| | Class 3 Device Recall Dual Invasive Blood Pressure (IBP) Adapter |  |
| Date Initiated by Firm | October 16, 2026 |
| Date Posted | November 25, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0554-2027 |
| Recall Event ID |
100156 |
| Product Classification |
Amplifier and signal conditioner, transducer signal - Product Code DRQ
|
| Product | Dual Invasive Blood Pressure (IBP) Adapter. Product/Model Number: 989803199741. Utilized in the below systems with corresponding Product/Model Numbers.
1. IntelliVue Multi-Measurement Module X3; Product/Model Number: 867030.
2. IntelliVue Patient Monitor MX100; Product/Model Number: 867033.
3. IntelliVue MMX; Product/Model Number: 867036.
4. IntelliVue Hemodynamic Extension; Product/Model Number: 867039.
5. IntelliVue Capnography Extension; Product/Model Number: 867040.
6. IntelliVue Microstream Extension; Product/Model Number: 867041. |
| Code Information |
Product/Model Number: 989803199741. UDI-DI: 00884838085299.
1. Product/Model Number: 867030. UDI: (01)00884838082588(21)DE61110031, (01)00884838082588(21)DE61111265, (01)00884838082588(21)DE61111290, (01)00884838082588(21)DE61114687, (01)00884838082588(21)DE61115539, (01)00884838082588(21)DE61115549, (01)00884838082588(21)DE61115735, (01)00884838082588(21)DE61116830, (01)00884838082588(21)DE61117433, (01)00884838082588(21)DE61117921, (01)00884838082588(21)DE61118030, (01)00884838082588(21)DE61118199, (01)00884838082588(21)DE61119328, (01)00884838082588(21)DE61119721, (01)00884838082588(21)DE61120392, (01)00884838082588(21)DE61120496, (01)00884838082588(21)DE61120683, (01)00884838082588(21)DE61121104, (01)00884838082588(21)DE61121185, (01)00884838082588(21)DE61121194, (01)00884838082588(21)DE61121453, (01)00884838082588(21)DE61122685, (01)00884838082588(21)DE61122698, (01)00884838082588(21)DE61123184, (01)00884838082588(21)DE61123219, (01)00884838082588(21)DE61123764, (01)00884838082588(21)DE61124383, (01)00884838082588(21)DE61124458, (01)00884838082588(21)DE61124863, (01)00884838082588(21)DE61125847, (01)00884838082588(21)DE61126277, (01)00884838082588(21)DE61126732, (01)00884838082588(21)DE61126744, (01)00884838082588(21)DE61127040, (01)00884838082588(21)DE61127516, (01)00884838082588(21)DE61127865, (01)00884838082588(21)DE61129238, (01)00884838082588(21)DE61129510, (01)00884838082588(21)DE61130319, (01)00884838082588(21)DE61133614, (01)00884838082588(21)DE61133970, (01)00884838082588(21)DE61134050, (01)00884838082588(21)DE61134275, (01)00884838082588(21)DE61135176, (01)00884838082588(21)DE61136862, (01)00884838082588(21)DE61137208, (01)00884838082588(21)DE61137637, (01)00884838082588(21)DE61138105, (01)00884838082588(21)DE61138208, (01)00884838082588(21)DE61138619, (01)00884838082588(21)DE61139164, (01)00884838082588(21)DE61139184, (01)00884838082588(21)DE61139562, (01)00884838082588(21)DE61139637, (01)00884838082588(21)DE61140703, (01)00884838082588(21)DE61141907, (01)00884838082588(21)DE61142326, (01)00884838082588(21)DE61142502, (01)00884838082588(21)DE61143169, (01)00884838082588(21)DE61143211, (01)00884838082588(21)DE61144511, (01)00884838082588(21)DE61145515, (01)00884838082588(21)DE61145737, (01)00884838082588(21)DE61145752, (01)00884838082588(21)DE61146888, (01)00884838082588(21)DE61147120, (01)00884838082588(21)DE61147126, (01)00884838082588(21)DE61147783, (01)00884838082588(21)DE61148215, (01)00884838082588(21)DE61150906, (01)00884838082588(21)DE61152421, (01)00884838082588(21)DE61153560, (01)00884838082588(21)DE61153564, (01)00884838082588(21)DE61153654, (01)00884838082588(21)DE61153768, (01)00884838082588(21)DE61154071, (01)00884838082588(21)DE61154808, (01)00884