| | Class 2 Device Recall Medline convenience kits |  |
| Date Initiated by Firm | September 24, 2026 |
| Date Posted | November 10, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0414-2027 |
| Recall Event ID |
100112 |
| Product Classification |
Cardiovascular procedure kit - Product Code OEZ
|
| Product | Medline convenience kits containing BD Urinary Catheter Trays that include CliniClean Pericare Wipes
ADULT HEART CIRCULATOR KIT
DYKS1120G
AHT CVOR CABG
CDS983346K
CARDIOPULMONARY MPH
DYNJT5258A
CV HEART GOWN & DRAPE PACK A
DYNJ912460
FEM POP
CDS982028W
OPEN HEART PROCEDURE
CDS980895X
VASCULAR
DYNJ911429
ENDOSCOPIC SKULL BASE
DYNJ908362F |
| Code Information |
DYKS1120G
UDI-DI 10198459335808
Kit lots 26DMI051
26EMJ713
CDS983346K
UDI-DI 10198459701993
Kit lot 26DBH084
DYNJT5258A
UDI-DI 10198459661495
Kit lot 26DBB503
DYNJ912460
UDI-DI 10198459726880
Kit lot 26DLA824
CDS982028W
UDI-DI 10198459385605
Kit lot 25LMH727
CDS980895X
UDI-DI 10198459662904
Kit lot 26EMA865
DYNJ911429
UDI-DI 10198459341014
Kit lots 26AMJ586
26CMF189
26FME829
26FMG294
DYNJ908362F
UDI-DI 10198459308581
Kit lot 25LLA930
|
| FEI Number |
1417592
|
Recalling Firm/ Manufacturer |
Medline Industries, LP
3 Lakes Dr
Northfield
IL
60093-2753
|
| For Additional Information Contact | Karin Johnson 800-633-5463 |
Manufacturer Reason for Recall | Convenience kits contain additional product codes and lots of recalled CliniClean Pericare Wipes identified by manufacturer. Wipes may be contaminated with a lubricant used in the manufacturing process. Use of affected wipes may cause localized skin effects such as irritation, erythema, or mild inflammation. Failure to perform pericare may increase the risk of localized infection or catheter-associated urinary tract infection. |
FDA Determined Cause 2 | Process change control |
| Action | This is an update to a previously issued letter due to the manufacturer/supplier identifying additional lots of affected product.
On September 24, 2026, customers were notified of the additional products and lots affected.
Medline Industries LP instructed customers to request stickers to over-label the affected convenience kits with a sticker indicating to remove the affected component and replace it with product from supply. No kits are to be returned. |
| Quantity in Commerce | 7,043 kits |
| Distribution | US Nationwide distribution in the states of AL, AR, CA, FL, GA, IL, IN, LA, MA, MN, MO, MS, MT, NC, NE, NV, NY, OH, OK, RI, SC, TN, TX, WI. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|