00884838099258(21)105, (01)00884838099258(21)291, (01)00884838099258(21)444, (01)00884838099258(21)640, (01)00884838099258(21)298, (01)00884838099258(21)240, (01)00884838099258(21)876, (01)00884838099258(21)94, (01)00884838099258(21)564, (01)00884838099258(21)33, (01)00884838099258(21)385, (01)00884838099258(21)461, (01)00884838099258(21)41, (01)00884838099258(21)129, (01)00884838099258(21)432, (01)00884838099258(21)439, (01)00884838099258(21)603, (01)00884838099258(21)595, (01)00884838099258(21)401, (01)00884838099258(21)400, (01)00884838099258(21)178, (01)00884838099258(21)494, (01)00884838099258(21)571, (01)00884838099258(21)151, (01)00884838099258(21)657, (01)00884838099258(21)115, (01)00884838099258(21)215, (01)00884838099258(21)674, (01)00884838099258(21)342, (01)00884838099258(21)478, (01)00884838099258(21)516, (01)00884838099258(21)187, (01)00884838099258(21)49, (01)00884838099258(21)154, (01)00884838099258(21)278, (01)00884838099258(21)552, (01)00884838099258(21)189, (01)00884838099258(21)217, (01)00884838099258(21)192, (01)00884838099258(21)108, (01)00884838099258(21)339, (01)00884838099258(21)471, (01)00884838099258(21)103, (01)00884838099258(21)47, (01)00884838099258(21)234, (01)00884838099258(21)637, (01)00884838099258(21)128, (01)00884838099258(21)265, (01)00884838099258(21)119, (01)00884838099258(21)587, (01)00884838099258(21)436, (01)00884838099258(21)438, (01)00884838099258(21)608, (01)00884838099258(21)112, (01)00884838099258(21)422, (01)00884838099258(21)464, (01)00884838099258(21)335, (01)00884838099258(21)841, (01)00884838099258(21)326, (01)00884838099258(21)134, (01)00884838099258(21)453, (01)00884838099258(21)507, (01)00884838099258(21)147, (01)00884838099258(21)148, (01)00884838099258(21)447, (01)00884838099258(21)107, (01)00884838099258(21)417, (01)00884838099258(21)374, (01)00884838099258(21)396, (01)00884838099258(21)482, (01)00884838099258(21)163, (01)00884838099258(21)492, (01)00884838099258(21)233, (01)00884838099258(21)645, (01)00884838099258(21)583, (01)00884838099258(21)662, (01)00884838099258(21)78, (01)00884838099258(21)268, (01)00884838099258(21)267, (01)00884838099258(21)346, (01)00884838099258(21)443, (01)00884838099258(21)252, (01)00884838099258(21)414, (01)00884838099258(21)45, (01)00884838099258(21)308, (01)00884838099258(21)285, (01)00884838099258(21)351, (01)00884838099258(21)300, (01)00884838099258(21)100, (01)00884838099258(21)591, (01)00884838099258(21)140, (01)00884838099258(21)479, (01)00884838099258(21)735, (01)00884838099258(21)594, (01)00884838099258(21)370, (01)00884838099258(21)437, (01)00884838099258(21)124, (01)00884838099258(21)125, (01)00884838099258(21)120, (01)00884838099258(21)183, (01)00884838099258(21)366, (01)00884838099258(21)839, (01)00884838099258(21)39, (01)00884838099258(21)785, (01)00884838099258(21)945, (01)00884838099258(21)88, (01)00884838099258(21)466, (01)00884838099258(21)101, (01)00884838099258(21)99, (01)00884838099258(21)592, (01)00884838099258(21)188, (01)00884838099258(21)106, (01)00884838099258(21)472, (01)00884838099258(21)36, (01)00884838099258(21)165, (01)00884838099258(21)72, (01)00884838099258(21)141, (01)00884838099258(21)627, (01)00884838099258(21)28, (01)00884838099258(21)46, (01)00884838099258(21)777, (01)00884838099258(21)177, (01)00884838099258(21)440, (01)00884838099258(21)364, (01)00884838099258(21)746, (01)00884838099258(21)80, (01)00884838099258(21)574, (01)00884838099258(21)113, (01)00884838099258(21)250, (01)00884838099258(21)194, (01)00884838099258(21)53, (01)00884838099258(21)332, (01)00884838099258(21)517, (01)00884838099258(21)254, (01)00884838099258(21)102, (01)00884838099258(21)290, (01)00884838099258(21)404, (01)00884838099258(21)621, (01)00884838099258(21