| | Class 2 Device Recall Presource Convenience Kit |  |
| Date Initiated by Firm | October 01, 2026 |
| Date Posted | November 13, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0463-2027 |
| Recall Event ID |
100062 |
| Product Classification |
Gynecological laparoscopic kit - Product Code OHD
|
| Product | Presource Convenience kits containing recalled WelMed Extension Tubing Sets
MPT HSG IR PACK
41-6606D
41-6606E
41-6606G
41-6606H
ULTRASOUND TRAY
41-8095B
41-8095C
41-8095D |
| Code Information |
41-6606D
UDI-DI 10197106449301
Kit lots 413943
434835
505458
554305
41-6606E
UDI-DI 10198956150164
Kit lot 579365
41-6606G
UDI-DI 10198956178304
Kit lot 654119
41-6606H
UDI-DI 10198956254107
Kit lot 657225
41-8095B
UDI-DI 10195594705220
Kit lots 509564
510534
511507
511604
511858
511948
512047
513402
514008
514343
514863
515284
515450
515657
515895
516022
516648
517326
517438
517464
518601
519068
41-8095C
UDI-DI 10198956315044
Kit lot 520383
41-8095D
UDI-DI 10198956344327
Kit lot 20647
|
| FEI Number |
3001236905
|
Recalling Firm/ Manufacturer |
Cardinal Health 200, LLC
3651 Birchwood Dr
Waukegan
IL
60085-8337
|
| For Additional Information Contact | Janna Parks 847-887-1114 |
Manufacturer Reason for Recall | Kits include Extension Tubing Sets recalled by supplier WelMed. Affected tubing sets may leak. |
FDA Determined Cause 2 | Component design/selection |
| Action | Firm began notifying customers on October 1, 2026.
Prior to use, users should visually inspect the tubing connection area, including bonded junctions and connection points, for any visible irregularities, gaps, deformation, incomplete bonding, leakage, or other abnormalities consistent with the examples shown above.
If a defect is observed, or if leakage or connection difficulties are encountered during setup or priming, the product should not be used for patient care and discarded. Customers should report affected product through their normal complaint reporting process and retain the product for evaluation if requested.
Direct consignees were instructed to notify retail customers of the market action. |
| Quantity in Commerce | 1,155,430 |
| Distribution | Worldwide distribution - US Nationwide and the country of Canada. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|