838082588(21)DE61155270, (01)00884838082588(21)DE61158450, (01)00884838082588(21)DE61158870, (01)00884838082588(21)DE61158941, (01)00884838082588(21)DE61158949, (01)00884838082588(21)DE61159033, (01)00884838082588(21)DE61160154, (01)00884838082588(21)DE61160233, (01)00884838082588(21)DE61161366, (01)00884838082588(21)DE61161541, (01)00884838082588(21)DE61162297, (01)00884838082588(21)DE61162580, (01)00884838082588(21)DE61163068, (01)00884838082588(21)DE61163241, (01)00884838082588(21)DE61163289, (01)00884838082588(21)DE61163808, (01)00884838082588(21)DE61165407, (01)00884838082588(21)DE61165542, (01)00884838082588(21)DE61166113, (01)00884838082588(21)DE61169398, (01)00884838082588(21)DE69470381, (01)00884838082588(21)DE69472099, (01)00884838082588(21)DE69472860, (01)00884838082588(21)DE69473929, (01)00884838082588(21)DE69474015, (01)00884838082588(21)DE69474020, (01)00884838082588(21)DE69475260, (01)00884838082588(21)DE69475416, (01)00884838082588(21)DE69475589, (01)00884838082588(21)DE69475933, (01)00884838082588(21)DE69482998, (01)00884838082588(21)DE69484111, (01)00884838082588(21)DE69485007, (01)00884838082588(21)DE69485710, (01)00884838082588(21)DE69485912, (01)00884838082588(21)DE69486075, (01)00884838082588(21)DE69492361, (01)00884838082588(21)DE69492849, (01)00884838082588(21)DE69497519, (01)00884838082588(21)DE69498581, (01)00884838082588(21)DE69498767, (01)00884838082588(21)DE694A2431, (01)00884838082588(21)DE694A4899, (01)00884838082588(21)DE694A6453, (01)00884838082588(21)DE694A9854, (01)00884838082588(21)DE694B0400, (01)00884838082588(21)DE694B1181, (01)00884838082588(21)DE694B1204, (01)00884838082588(21)DE694B1324, (01)00884838082588(21)DE694B2173, (01)00884838082588(21)DE694B3746, (01)00884838082588(21)DE694B5547, (01)00884838082588(21)DE694B6492, (01)00884838082588(21)DE694B6519, (01)00884838082588(21)DE694B7327, (01)00884838082588(21)DE694B7651, (01)00884838082588(21)DE694B8338, (01)00884838082588(21)DE694C1648, (01)00884838082588(21)DE694C1837, (01)00884838082588(21)DE694C7709, (01)00884838082588(21)DE694C8850, (01)00884838082588(21)DE694D0264, (01)00884838082588(21)DE694D0498, (01)00884838082588(21)DE694D1175, (01)00884838082588(21)DE694D1671, (01)00884838082588(21)DE694D3152, (01)00884838082588(21)DE694D3984, (01)00884838082588(21)DE694D4276, (01)00884838082588(21)DE694D8280, (01)00884838082588(21)DE694D8284, (01)00884838082588(21)DE694E0044, (01)00884838082588(21)DE694E2396, (01)00884838082588(21)DE694E4497, (01)00884838082588(21)DE694E6012, (01)00884838082588(21)DE694E6029, (01)00884838082588(21)DE694E7553, (01)00884838082588(21)DE694E9324, (01)00884838082588(21)DE694F3852, (01)00884838082588(21)DE694F4366, (01)00884838082588(21)DE694F6607, (01)00884838082588(21)DE694F6820, (01)00884838082588(21)DE694F6991, (01)00884838082588(21)DE694F7289, (01)00884838082588(21)DE694F9477, (01)00884838082588(21)DE694F9549, (01)00884838082588(21)DE694G1639, (01)00884838082588(21)DE694G2454, (01)00884838082588(21)DE694G3113, (01)00884838082588(21)DE694G4663, (01)00884838082588(21)DE694G4826, (01)00884838082588(21)DE694G4844, (01)00884838082588(21)DE694G5356, (01)00884838082588(21)DE694G5423, (01)00884838082588(21)DE694G5450, (01)00884838082588(21)DE694G5849, (01)00884838082588(21)DE694G6256, (01)00884838082588(21)DE694G6930, (01)00884838082588(21)DE694G6934, (01)00884838082588(21)DE694G7264, (01)00884838082588(21)DE694G7714, (01)00884838082588(21)DE694G7730, (01)00884838082588(21)DE694G7753, (01)00884838082588(21)DE694G7948, (01)00884838082588(21)DE694G8356, (01)00884838082588(21)DE694G8515, (01)00884838082588(21)DE694G9283, (01)00884838082588(21)DE694H0344, (01)00884838