Class 1 Device Recall STREAMLINE BLOODLINE SET FOR DIALOG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall STREAMLINE BLOODLINE SET FOR DIALOGsee related information
Date Initiated by FirmOctober 22, 2026
Date PostedNovember 18, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0313-2027
Recall Event ID 100148
510(K)NumberK080807 
Product Classification Set, tubing, blood, with and without anti-regurgitation valve - Product Code FJK
ProductHemodialysis Bloodlines STREAMLINE BLOODLINE SET FOR DIALOG+. Material Number: SL-2010M2096. STREAMLINE BLOODLINE SET FOR DIALOG
Code Information Material Number: SL-2010M2096; UDI-DI Primary: 04046964367786; UDI-DI Unit: 04046964367779; Lot Number (Expiration date): A2600099 (07JAN2029), A2600128 (Exp 07JAN2029), A2600136 (07FEB2029).
FEI Number 2521402
Recalling Firm/
Manufacturer
B Braun Medical Inc
824 12th Ave
Bethlehem PA 18018-3524
For Additional Information ContactAllison Longenhagen
001-484-2408373
Manufacturer Reason
for Recall
Potential inadequate bonding of the arterial tubing connection to the dialyzer connector may result in detachment, separation, or leaking of the arterial tubing during priming or dialysis treatment.
FDA Determined
Cause 2
Process control
ActionB. Braun Medical. Inc. notified consignees on about 10/22/2026 via mailed "URGENT MEDICAL DEVICE RECALL NOTIFICATION" letter. Consignees were made aware of the issue and potential hazard. They were instructed to review the notification letter entirely for awareness, ensure all users withing the organization are informed of the issue, notify customers if affected units have been further distributed, and determine current inventory of affected units and cease use and quarantine them immediately. Complete and return the "Urgent Medical Device Recall Acknowledgment Form" provided and arrange for the return of all affected units.
Quantity in Commerce66,262 units
DistributionNationwide distribution in the US.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.