| | Class 2 Device Recall Presource Convenience Kit |  |
| Date Initiated by Firm | October 01, 2026 |
| Date Posted | November 13, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0453-2027 |
| Recall Event ID |
100062 |
| Product Classification |
Cardiac catheterization kit - Product Code OES
|
| Product | Presource Convenience kits containing recalled WelMed Extension Tubing Sets
INTERVENTIONAL PK
SAN11BICH8
SAN11BICHM
SAN11BICHN |
| Code Information |
SAN11BICH8
UDI-DI 10198956239623
Kit lot 645007
SAN11BICHM
UDI-DI 10888439904475
Kit lots 239409
252244
SAN11BICHN
UDI-DI 10197106640029
Kit lots 64523
279504
315006
327529
347210
406855
433230
443005
470724
497502
514680
528968
552592
577383
601044
621522
672712
694830
717947
744553
|
| FEI Number |
3001236905
|
Recalling Firm/ Manufacturer |
Cardinal Health 200, LLC
3651 Birchwood Dr
Waukegan
IL
60085-8337
|
| For Additional Information Contact | Janna Parks 847-887-1114 |
Manufacturer Reason for Recall | Kits include Extension Tubing Sets recalled by supplier WelMed. Affected tubing sets may leak. |
FDA Determined Cause 2 | Component design/selection |
| Action | Firm began notifying customers on October 1, 2026.
Prior to use, users should visually inspect the tubing connection area, including bonded junctions and connection points, for any visible irregularities, gaps, deformation, incomplete bonding, leakage, or other abnormalities consistent with the examples shown above.
If a defect is observed, or if leakage or connection difficulties are encountered during setup or priming, the product should not be used for patient care and discarded. Customers should report affected product through their normal complaint reporting process and retain the product for evaluation if requested.
Direct consignees were instructed to notify retail customers of the market action. |
| Quantity in Commerce | 1,155,430 |
| Distribution | Worldwide distribution - US Nationwide and the country of Canada. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|