)331, (01)00884838099258(21)468, (01)00884838099258(21)467, (01)00884838099258(21)465, (01)00884838099258(21)328, (01)00884838099258(21)329, (01)00884838099258(21)139, (01)00884838099258(21)159, (01)00884838099258(21)490, (01)00884838099258(21)593, (01)00884838099258(21)279, (01)00884838099258(21)253, (01)00884838099258(21)307, (01)00884838099258(21)64, (01)00884838099258(21)130, (01)00884838099258(21)116, (01)00884838099258(21)607, (01)00884838099258(21)327, (01)00884838099258(21)897, (01)00884838099258(21)176, (01)00884838099258(21)219, (01)00884838099258(21)582, (01)00884838099258(21)441, (01)00884838099258(21)84, (01)00884838099258(21)174, (01)00884838099258(21)350, (01)00884838099258(21)271, (01)00884838099258(21)76, (01)00884838099258(21)109, (01)00884838099258(21)160, (01)00884838099258(21)34, (01)00884838099258(21)503, (01)00884838099258(21)185, (01)00884838099258(21)220, (01)00884838099258(21)225, (01)00884838099258(21)235, (01)00884838099258(21)714, (01)00884838099258(21)452, (01)00884838099258(21)518, (01)00884838099258(21)523, (01)00884838099258(21)522, (01)00884838099258(21)343, (01)00884838099258(21)1527, (01)00884838099258(21)104, (01)00884838099258(21)196, (01)00884838099258(21)833, (01)00884838099258(21)258, (01)00884838099258(21)330, (01)00884838099258(21)372, (01)00884838099258(21)379, (01)00884838099258(21)585, (01)00884838099258(21)649, (01)00884838099258(21)173, (01)00884838099258(21)352, (01)00884838099258(21)353, (01)00884838099258(21)257, (01)00884838099258(21)181, (01)00884838099258(21)534, (01)00884838099258(21)497, (01)00884838099258(21)578, (01)00884838099258(21)394, (01)00884838099258(21)334, (01)00884838099258(21)63, (01)00884838099258(21)558, (01)00884838099258(21)801, (01)00884838099258(21)937, (01)00884838099258(21)149, (01)00884838099258(21)734, (01)00884838099258(21)201, (01)00884838099258(21)531, (01)00884838099258(21)501, (01)00884838099258(21)546, (01)00884838099258(21)167, (01)00884838099258(21)513, (01)00884838099258(21)429, (01)00884838099258(21)828, (01)00884838099258(21)87, (01)00884838099258(21)294, (01)00884838099258(21)184, (01)00884838099258(21)757, (01)00884838099258(21)720, (01)00884838099258(21)877, (01)00884838099258(21)675, (01)00884838099258(21)672, (01)00884838099258(21)9, (01)00884838099258(21)86, (01)00884838099258(21)791, (01)00884838099258(21)499, (01)00884838099258(21)530, (01)00884838099258(21)916, (01)00884838099258(21)340, (01)00884838099258(21)771, (01)00884838099258(21)906, (01)00884838099258(21)890, (01)00884838099258(21)887, (01)00884838099258(21)361, (01)00884838099258(21)532, (01)00884838099258(21)596, (01)00884838099258(21)57, (01)00884838099258(21)62, (01)00884838099258(21)37, (01)00884838099258(21)600, (01)00884838099258(21)65, (01)00884838099258(21)30, (01)00884838099258(21)127, (01)00884838099258(21)620, (01)00884838099258(21)1, (01)00884838099258(21)456, (01)00884838099258(21)386, (01)00884838099258(21)632, (01)00884838099258(21)256, (01)00884838099258(21)367, (01)00884838099258(21)706, (01)00884838099258(21)707, (01)00884838099258(21)708, (01)00884838099258(21)783, (01)00884838099258(21)604, (01)00884838099258(21)544, (01)00884838099258(21)725, (01)00884838099258(21)24, (01)00884838099258(21)406, (01)00884838099258(21)16, (01)00884838099258(21)13, (01)00884838099258(21)816, (01)00884838099258(21)93, (01)00884838099258(21)171, (01)00884838099258(21)573, (01)00884838099258(21)542, (01)00884838099258(21)48, (01)00884838099258(21)463, (01)00884838099258(21)1051, (01)00884838099258(21)1052, (01)00884838099258(21)658, (01)00884838099258(21)1050, (01)00884838099258(21)1053, (01)00884838099258(21)1054, (01)00884838099258(21)383, (01)00884838099258(21)779, (01)00884838099258(21)891, (01)00884