| | Class 2 Device Recall FlexLab X |  |
| Date Initiated by Firm | October 19, 2026 |
| Date Posted | November 24, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0550-2027 |
| Recall Event ID |
100107 |
| 510(K)Number | K121012 |
| Product Classification |
Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
|
| Product | FlexLab X;
Model/Version: FXX;
Component (Description): 1. MEE0074 (Power Cable End Plug at the Automation Track Head, at T Turn Module and L Turn Module), 2. MEE0075 (Power Cable End Socket at the Automation Track Head, at T Turn Module and L Turn Module), 3. MEE0082 (Power Cable Central Power Junction Box at the Automation Track and Track T Module); |
| Code Information |
Model/Version: FXX;
UDI-DI: 07640172344002;
Serial Numbers: FXX.0062, FXX.0065, FXX.0067, FXX.0068, FXX.0069, FXX.0070, FXX.0071, FXX.0072, FXX.0073, FXX.0074, FXX.0075, FXX.0076, FXX.0077, FXX.0082, FXX.0083, FXX.0084, FXX.0085, FXX.0086, FXX.0087, FXX.0088, FXX.0089, FXX.0090, FXX.0092, FXX.0093, FXX.0098, FXX.0099, FXX.0100, FXX.0101, FXX.0103, FXX.0104, FXX.0105, FXX.0106, FXX.0107, FXX.0108, FXX.0110; |
| FEI Number |
3010825766
|
Recalling Firm/ Manufacturer |
Inpeco S.A.
Via San Gottardo 10
Lugano
Switzerland
|
Manufacturer Reason for Recall | Certain power cables used in the system may have loose electrical connections due to screws or terminals that were not tightened to the required level. This could affect one or more of the cable's internal wires, and may result in the system malfunctioning, causing delays in sample processing or potentially injuring laboratory staff. |
FDA Determined Cause 2 | Process control |
| Action | On October 19, 2026 Urgent: Medical Device Correction letters were sent to customers.
Actions to be taken:
The User is advised to apply a notice on/around the FlexLab X product in order to be sure that all
personnel who operate or service the automation system are aware about the issue and the
action listed above.
These precautions do not correct the potentially loose terminal connection.
Your service provider will contact you to schedule the corrective intervention needed to inspect the affected components and, if required, replace them.
Until the service visit takes place, users are requested to follow the recommendations in this notice.
Please forward this notice to all personnel who operate or service the affected automation
system. Inpeco recommends that all personnel are made aware of this notice and follow the
instruction reported in the Action to be taken by the user paragraph.
Please complete and return the Customer Letter Receipt Confirmation and Implementation
Check form attached to this letter within 30 days directly to the email address in the
accompanying email. |
| Quantity in Commerce | 35 units; |
| Distribution | Worldwide - US Nationwide distribution in the states of MO and the countries of Denmark, Egypt, France, Greece, Hungary, Italy, Japan, Malaysia, Netherlands, Romania, Saudi Arabia, Spain, Switzerland, Turkey.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = JJE
|
|
|
|