| Date Initiated by Firm | November 20, 2025 |
| Date Posted | November 06, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0389-2027 |
| Recall Event ID |
100092 |
| PMA Number | P200031 |
| Product Classification |
Normothermic machine perfusion system for the preservation of donor livers prior to transplantation - Product Code QQK
|
| Product | OCS Liver System for organ transport
OCS Liver Perfusion Set
Model/Catalog Number: 3200
Product Description: The recalled product is a standard off the shelf syringe. It is used as part of solution delivery cassette (SDS Cassette) to be used with the perfusion module of OCS Liver System. OCS Liver Perfusion Sets within the serial number range between 4470-3-18076 and 4470-3-20343 may contain SDS cassettes within the impacted lot range of 4459-6-737116 to 4459-6-739965. Each Liver Perfusion Set contains three SDS cassettes.
The BD syringe is a component of SDS Cassette PN 100004459, which is an accessory of the Liver Perfusion Set. |
| Code Information |
OCS Liver Perfusion Set Serial Numbers from 4470-3-18076 to 4470-3-20343;
|
| FEI Number |
3003152463
|
Recalling Firm/ Manufacturer |
TransMedics, Inc.
200 Minuteman Rd Ste 302
Andover
MA
01810-1046
|
| For Additional Information Contact | Christy Foreman 978-8098795 |
Manufacturer Reason for Recall | Faulty Solution Delivery System (SDS) Cassettes within the device set were assembled with syringes that
were manufactured with an atypical lubricant, which results in sub-par performance and has the potential to cause leaks. |
FDA Determined Cause 2 | Component change control |
| Action | On or about November 20, 2025, "OCS Liver Solution Delivery (SDS) Cassettes" letters were sent to customers:
Actions to be taken:
TransMedics requests that you discard any unused, affected, SDS Cassettes.
Actions Taken by TransMedics
We plan to ship replacement units to you to replace the affected product.
If you have any questions, please contact myself at the number below, your local support representative, or
TransMedics Customer Service by phone at +1 (978) 552-0999 (US), +31(0) 20-7084561 (EU) or email (recall@transmedics.com). |
| Quantity in Commerce | 1179 (Liver Perfusion Sets) |
| Distribution | Worldwide - US Nationwide distribution in the state of NE and the country of Italy. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| PMA Database | PMAs with Product Code = QQK
|