| Date Initiated by Firm | November 20, 2025 |
| Date Posted | November 06, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0391-2027 |
| Recall Event ID |
100092 |
| PMA Number | P200031 |
| Product Classification |
Normothermic machine perfusion system for the preservation of donor livers prior to transplantation - Product Code QQK
|
| Product | OCS Liver System for organ transport
OCS Solution Delivery Cassettes
Model/Catalog Number: 1467
Product Description: OCS Solution Delivery Cassettes
The recalled product is a standard off the shelf syringe. It is used as part of solution delivery cassette (SDS Cassette) to be used with the perfusion module of OCS Liver System. Each OCS Solution Delivery Cassettes contains three SDS cassettes.
The BD syringe is a component of the SDS Cassette PN 100004459 |
| Code Information |
Lot Code: Lot: 4460-4-737189;
Expiration Date: 2028-10-08;
UDI: +B00944605/$$74460-4-737189/14D20281008/Q3L;
Lot: 4460-4-738008;
Expiration Date: 2028-11-07;
UDI: +B00944605/$$74460-4-738008/14D20281107/Q3C;
Lot: 4460-4-738009;
Expiration Date: 2028-11-07;
UDI: +B00944605/$$74460-4-738009/14D20281107/Q3D;
Lot: 4460-4-739728;
Expiration Date: 2029-01-14;
UDI: +B00944605/$$74460-4-739728/14D20290114/Q3K;
|
| FEI Number |
3003152463
|
Recalling Firm/ Manufacturer |
TransMedics, Inc.
200 Minuteman Rd Ste 302
Andover
MA
01810-1046
|
| For Additional Information Contact | Christy Foreman 978-8098795 |
Manufacturer Reason for Recall | Faulty Solution Delivery System (SDS) Cassettes within the device set were assembled with syringes that
were manufactured with an atypical lubricant, which results in sub-par performance and has the potential to cause leaks. |
FDA Determined Cause 2 | Component change control |
| Action | On or about November 20, 2025, "OCS Liver Solution Delivery (SDS) Cassettes" letters were sent to customers:
Actions to be taken:
TransMedics requests that you discard any unused, affected, SDS Cassettes.
Actions Taken by TransMedics
We plan to ship replacement units to you to replace the affected product.
If you have any questions, please contact myself at the number below, your local support representative, or
TransMedics Customer Service by phone at +1 (978) 552-0999 (US), +31(0) 20-7084561 (EU) or email (recall@transmedics.com). |
| Quantity in Commerce | 55 OCS Solution Delivery Cassettes |
| Distribution | Worldwide - US Nationwide distribution in the state of NE and the country of Italy. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| PMA Database | PMAs with Product Code = QQK
|