838099258(21)305, (01)00884838099258(21)506, (01)00884838099258(21)311, (01)00884838099258(21)98, (01)00884838099258(21)302, (01)00884838099258(21)77, (01)00884838099258(21)145, (01)00884838099258(21)81, (01)00884838099258(21)519, (01)00884838099258(21)368, (01)00884838099258(21)563, (01)00884838099258(21)727, (01)00884838099258(21)639, (01)00884838099258(21)8, (01)00884838099258(21)111, (01)00884838099258(21)259, (01)00884838099258(21)248, (01)00884838099258(21)926, (01)00884838099258(21)249, (01)00884838099258(21)264, (01)00884838099258(21)496, (01)00884838099258(21)545, (01)00884838099258(21)803, (01)00884838099258(21)729, (01)00884838099258(21)768, (01)00884838099258(21)520, (01)00884838099258(21)423, (01)00884838099258(21)193, (01)00884838099258(21)31, (01)00884838099258(21)358, (01)00884838099258(21)299, (01)00884838099258(21)489, (01)00884838099258(21)378, (01)00884838099258(21)208, (01)00884838099258(21)407, (01)00884838099258(21)537, (01)00884838099258(21)175, (01)00884838099258(21)752, (01)00884838099258(21)505, (01)00884838099258(21)555, (01)00884838099258(21)7, (01)00884838099258(21)133, (01)00884838099258(21)349, (01)00884838099258(21)186, (01)00884838099258(21)60, (01)00884838099258(21)19, (01)00884838099258(21)6, (01)00884838099258(21)551, (01)00884838099258(21)460, (01)00884838099258(21)200, (01)00884838099258(21)880, (01)00884838099258(21)410, (01)00884838099258(21)459, (01)00884838099258(21)412, (01)00884838099258(21)20, (01)00884838099258(21)1035, (01)00884838099258(21)508, (01)00884838099258(21)804, (01)00884838099258(21)399, (01)00884838099258(21)211, (01)00884838099258(21)227, (01)00884838099258(21)337, (01)00884838099258(21)795, (01)00884838099258(21)66, (01)00884838099258(21)67, (01)00884838099258(21)223, (01)00884838099258(21)198, (01)00884838099258(21)848, (01)00884838099258(21)634, (01)00884838099258(21)416, (01)00884838099258(21)32, (01)00884838099258(21)85, (01)00884838099258(21)872, (01)00884838099258(21)985, (01)00884838099258(21)59, (01)00884838099258(21)898, (01)00884838099258(21)371, (01)00884838099258(21)156, (01)00884838099258(21)347, (01)00884838099258(21)1017, (01)00884838099258(21)991, (01)00884838099258(21)569, (01)00884838099258(21)533, (01)00884838099258(21)270, (01)00884838099258(21)29, (01)00884838099258(21)52, (01)00884838099258(21)71, (01)00884838099258(21)424, (01)00884838099258(21)671, (01)00884838099258(21)310, (01)00884838099258(21)209, (01)00884838099258(21)629, (01)00884838099258(21)631, (01)00884838099258(21)11, (01)00884838099258(21)304, (01)00884838099258(21)303, (01)00884838099258(21)132, (01)00884838099258(21)653, (01)00884838099258(21)660, (01)00884838099258(21)972, (01)00884838099258(21)74, (01)00884838099258(21)737, (01)00884838099258(21)68, (01)00884838099258(21)938, (01)00884838099258(21)528, (01)00884838099258(21)529, (01)00884838099258(21)293, (01)00884838099258(21)541, (01)00884838099258(21)205, (01)00884838099258(21)221, (01)00884838099258(21)509, (01)00884838099258(21)387, (01)00884838099258(21)597, (01)00884838099258(21)822, (01)00884838099258(21)243, (01)00884838099258(21)207, (01)00884838099258(21)180, (01)00884838099258(21)664, (01)00884838099258(21)155, (01)00884838099258(21)655, (01)00884838099258(21)23, (01)00884838099258(21)774, (01)00884838099258(21)510, (01)00884838099258(21)610, (01)00884838099258(21)375, (01)00884838099258(21)122, (01)00884838099258(21)237, (01)00884838099258(21)758, (01)00884838099258(21)97, (01)00884838099258(21)203, (01)00884838099258(21)821, (01)00884838099258(21)162, (01)00884838099258(21)348, (01)00884838099258(21)838, (01)00884838099258(21)393, (01)00884838099258(21)344, (01)00884838099258(21)474, (01)00884838099258(21)2137, (01)00884838099258(21)498, (01)00