 Class 2 Device Recall FlexLab Xsee related information
Date Initiated by FirmOctober 19, 2026
Date PostedNovember 24, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0550-2027
Recall Event ID 100107
510(K)NumberK121012 
Product Classification Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
ProductFlexLab X; Model/Version: FXX; Component (Description): 1. MEE0074 (Power Cable End Plug at the Automation Track Head, at T Turn Module and L Turn Module), 2. MEE0075 (Power Cable End Socket at the Automation Track Head, at T Turn Module and L Turn Module), 3. MEE0082 (Power Cable Central Power Junction Box at the Automation Track and Track T Module);
Code Information Model/Version: FXX; UDI-DI: 07640172344002; Serial Numbers: FXX.0062, FXX.0065, FXX.0067, FXX.0068, FXX.0069, FXX.0070, FXX.0071, FXX.0072, FXX.0073, FXX.0074, FXX.0075, FXX.0076, FXX.0077, FXX.0082, FXX.0083, FXX.0084, FXX.0085, FXX.0086, FXX.0087, FXX.0088, FXX.0089, FXX.0090, FXX.0092, FXX.0093, FXX.0098, FXX.0099, FXX.0100, FXX.0101, FXX.0103, FXX.0104, FXX.0105, FXX.0106, FXX.0107, FXX.0108, FXX.0110;
FEI Number 3010825766
Recalling Firm/
Manufacturer
Inpeco S.A.
Via San Gottardo 10
Lugano Switzerland
Manufacturer Reason
for Recall
Certain power cables used in the system may have loose electrical connections due to screws or terminals that were not tightened to the required level. This could affect one or more of the cable's internal wires, and may result in the system malfunctioning, causing delays in sample processing or potentially injuring laboratory staff.
FDA Determined
Cause 2
Process control
ActionOn October 19, 2026 Urgent: Medical Device Correction letters were sent to customers. Actions to be taken: The User is advised to apply a notice on/around the FlexLab X product in order to be sure that all personnel who operate or service the automation system are aware about the issue and the action listed above. These precautions do not correct the potentially loose terminal connection. Your service provider will contact you to schedule the corrective intervention needed to inspect the affected components and, if required, replace them. Until the service visit takes place, users are requested to follow the recommendations in this notice. Please forward this notice to all personnel who operate or service the affected automation system. Inpeco recommends that all personnel are made aware of this notice and follow the instruction reported in the Action to be taken by the user paragraph. Please complete and return the Customer Letter Receipt Confirmation and Implementation Check form attached to this letter within 30 days directly to the email address in the accompanying email.
Quantity in Commerce35 units;
DistributionWorldwide - US Nationwide distribution in the states of MO and the countries of Denmark, Egypt, France, Greece, Hungary, Italy, Japan, Malaysia, Netherlands, Romania, Saudi Arabia, Spain, Switzerland, Turkey.
Total Product Life CycleTPLC Device Report