082588(21)DE694H0505, (01)00884838082588(21)DE694H1205, (01)00884838082588(21)DE694H1597, (01)00884838082588(21)DE694H2244, (01)00884838082588(21)DE694H2714, (01)00884838082588(21)DE694H2833, (01)00884838082588(21)DE694H3104, (01)00884838082588(21)DE694H3204, (01)00884838082588(21)DE694H4276, (01)00884838082588(21)DE694H5395, (01)00884838082588(21)DE694H6200, (01)00884838082588(21)DE694H6219, (01)00884838082588(21)DE694H6891, (01)00884838082588(21)DE694H8145, (01)00884838082588(21)DE694H8160, (01)00884838082588(21)DE694H8180, (01)00884838082588(21)DE694H8563, (01)00884838082588(21)DE694H8670, (01)00884838082588(21)DE694H9051, (01)00884838082588(21)DE694J1918, (01)00884838082588(21)DE694J2924, (01)00884838082588(21)DE694J3043, (01)00884838082588(21)DE694J3958, (01)00884838082588(21)DE694J4251, (01)00884838082588(21)DE694J4572, (01)00884838082588(21)DE694J4783, (01)00884838082588(21)DE694J6216, (01)00884838082588(21)DE694J6359, (01)00884838082588(21)DE694J6499, (01)00884838082588(21)DE694J6785, (01)00884838082588(21)DE694J7004, (01)00884838082588(21)DE694J7035, (01)00884838082588(21)DE694J7245, (01)00884838082588(21)DE694J7488, (01)00884838082588(21)DE694J8117, (01)00884838082588(21)DE694J8362, (01)00884838082588(21)DE694J8509, (01)00884838082588(21)DE694J8519, (01)00884838082588(21)DE694J8905, (01)00884838082588(21)DE694J9157, (01)00884838082588(21)DE694J9648, (01)00884838082588(21)DE694J9666, (01)00884838082588(21)DE694J9680, (01)00884838082588(21)DE694K0137, (01)00884838082588(21)DE694K0589, (01)00884838082588(21)DE694K0698, (01)00884838082588(21)DE694K1202, (01)00884838082588(21)DE694K1591, (01)00884838082588(21)DE694K2047, (01)00884838082588(21)DE752K2471, (01)00884838082588(21)DE752K2779, (01)00884838082588(21)DE752K2849, (01)00884838082588(21)DE752K2900, (01)00884838082588(21)DE752K4303, (01)00884838082588(21)DE752K4716, (01)00884838082588(21)DE752K5525, (01)00884838082588(21)DE752K6493, (01)00884838082588(21)DE752K6588, (01)00884838082588(21)DE752K6721, (01)00884838082588(21)DE752K6797, (01)00884838082588(21)DE752K6834, (01)00884838082588(21)DE752K7293, (01)00884838082588(21)DE752K7319, (01)00884838082588(21)DE752K8613, (01)00884838082588(21)DE752K9346, (01)00884838082588(21)DE752K9443, (01)00884838082588(21)DE752K9999, (01)00884838082588(21)DE752L0295, (01)00884838082588(21)DE752L1080, (01)00884838082588(21)DE752L1714, (01)00884838082588(21)DE752L1839, (01)00884838082588(21)DE752L1851, (01)00884838082588(21)DE752L2605, (01)00884838082588(21)DE752L3141, (01)00884838082588(21)DE752L3165, (01)00884838082588(21)DE752L3252, (01)00884838082588(21)DE752L3392, (01)00884838082588(21)DE752L3528, (01)00884838082588(21)DE752L3807, (01)00884838082588(21)DE752L3884, (01)00884838082588(21)DE752L3896, (01)00884838082588(21)DE752L4170, (01)00884838082588(21)DE752L4193, (01)00884838082588(21)DE752L4302, (01)00884838082588(21)DE752L5092, (01)00884838082588(21)DE752L5415, (01)00884838082588(21)DE752L5422, (01)00884838082588(21)DE752L5438, (01)00884838082588(21)DE752L5561, (01)00884838082588(21)DE752L5880, (01)00884838082588(21)DE752L6157, (01)00884838082588(21)DE752L6740, (01)00884838082588(21)DE752L6750, (01)00884838082588(21)DE752L7727, (01)00884838082588(21)DE752L7734, (01)00884838082588(21)DE752L7800, (01)00884838082588(21)DE752L9046, (01)00884838082588(21)DE752L9085, (01)00884838082588(21)DE752L9205, (01)00884838082588(21)DE752L9260, (01)00884838082588(21)DE752L9516, (01)00884838082588(21)DE752L9785, (01)00884838082588(21)DE752M0931, (01)00884838082588(21)DE752M0937, (01)00884838082588(21)DE752M1075, (01)00884838082588(21)DE752M2031, (01)00884838082588(21)DE752M3146, (01)00884838082588