| Date Initiated by Firm | October 22, 2026 |
| Date Posted | November 18, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0313-2027 |
| Recall Event ID |
100148 |
| 510(K)Number | K080807 |
| Product Classification |
Set, tubing, blood, with and without anti-regurgitation valve - Product Code FJK
|
| Product | Hemodialysis Bloodlines STREAMLINE BLOODLINE SET FOR DIALOG+. Material Number: SL-2010M2096.
STREAMLINE BLOODLINE SET FOR DIALOG |
| Code Information |
Material Number: SL-2010M2096;
UDI-DI Primary: 04046964367786;
UDI-DI Unit: 04046964367779;
Lot Number (Expiration date): A2600099 (07JAN2029), A2600128 (Exp 07JAN2029), A2600136 (07FEB2029).
|
| FEI Number |
2521402
|
Recalling Firm/ Manufacturer |
B Braun Medical Inc
824 12th Ave
Bethlehem
PA
18018-3524
|
| For Additional Information Contact | Allison Longenhagen 001-484-2408373 |
Manufacturer Reason for Recall | Potential inadequate bonding of the arterial tubing connection to the dialyzer connector may result in detachment, separation, or leaking of the arterial tubing during priming or dialysis treatment. |
FDA Determined Cause 2 | Process control |
| Action | B. Braun Medical. Inc. notified consignees on about 10/22/2026 via mailed "URGENT MEDICAL DEVICE RECALL NOTIFICATION" letter. Consignees were made aware of the issue and potential hazard. They were instructed to review the notification letter entirely for awareness, ensure all users withing the organization are informed of the issue, notify customers if affected units have been further distributed, and determine current inventory of affected units and cease use and quarantine them immediately. Complete and return the "Urgent Medical Device Recall Acknowledgment Form" provided and arrange for the return of all affected units. |
| Quantity in Commerce | 66,262 units |
| Distribution | Nationwide distribution in the US. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = FJK
|