| | Class 2 Device Recall Presource Convenience Kit |  |
| Date Initiated by Firm | October 01, 2026 |
| Date Posted | November 13, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0458-2027 |
| Recall Event ID |
100062 |
| Product Classification |
Dressing change tray - Product Code OXQ
|
| Product | Presource Convenience kits containing recalled WelMed Extension Tubing Sets
MIDLINE DRESSING CHANGE TRAY
03B9836B
03B9836D |
| Code Information |
03B9836B
UDI-DI 10195594966355
Kit lot 228524
03B9836D
UDI-DI 10197106647318
Kit lots 268705
305691
315304
343765
387319
512707
589157
617837
635838
645416
733424
757623
|
| FEI Number |
3001236905
|
Recalling Firm/ Manufacturer |
Cardinal Health 200, LLC
3651 Birchwood Dr
Waukegan
IL
60085-8337
|
| For Additional Information Contact | Janna Parks 847-887-1114 |
Manufacturer Reason for Recall | Kits include Extension Tubing Sets recalled by supplier WelMed. Affected tubing sets may leak. |
FDA Determined Cause 2 | Component design/selection |
| Action | Firm began notifying customers on October 1, 2026.
Prior to use, users should visually inspect the tubing connection area, including bonded junctions and connection points, for any visible irregularities, gaps, deformation, incomplete bonding, leakage, or other abnormalities consistent with the examples shown above.
If a defect is observed, or if leakage or connection difficulties are encountered during setup or priming, the product should not be used for patient care and discarded. Customers should report affected product through their normal complaint reporting process and retain the product for evaluation if requested.
Direct consignees were instructed to notify retail customers of the market action. |
| Quantity in Commerce | 1,155,430 |
| Distribution | Worldwide distribution - US Nationwide and the country of Canada. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|