(21)DE752M3517, (01)00884838082588(21)DE752M4111, (01)00884838082588(21)DE752M4295, (01)00884838082588(21)DE752M4871, (01)00884838082588(21)DE752M4926, (01)00884838082588(21)DE752M4943, (01)00884838082588(21)DE752M5063, (01)00884838082588(21)DE752M5918, (01)00884838082588(21)DE752M6029, (01)00884838082588(21)DE752M6055, (01)00884838082588(21)DE752M6203, (01)00884838082588(21)DE752M6368, (01)00884838082588(21)DE752M6473, (01)00884838082588(21)DE752M6544, (01)00884838082588(21)DE752M6918, (01)00884838082588(21)DE752M6936, (01)00884838082588(21)DE752M7271, (01)00884838082588(21)DE752M7469, (01)00884838082588(21)DE752M8350, (01)00884838082588(21)DE752M8884, (01)00884838082588(21)DE752N0135, (01)00884838082588(21)DE752N0168, (01)00884838082588(21)DE752N1275, (01)00884838082588(21)DE752N1555, (01)00884838082588(21)DE752N1962, (01)00884838082588(21)DE752N1965, (01)00884838082588(21)DE752N2052, (01)00884838082588(21)DE752N2055, (01)00884838082588(21)DE752N2107, (01)00884838082588(21)DE752N2832, (01)00884838082588(21)DE752N3207, (01)00884838082588(21)DE752N3348, (01)00884838082588(21)DE752N4261, (01)00884838082588(21)DE752N4574, (01)00884838082588(21)DE752N5030, (01)00884838082588(21)DE752N5781, (01)00884838082588(21)DE752N6432, (01)00884838082588(21)DE752N6485, (01)00884838082588(21)DE752N7318, (01)00884838082588(21)DE752N7497, (01)00884838082588(21)DE752N7849, (01)00884838082588(21)DE752N8076, (01)00884838082588(21)DE752N8697, (01)00884838082588(21)DE752N9035, (01)00884838082588(21)DE752N9129, (01)00884838082588(21)DE752N9487, (01)00884838082588(21)DE752N9537, (01)00884838082588(21)DE752P0484, (01)00884838082588(21)DE752P0565, (01)00884838082588(21)DE752P0632, (01)00884838082588(21)DE752P1013, (01)00884838082588(21)DE752P2670, (01)00884838082588(21)DE752P2809, (01)00884838082588(21)DE752P2908, (01)00884838082588(21)DE752P3122, (01)00884838082588(21)DE752P3128, (01)00884838082588(21)DE752P3133, (01)00884838082588(21)DE752P3191, (01)00884838082588(21)DE752P3399, (01)00884838082588(21)DE752P3577, (01)00884838082588(21)DE752P3578, (01)00884838082588(21)DE752P3580, (01)00884838082588(21)DE752P3706, (01)00884838082588(21)DE752P4021, (01)00884838082588(21)DE752P4034, (01)00884838082588(21)DE752P4038, (01)00884838082588(21)DE752P4110, (01)00884838082588(21)DE752P4403, (01)00884838082588(21)DE752P4866, (01)00884838082588(21)DE752P5042, (01)00884838082588(21)DE752P6405, (01)00884838082588(21)DE752P6476, (01)00884838082588(21)DE752P7180, (01)00884838082588(21)DE752P7186, (01)00884838082588(21)DE752P7194, (01)00884838082588(21)DE752P7268, (01)00884838082588(21)DE752P7426, (01)00884838082588(21)DE752P8260, (01)00884838082588(21)DE752P8849, (01)00884838082588(21)DE752P8990, (01)00884838082588(21)DE752P9028, (01)00884838082588(21)DE752P9049, (01)00884838082588(21)DE752P9166, (01)00884838082588(21)DE752P9951, (01)00884838082588(21)DE752R0182, (01)00884838082588(21)DE752R0619, (01)00884838082588(21)DE752R2497, (01)00884838082588(21)DE752R3486, (01)00884838082588(21)DE752R3918, (01)00884838082588(21)DE752R3980, (01)00884838082588(21)DE752R4151, (01)00884838082588(21)DE752R4922, (01)00884838082588(21)DE752R4960, (01)00884838082588(21)DE752R5283, (01)00884838082588(21)DE752R6113, (01)00884838082588(21)DE752R6362, (01)00884838082588(21)DE752R8269, (01)00884838082588(21)DE752R8614, (01)00884838082588(21)DE752R9127, (01)00884838082588(21)DE752T0192, (01)00884838082588(21)DE752T0430, (01)00884838082588(21)DE752T0592, (01)00884838082588(21)DE752T0685, (01)00884838082588(21)DE752T0727, (01)00884838082588(21)DE752T0865, (01)00884838082588(21)DE752T0888, (01)00884838082588(21)DE752T0934, (01)00884838082588(21