884838099258(21)919, (01)00884838099258(21)126, (01)00884838099258(21)345, (01)00884838099258(21)117, (01)00884838099258(21)241, (01)00884838099258(21)92, (01)00884838099258(21)576, (01)00884838099258(21)51, (01)00884838099258(21)918, (01)00884838099258(21)536, (01)00884838099258(21)535, (01)00884838099258(21)213, (01)00884838099258(21)749, (01)00884838099258(21)984, (01)00884838099258(21)863, (01)00884838099258(21)712, (01)00884838099258(21)549, (01)00884838099258(21)70, (01)00884838099258(21)405, (01)00884838099258(21)382, (01)00884838099258(21)855, (01)00884838099258(21)853, (01)00884838099258(21)724, (01)00884838099258(21)810, (01)00884838099258(21)688, (01)00884838099258(21)700, (01)00884838099258(21)701, (01)00884838099258(21)743, (01)00884838099258(21)751, (01)00884838099258(21)705, (01)00884838099258(21)698, (01)00884838099258(21)703, (01)00884838099258(21)704, (01)00884838099258(21)477, (01)00884838099258(21)556, (01)00884838099258(21)721, (01)00884838099258(21)677, (01)00884838099258(21)646, (01)00884838099258(21)168, (01)00884838099258(21)384, (01)00884838099258(21)61, (01)00884838099258(21)392, (01)00884838099258(21)568, (01)00884838099258(21)676, (01)00884838099258(21)562, (01)00884838099258(21)487, (01)00884838099258(21)96, (01)00884838099258(21)142, (01)00884838099258(21)397, (01)00884838099258(21)137, (01)00884838099258(21)473, (01)00884838099258(21)794, (01)00884838099258(21)197, (01)00884838099258(21)143, (01)00884838099258(21)692, (01)00884838099258(21)557, (01)00884838099258(21)764, (01)00884838099258(21)83, (01)00884838099258(21)157, (01)00884838099258(21)170, (01)00884838099258(21)718, (01)00884838099258(21)731, (01)00884838099258(21)118, (01)00884838099258(21)212, (01)00884838099258(21)144, (01)00884838099258(21)538) 450, 409, 261, 262, 275, 277, 295, 272, 630, 381, 232, 260, 246, 475, 526, 539, 421, 288, 90, 50, 239, 589, 91, 43, 362, 210, 287, 403, 723, 226, 365, 266, 512, 301, 553, 514, 82, 40, 369, 527, 297, 110, 584, 425, 611, 826, 565, 566, 742, 434, 164, 224, 190, 153, 79, 650, 609, 667, 395, 182, 75, 380, 286, 289, 363, 580, 273, 333, 306, 218, 391, 216, 191, 454, 867, 244, 360, 229, 476, 238, 42, 166, 56, 105, 291, 444, 640, 298, 240, 876, 94, 564, 33, 385, 461, 41, 129, 432, 439, 603, 595, 401, 400, 178, 494, 571, 151, 657, 115, 215, 674, 342, 478, 516, 187, 49, 154, 278, 552, 189, 217, 192, 108, 339, 471, 103, 47, 234, 637, 128, 265, 119, 587, 436, 438, 608, 112, 422, 464, 335, 841, 326, 134, 453, 507, 147, 148, 447, 107, 417, 374, 396, 482, 163, 492, 233, 645, 583, 662, 78, 268, 267, 346, 443, 252, 414, 45, 308, 285, 351, 300, 100, 591, 140, 479, 735, 594, 370, 437, 124, 125, 120, 183, 366, 839, 39, 785, 945, 88, 466, 101, 99, 592, 188, 106, 472, 36, 165, 72, 141, 627, 28, 46, 777, 177, 440, 364, 746, 80, 574, 113, 250, 194, 53, 332, 517, 254, 102, 290, 404, 621, 331, 468, 467, 465, 328, 329, 139, 159, 490, 593, 279, 253, 307, 64, 130, 116, 607, 327, 897, 176, 219, 582, 441, 84, 174, 350, 271, 76, 109, 160, 34, 503, 185, 220, 225, 235, 714, 452, 518, 523, 522, 343, 1527, 104, 196, 833, 258, 330, 372, 379, 585, 649, 173, 352, 353, 257, 181, 534, 497, 578, 394, 334, 63, 558, 801, 937, 149, 734, 201, 531, 501, 546, 167, 513, 429, 828, 87, 294, 184, 757, 720, 877, 675, 672, 9, 86, 791, 499, 530, 916, 340, 771, 906, 890, 887, 361, 532, 596, 57, 62, 37, 600, 65, 30, 127, 620, 1, 456, 386, 632, 256, 367, 706, 707, 708, 783, 604, 544, 725, 24, 406, 16, 13, 816, 93, 171, 573, 542, 48, 463, 1051, 1052, 658, 1050, 1053, 1054, 383, 779, 891, 305, 506, 311, 98, 302, 77, 145, 81, 519, 368, 563, 727, 639, 8, 111, 259, 248, 926, 249, 264, 496, 545, 803, 729, 768, 520, 423, 193, 31, 358, 299, 489, 378, 208, 407, 537, 175, 752, 505, 555, 7, 