| | Class 2 Device Recall Skytron INTEGRITY 215 SG Steam Sterilizer |  |
| Date Initiated by Firm | September 07, 2026 |
| Date Posted | November 20, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0506-2027 |
| Recall Event ID |
100108 |
| 510(K)Number | K120149 |
| Product Classification |
Sterilizer, steam - Product Code FLE
|
| Product | SKYTRON INTEGRITY Model 215 SG Steam Sterilizer |
| Code Information |
UDI/DI: 10841736100087
Serial Number: 23074986, 23074987, 23074988, 23074989, 23074990, 23074991, 23074992, 23074993, 23074994, 23074995, 23105060, 23105063, 23105061, 23105064, 23105062, 23115065, 23115067, 23115066, 23125070, 23115069, 23125071, 23115068, 24095224, 23125072, 23125073, 24095225, 24095220, 24095229, 24095226, 24095227, 24095228, 25025380, 25025381, 25025382, 25025383, 25035384, 25035386, 25035387, 25035385, 25035388, 25045418, 25045419, 25045417, 25045420, 25045416, 25045421, 25045422, 25045423, 25055426, 25055424, 25055425, 25055427, 25055430, 25055429, 25055431, 25065432, 25065434, 25065433, 25065435, 25065436, 25065439, 25065437, 25065438, 25055428, 25075443, 25075440, 25075441, 25075442, 25075444, 25075445, 26015588, 26015589, 26015590, 26015591, 26015592, 26075593, 26075594, 26075595, 26075596, 26075597, 26075598
|
| FEI Number |
3009685834
|
Recalling Firm/ Manufacturer |
Sakura Seiki Co., Ltd. Chikuma Factory
46-1
Oazashinden
Chikuma
Japan
|
Manufacturer Reason for Recall | Potential for the device to overheat and shut down |
FDA Determined Cause 2 | Process control |
| Action | Sakura Seiki Co., LTD issued a Notice of Corrective Service for Your Steam Sterilizer to its consignees on 9/7/2026 via email. The notice explained the issue and stated that their authorized service personnel would visit each consignee's facility to perform the necessary correction. Each consignee will be contacted to arrange a date and time for the service visit. |
| Quantity in Commerce | 81 units |
| Distribution | US Nationwide distribution in the states of AL, AR, CA, FL, GA, IA, ID, IL, KS, LA, MA, MI, MO, MS, NE, OH, OR, TN, TX, UT. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|