)DE752T0967, (01)00884838082588(21)DE752T0978, (01)00884838082588(21)DE752T1152, (01)00884838082588(21)DE752T1524, (01)00884838082588(21)DE752T1570, (01)00884838082588(21)DE752T1635, (01)00884838082588(21)DE752T1660, (01)00884838082588(21)DE752T1729, (01)00884838082588(21)DE752T3436, (01)00884838082588(21)DE752T3831, (01)00884838082588(21)DE752T4474, (01)00884838082588(21)DE752T4668, (01)00884838082588(21)DE752T5228, (01)00884838082588(21)DE752T5742, (01)00884838082588(21)DE752T7696, (01)00884838082588(21)DE752T7931, (01)00884838082588(21)DE752T7953, (01)00884838082588(21)DE752T8237, (01)00884838082588(21)DE752T8563, (01)00884838082588(21)DE752T8851, (01)00884838082588(21)DE752T9358, (01)00884838082588(21)DE752T9432, (01)00884838082588(21)DE752T9567, (01)00884838082588(21)DE752U0237, (01)00884838082588(21)DE752U0245, (01)00884838082588(21)DE752U0373, (01)00884838082588(21)DE752U1018, (01)00884838082588(21)DE752U1102, (01)00884838082588(21)DE752U1530, (01)00884838082588(21)DE752U2596, (01)00884838082588(21)DE752U4704, (01)00884838082588(21)DE752U5091, (01)00884838082588(21)DE7610A061, (01)00884838082588(21)DE761U8112, (01)00884838082588(21)DE761U8653, (01)00884838082588(21)DE761U8674, (01)00884838082588(21)DE761V0414, (01)00884838082588(21)DE761V0778, (01)00884838082588(21)DE761V0959, (01)00884838082588(21)DE761V1169, (01)00884838082588(21)DE761V1353, (01)00884838082588(21)DE761V4144, (01)00884838082588(21)DE761V4681, (01)00884838082588(21)DE761V4782, (01)00884838082588(21)DE761V5392, (01)00884838082588(21)DE761V5997, (01)00884838082588(21)DE761V7511, (01)00884838082588(21)DE761V9416, (01)00884838082588(21)DE761W0239, (01)00884838082588(21)DE761W1516, (01)00884838082588(21)DE761W1606, (01)00884838082588(21)DE761W1773, (01)00884838082588(21)DE761W1781, (01)00884838082588(21)DE761W2395, (01)00884838082588(21)DE761W4839, (01)00884838082588(21)DE761W5192, (01)00884838082588(21)DE761W5221, (01)00884838082588(21)DE761W5476, (01)00884838082588(21)DE761W5760, (01)00884838082588(21)DE761W6213, (01)00884838082588(21)DE761W6906, (01)00884838082588(21)DE761W7944, (01)00884838082588(21)DE761W8005, (01)00884838082588(21)DE761Y0501, (01)00884838082588(21)DE761Y0515, (01)00884838082588(21)DE761Y1038, (01)00884838082588(21)DE761Y1484, (01)00884838082588(21)DE761Y2391, (01)00884838082588(21)DE761Y3415, (01)00884838082588(21)DE761Y3416, (01)00884838082588(21)DE761Y3436, (01)00884838082588(21)DE761Y4449, (01)00884838082588(21)DE761Y4531, (01)00884838082588(21)DE761Y4604, (01)00884838082588(21)DE761Y4609, (01)00884838082588(21)DE761Y4616, (01)00884838082588(21)DE761Y4743, (01)00884838082588(21)DE761Y5289, (01)00884838082588(21)DE761Y6503, (01)00884838082588(21)DE761Y6908, (01)00884838082588(21)DE761Y7811, (01)00884838082588(21)DE761Y8178, (01)00884838082588(21)DE7730A5E9, (01)00884838082588(21)DE7730A6L8, (01)00884838082588(21)DE7730A74G, (01)00884838082588(21)DE7730A7R7, (01)00884838082588(21)DE7730A81Y, (01)00884838082588(21)DE7730A8W9, (01)00884838082588(21)DE7730A935, (01)00884838082588(21)DE7730AAPM, (01)00884838082588(21)DE7730AAPR, (01)00884838082588(21)DE7730ABKJ, (01)00884838082588(21)DE7730ABTE, (01)00884838082588(21)DE7730AC52, (01)00884838082588(21)DE7730ACJ3, (01)00884838082588(21)DE7730ACJJ, (01)00884838082588(21)DE7730ACLA, (01)00884838082588(21)DE7840AGK8, (01)00884838082588(21)DE7840AJ6K, (01)00884838082588(21)DE7840AJV5, (01)00884838082588(21)DE7840AK3T, (01)00884838082588(21)DE7840ALZL, (01)00884838082588(21)DE7840AMZ7, (01)00884838082588(21)DE7840AN0G, (01)00884838082588(21)DE7840AT56, (01)00884838082588(21)DE7840AU4D, (01)00884838082588(21)DE7840AUB1, (01)00884838082588(21)DE7