• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Presource Convenience Kit

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Presource Convenience Kitsee related information
Date Initiated by FirmOctober 01, 2026
Date PostedNovember 13, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0456-2027
Recall Event ID 100062
Product Classification Accessories, catheter - Product Code KGZ
ProductPresource Convenience kits containing recalled WelMed Extension Tubing Sets 20IN EXTENSION TBG PACK SAN76ETMDI
Code Information SAN76ETMDI UDI-DI 10195594230456 Kit lots 276274 378811
FEI Number 3001236905
Recalling Firm/
Manufacturer
Cardinal Health 200, LLC
3651 Birchwood Dr
Waukegan IL 60085-8337
For Additional Information ContactJanna Parks
847-887-1114
Manufacturer Reason
for Recall
Kits include Extension Tubing Sets recalled by supplier WelMed. Affected tubing sets may leak.
FDA Determined
Cause 2
Component design/selection
ActionFirm began notifying customers on October 1, 2026. Prior to use, users should visually inspect the tubing connection area, including bonded junctions and connection points, for any visible irregularities, gaps, deformation, incomplete bonding, leakage, or other abnormalities consistent with the examples shown above. If a defect is observed, or if leakage or connection difficulties are encountered during setup or priming, the product should not be used for patient care and discarded. Customers should report affected product through their normal complaint reporting process and retain the product for evaluation if requested. Direct consignees were instructed to notify retail customers of the market action.
Quantity in Commerce1,155,430
DistributionWorldwide distribution - US Nationwide and the country of Canada.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-