133, 349, 186, 60, 19, 6, 551, 460, 200, 880, 410, 459, 412, 20, 1035, 508, 804, 399, 211, 227, 337, 795, 66, 67, 223, 198, 848, 634, 416, 32, 85, 872, 985, 59, 898, 371, 156, 347, 1017, 991, 569, 533, 270, 29, 52, 71, 424, 671, 310, 209, 629, 631, 11, 304, 303, 132, 653, 660, 972, 74, 737, 68, 938, 528, 529, 293, 541, 205, 221, 509, 387, 597, 822, 243, 207, 180, 664, 155, 655, 23, 774, 510, 610, 375, 122, 237, 758, 97, 203, 821, 162, 348, 838, 393, 344, 474, 2137, 498, 919, 126, 345, 117, 241, 92, 576, 51, 918, 536, 535, 213, 749, 984, 863, 712, 549, 70, 405, 382, 855, 853, 724, 810, 688, 700, 701, 743, 751, 705, 698, 703, 704, 477, 556, 721, 677, 646, 168, 384, 61, 392, 568, 676, 562, 487, 96, 142, 397, 137, 473, 794, 197, 143, 692, 557, 764, 83, 157, 170, 718, 731, 118, 212, 144, 538.
|
| FEI Number |
3042175844
|
Recalling Firm/ Manufacturer |
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Veenpluis 6
Best
Netherlands
|
Manufacturer Reason for Recall | When a power supply failure occurs during a start or restart, the Power Distribution Unit does not provide the required direct current power to essential system components. As a result, the system does not successfully complete the (re-)start, imaging functionality is unavailable, and the system remains unavailable for clinical use. |
FDA Determined Cause 2 | Process control |
| Action | Philips notified consignees on about 10/15/2026 via "URGENT Medical Device Correction" letter. Consignees were informed that they may continue to use the affected systems in accordance with their intended use, subject to the following actions: Ensure the notice is distributed to all relevant users and retained with the device documentation until corrective actions are completed, post the notification near affected devices for ease of reference, and if affected product has been transferred to another organization, forward the notification to that organization and inform Philips through local representatives. They were also instructed to keep the notification with the documentation of the system until Philips corrects each system, if the issue is experienced report it to the local Philips representative, and complete and return the provided response form.
Philips will replace the existing Power Distribution Unit assembly with a redesigned assembly that includes power-supply redundancy.
With the redesigned assembly, a failure of a single power supply during system start-up may still result in the system not starting correctly. However, upon performing a cold restart of the system, the redundant power supply will automatically take over the function of the failed power supply, allowing the system to start successfully and restoring system functionality. During a subsequent preventive maintenance visit, the affected assembly will be replaced to restore full redundancy.
Philips currently expects the redesigned Power Distribution Unit to be available by Q2 2028. Each local Philips representative will contact consignees to schedule a visit for implementation of the correction once it becomes available. |
| Quantity in Commerce | 2,632 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia, Brazil, Brunei Darussal, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, French Polynesia, Germany, Ghana, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kosovo, Kuwait, Lao, Latvia, Lebanon, Lithuania, Malaysia, Mexico, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Puerto Rico, Reunion, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Syria, Taiwan, Thailand, Tunisia, Turrkiye, Turkmenistan, Ukraine, United Kingdom, United Arab Emirates, Uzbekistan, Vietnam, Yemen. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|