840AVWF, (01)00884838082588(21)DE7840AW3W, (01)00884838082588(21)DE7840AZR9, (01)00884838082588(21)DE7840B291, (01)00884838082588(21)DE7840B2MM, (01)00884838082588(21)DE7840B2PW, (01)00884838082588(21)DE7840B2TN, (01)00884838082588(21)DE7840B31C, (01)00884838082588(21)DE7840B4PY, (01)00884838082588(21)DE7840B51J, (01)00884838082588(21)DE7840B5B4, (01)00884838082588(21)DE7840B67P, (01)00884838082588(21)DE7840B6W4, (01)00884838082588(21)DE7840B6WD, (01)00884838082588(21)DE7840B7CU, (01)00884838082588(21)DE7840B7L5, (01)00884838082588(21)DE7840B81T, (01)00884838082588(21)DE7840B85V, (01)00884838082588(21)DE7840B89A, (01)00884838082588(21)DE7840B8AE, (01)00884838082588(21)DE7840B8FB, (01)00884838082588(21)DE7840B8FM, (01)00884838082588(21)DE7840BAM3, (01)00884838082588(21)DE7840BDV8, (01)00884838082588(21)DE7840BG1Y, (01)00884838082588(21)DE7840BHLJ, (01)00884838082588(21)DE7840BJDP, (01)00884838082588(21)DE7840BK4J, (01)00884838082588(21)DE7840BL4K, (01)00884838082588(21)DE7840BPM4, (01)00884838082588(21)DE7840BRC7, (01)00884838082588(21)DE7840BU5K, (01)00884838082588(21)DE7840BULC, (01)00884838082588(21)DE7840BUPR, (01)00884838082588(21)DE7840BV5E, (01)00884838082588(21)DE7840BVBA, (01)00884838082588(21)DE7840BVUD, (01)00884838082588(21)DE7840BVUU, (01)00884838082588(21)DE7840BW63, (01)00884838082588(21)DE7840BY89, (01)00884838082588(21)DE7840BYPT, (01)00884838082588(21)DE7840BYR6, (01)00884838082588(21)DE7840BZ9T, (01)00884838082588(21)DE7840C17D, (01)00884838082588(21)DE7840C1NF, (01)00884838082588(21)DE7840C26L, (01)00884838082588(21)DE7840C3J4, (01)00884838082588(21)DE7840C78Y, (01)00884838082588(21)DE7840C7A0, (01)00884838082588(21)DE7840C873, (01)00884838082588(21)DE7840C9TH, (01)00884838082588(21)DE7840CAYW, (01)00884838082588(21)DE8090CBN0, (01)00884838082588(21)DE8090CFH3, (01)00884838082588(21)DE8090CHFP, (01)00884838082588(21)DE8090CJ9D, (01)00884838082588(21)DE8090CJW9, (01)00884838082588(21)DE8090CK2G, (01)00884838082588(21)DE8090CKCM, (01)00884838082588(21)DE8090CKKH, (01)00884838082588(21)DE8090CKN3, (01)00884838082588(21)DE8090CLL8, (01)00884838082588(21)DE8090CRBH, (01)00884838082588(21)DE8090CRNF, (01)00884838082588(21)DE8090CRNN, (01)00884838082588(21)DE8090CUCV, (01)00884838082588(21)DE8090CUT1, (01)00884838082588(21)DE8090CVK7, (01)00884838082588(21)DE8090CWUL, (01)00884838082588(21)DE8090CY4W, (01)00884838082588(21)DE8090CYCH, (01)00884838082588(21)DE8090D0F0, (01)00884838082588(21)DE8090D0J5, (01)00884838082588(21)DE8090D1RV, (01)00884838082588(21)DE8090D1UV, (01)00884838082588(21)DE8090D2AM, (01)00884838082588(21)DE8090D2DH, (01)00884838082588(21)DE8090D2G6, (01)00884838082588(21)DE8090D33Z, (01)00884838082588(21)DE8090D3JL, (01)00884838082588(21)DE8090D3KY, (01)00884838082588(21)DE8090D3PR, (01)00884838082588(21)DE8090D4P1, (01)00884838082588(21)DE8090D6EW, (01)00884838082588(21)DE8090D6T6, (01)00884838082588(21)DE8090D7UT, (01)00884838082588(21)DE8090D8JF, (01)00884838082588(21)DE8090D9AH, (01)00884838082588(21)DE8090D9KW, (01)00884838082588(21)DE8090D9L9, (01)00884838082588(21)DE8090DBMZ, (01)00884838082588(21)DE8090DCBE, (01)00884838082588(21)DE8090DCDL, (01)00884838082588(21)DE8090DCNP, (01)00884838082588(21)DE8090DDJE, (01)00884838082588(21)DE8090DE4E, (01)00884838082588(21)DE8090DEJD, (01)00884838082588(21)DE8090DER2, (01)00884838082588(21)DE8090DEYN, (01)00884838082588(21)DE8090DF0B, (01)00884838082588(21)DE8090DFER, (01)00884838082588(21)DE8090DH21, (01)00884838082588(21)DE8090DJE7, (01)00884838082588(21)DE8090DLZK, (01)00884838082588(21)DE8090DLZV, (01)00884838082588(21)DE8090DN5W, (01)00884838082588(21)DE8090DTNU, (01)00884838082588(21)DE8090DTRE, (01)00884838082588(21)DE8090DTTL, (01)00884838082588(21)DE8090DUY1, (01)00884838082588(21)DE8090DVUF, (01)00884838082588(21)DE8090DVZ0, (01)00884838082588(21)DE8090DW0J, (01)00884838082588(21)DE8090DW8T, (01)00884838082588(21)DE8090E0PK, (01)00884838082588(21)DE8090E1A3, (01)00884838082588(21)DE8090E1LL, (01)00884838082588(21)DE8090E2LJ, (01)00884838082588(21)DE8090E35G, (01)00884838082588(21)DE8090E38W, (01)00884838082588(21)DE8090E3MG, (01)00884838082588(21)DE8090E6NG, (01)00884838082588(21)DE8090E79J, (01)00884838082588(21)DE8090E7BG, (01)00884838082588(21)DE8090E7CC, (01)00884838082588(21)DE8090E7CF, (01)00884838082588(21)DE8090E7EU, (01)00884838082588(21)DE8090E9CD, (01)00884838082588(21)DE8090EBG7, (01)00884838082588(21)DE8090EDGW, (01)00884838082588(21)DE8090EDJF, (01)00884838082588(21)DE8090EDMY, (01)00884838082588(21)DE8090EDPV, (01)00884838082588(21)DE8090EDR0, (01)00884838082588(21)DE8090EFJP, (01)00884838082588(21)DE8090EG5Y, (01)00884838082588(21)DE8090EJJ6, (01)00884838082588(21)DE8090EJPW, (01)00884838082588(21)DE8090EK97, (01)00884838082588(21)DE8090EK9R, (01)00884838082588(21)DE8090EKF0, (01)00884838082588(21)DE8090EKH1, (01)00884838082588(21)DE8090EL5N, (01)00884838082588(21)DE8090EMCK, (01)00884838082588(21)DE8090EN03, (01)00884838082588(21)DE8090EN37, (01)00884838082588(21)DE8090EN58, (01)00884838082588(21)DE8090ENU1, (01)00884838082588(21)DE8090EP6D, (01)00884838082588(21)DE8090EPEE, (01)00884838082588(21)DE8090ERJB, (01)00884838082588(21)DE8090EV0A, (01)00884838082588(21)DE8090EVTL, (01)00884838082588(21)DE8090EVTZ, (01)00884838082588(21)DE8090EYDV, (01)00884838082588(21)DE8090EZ72, (01)00884838082588(21)DE8090EZUL, (01)00884838082588(21)DE8090F02H, (01)00884838082588(21)DE8090F0AJ, (01)00884838082588(21)DE8090F38V, (01)00884838082588(21)DE8090F3ZU, (01)00884838082588(21)DE8090F4EH, (01)00884838082588(21)DE8090F5CG, (01)00884838082588(21)DE8090F65E, (01)00884838082588(21)DE8090F780, (01)00884838082588(21)DE8090F7FL, (01)00884838082588(21)DE8090F840, (01)00884838082588(21)DE8090F9DR, (01)00884838082588(21)DE8090FAFC, (01)00884838082588(21)DE8090FAZZ, (01)00884838082588(21)DE8090FD14, (01)00884838082588(21)DE8090FFH6, (01)00884838082588(21)DE8090FGB7, (01)00884838082588(21)DE8090FH0A, (01)00884838082588(21)DE8090FHT6, (01)00884838082588(21)DE8090FKJ5, (01)00884838082588(21)DE8090FKUH, (01)00884838082588(21)DE8090FNFH, (01)00884838082588(21)DE8090FNFU, (01)00884838082588(21)DE8090FNR2, (01)00884838082588(21)DE8090FNW9, (01)00884838082588(21)DE8090FWCN, (01)00884838082588(21)DE8090FZNR, (01)00884838082588(21)DE8090G2G8, (01)00884838082588(21)DE8090G4JB, (01)00884838082588(21)DE8090G4L4, (01)00884838082588(21)DE8090G57E, (01)00884838082588(21)DE8090G58R, (01)00884838082588(21)DE8090G599, (01)00884838082588(21)DE8090G661, (01)00884838082588(21)DE8090G664, (01)00884838082588(21)DE8090G66M, (01)00884838082588(21)DE8090G6EK, (01)00884838082588(21)DE8090G9MB, (01)00884838082588(21)DE8090GAER, (01)00884838082588(21)DE8090GC6M, (01)00884838082588(21)DE8090GDCN, (01)00884838082588(21)DE8090GFTH, (01)00884838082588(21)DE8090GHNE, (01)00884838082588(21)DE8090GHW4, (01)00884838082588(21)DE8090GJ0C, (01)00884838082588(21)DE8090GN07, (01)00884838082588(21)DE8090GPJM, (01)00884838082588(21)DE8090GRFA, (01)00884838082588(21)DE8090GRVR, (01)00884838082588(21)DE8090GUD5, (01)00884838082588(21)DE8090GVUN, (01)00884838082588(21)DE8090GYPR, (01)00884838082588(21)DE8090GYPT, (01)00884838082588(21)DE8090GYRP, (01)00884838082588(21)DE8090GZN9, (01)00884838082588(21)DE8090H16Y, (01)00884838082588(21)DE8090H388, (01)00884838082588(21)DE8090H390, (01)00884838082588(21)DE8090H3K6.
