U.S. Department of Health and Human Services

Class 2 Device Recall YeastOne Broth

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall YeastOne Brothsee related information
Date Initiated by FirmOctober 19, 2026
Date PostedNovember 03, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0355-2027
Recall Event ID 100139
510(K)NumberK090968 
Product Classification Susceptibility test plate, antifungal - Product Code NGZ
ProductYeastone Broth, 11ML, 10/BOX, used in antibiotic susceptibility testing of mycological clinical specimens
Code Information YY3462 lot 440400 UDI-DI 848838018817
FEI Number 1924669
Recalling Firm/
Manufacturer
Remel, Inc
12076 Santa Fe Trail Dr
Lenexa KS 66215-3519
For Additional Information ContactErica Knox
800-255-6730
Manufacturer Reason
for Recall
Affected lot did not meet specifications. Product may report false susceptibility to antifungal agents, specifically within the Azole family.
FDA Determined
Cause 2
Under Investigation by firm
ActionThe firm began notifying affected customers on October 19, 2026. Customers were instructed to notify any personnel who needed to be aware of the issue. Customers should review previously reported susceptibility results generated using the affected lot and assess whether retesting is appropriate. Customers should destroy any remaining inventory of the affected lot. Do not use it for any clinical laboratory testing. Customers should complete and return the acknowledgement form accompanying the letter. Firm will replace product or issue credit. If you have any questions, please contact our Technical Services Department at 800-255-6730 (US) or email microbiology.ts.us@thermofisher.com. You should complete the accompanying Acknowledgment Form upon receipt and return to MBD.vigilance@thermofisher.com
Quantity in Commerce33 US, 54 OUS
DistributionWorldwide - US Nationwide distribution in the states of MA and CA. The countries of India, Saudi Arabia, UK.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = NGZ
-
-