 Class 2 Device Recall Presource Convenience Kitsee related information
Date Initiated by FirmOctober 01, 2026
Date PostedNovember 13, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0451-2027
Recall Event ID 100062
Product Classification Epidural anesthesia kit - Product Code OGE
ProductPresource Convenience kits containing recalled WelMed Extension Tubing Sets LUMBAR EPIDURAL KIT LG 34A1256D SINGLE SHOT EPIDURAL TRAY 34-2703D 34-4623B 34-4882E
Code Information 34A1256D UDI-DI 10195594660239 Kit lot 517607 34-2703D UDI-DI 10197106222379 Kit lots 243328 297284 390599 475326 643185 665520 751245 34-4623B UDI-DI 10197106494820 Kit lots 330874 345915 443088 514878 531968 595253 597882 760845 34-4882E UDI-DI 10197106174364 Kit lots 171937 250913 601741 703682 751247 759293
FEI Number 3001236905
Recalling Firm/
Manufacturer
Cardinal Health 200, LLC
3651 Birchwood Dr
Waukegan IL 60085-8337
For Additional Information ContactJanna Parks
847-887-1114
Manufacturer Reason
for Recall
Kits include Extension Tubing Sets recalled by supplier WelMed. Affected tubing sets may leak.
FDA Determined
Cause 2
Component design/selection
ActionFirm began notifying customers on October 1, 2026. Prior to use, users should visually inspect the tubing connection area, including bonded junctions and connection points, for any visible irregularities, gaps, deformation, incomplete bonding, leakage, or other abnormalities consistent with the examples shown above. If a defect is observed, or if leakage or connection difficulties are encountered during setup or priming, the product should not be used for patient care and discarded. Customers should report affected product through their normal complaint reporting process and retain the product for evaluation if requested. Direct consignees were instructed to notify retail customers of the market action.
Quantity in Commerce1,155,430
DistributionWorldwide distribution - US Nationwide and the country of Canada.
Total Product Life CycleTPLC Device Report