 Class 2 Device Recall Skytron INTEGRITY 215 SG Steam Sterilizersee related information
Date Initiated by FirmSeptember 07, 2026
Date PostedNovember 20, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0506-2027
Recall Event ID 100108
510(K)NumberK120149 
Product Classification Sterilizer, steam - Product Code FLE
ProductSKYTRON INTEGRITY Model 215 SG Steam Sterilizer
Code Information UDI/DI: 10841736100087 Serial Number: 23074986, 23074987, 23074988, 23074989, 23074990, 23074991, 23074992, 23074993, 23074994, 23074995, 23105060, 23105063, 23105061, 23105064, 23105062, 23115065, 23115067, 23115066, 23125070, 23115069, 23125071, 23115068, 24095224, 23125072, 23125073, 24095225, 24095220, 24095229, 24095226, 24095227, 24095228, 25025380, 25025381, 25025382, 25025383, 25035384, 25035386, 25035387, 25035385, 25035388, 25045418, 25045419, 25045417, 25045420, 25045416, 25045421, 25045422, 25045423, 25055426, 25055424, 25055425, 25055427, 25055430, 25055429, 25055431, 25065432, 25065434, 25065433, 25065435, 25065436, 25065439, 25065437, 25065438, 25055428, 25075443, 25075440, 25075441, 25075442, 25075444, 25075445, 26015588, 26015589, 26015590, 26015591, 26015592, 26075593, 26075594, 26075595, 26075596, 26075597, 26075598
FEI Number 3009685834
Recalling Firm/
Manufacturer
Sakura Seiki Co., Ltd. Chikuma Factory
46-1
Oazashinden
Chikuma Japan
Manufacturer Reason
for Recall
Potential for the device to overheat and shut down
FDA Determined
Cause 2
Process control
ActionSakura Seiki Co., LTD issued a Notice of Corrective Service for Your Steam Sterilizer to its consignees on 9/7/2026 via email. The notice explained the issue and stated that their authorized service personnel would visit each consignee's facility to perform the necessary correction. Each consignee will be contacted to arrange a date and time for the service visit.
Quantity in Commerce81 units
DistributionUS Nationwide distribution in the states of AL, AR, CA, FL, GA, IA, ID, IL, KS, LA, MA, MI, MO, MS, NE, OH, OR, TN, TX, UT.
Total Product Life CycleTPLC Device Report