2. Product/Model Number: 867033. UDI: (01)00884838082595(21)DE61120960, (01)00884838082595(21)DE61163620, (01)00884838082595(21)DE69481908, (01)00884838082595(21)DE69490222, (01)00884838082595(21)DE69498206, (01)00884838082595(21)DE752N9431, (01)00884838082595(21)DE752P4426, (01)00884838082595(21)DE752P5708, (01)00884838082595(21)DE752P9350, (01)00884838082595(21)DE752U1337, (01)00884838082595(21)DE761V3840, (01)00884838082595(21)DE761V9617, (01)00884838082595(21)DE8090EB5N.
3. Product/Model Number: 867036. UDI: (01)00884838086760(21)DE64107037, (01)00884838086760(21)DE64111363, (01)00884838086760(21)DE69513604, (01)00884838086760(21)DE69513635, (01)00884838086760(21)DE69513967, (01)00884838086760(21)DE69515319, (01)00884838086760(21)DE69515489, (01)00884838086760(21)DE69518525, (01)00884838086760(21)DE69519038, (01)00884838086760(21)DE69520426, (01)00884838086760(21)DE69520950, (01)00884838086760(21)DE69521640, (01)00884838086760(21)DE69523609, (01)00884838086760(21)DE69524730, (01)00884838086760(21)DE69524825, (01)00884838086760(21)DE69525232, (01)00884838086760(21)DE69525470, (01)00884838086760(21)DE69533993, (01)00884838086760(21)DE69534306, (01)00884838086760(21)DE69535734, (01)00884838086760(21)DE69552022, (01)00884838086760(21)DE69557256, (01)00884838086760(21)DE69557419, (01)00884838086760(21)DE69557720, (01)00884838086760(21)DE69560076, (01)00884838086760(21)DE69560144, (01)00884838086760(21)DE69561067, (01)00884838086760(21)DE69561933, (01)00884838086760(21)DE69562201, (01)00884838086760(21)DE69562377, (01)00884838086760(21)DE69562378, (01)00884838086760(21)DE69563388, (01)00884838086760(21)DE69563698, (01)00884838086760(21)DE69564264, (01)00884838086760(21)DE69564439, (01)00884838086760(21)DE69564518, (01)00884838086760(21)DE69564544, (01)00884838086760(21)DE69564546, (01)00884838086760(21)DE69564556, (01)00884838086760(21)DE69564613, (01)00884838086760(21)DE69564780, (01)00884838086760(21)DE69566055, (01)00884838086760(21)DE69566489, (01)00884838086760(21)DE69566532, (01)00884838086760(21)DE69566637, (01)00884838086760(21)DE69566650, (01)00884838086760(21)DE69567035, (01)00884838086760(21)DE69567313, (01)00884838086760(21)DE69567546, (01)00884838086760(21)DE69568457, (01)00884838086760(21)DE69568995, (01)00884838086760(21)DE69569023, (01)00884838086760(21)DE69570015, (01)00884838086760(21)DE69570463, (01)00884838086760(21)DE69571459, (01)00884838086760(21)DE69571462, (01)00884838086760(21)DE69571846, (01)00884838086760(21)DE69572229, (01)00884838086760(21)DE69574163, (01)00884838086760(21)DE69575698, (01)00884838086760(21)DE69576754, (01)00884838086760(21)DE69578100, (01)00884838086760(21)DE69580407, (01)00884838086760(21)DE69581475, (01)00884838086760(21)DE69582278, (01)00884838086760(21)DE69582350, (01)00884838086760(21)DE69582754, (01)00884838086760(21)DE69583581, (01)00884838086760(21)DE69584046, (01)00884838086760(21)DE69586342, (01)00884838086760(21)DE69586448, (01)00884838086760(21)DE69586687, (01)00884838086760(21)DE69586956, (01)00884838086760(21)DE69586996, (01)00884838086760(21)DE69587023, (01)00884838086760(21)DE69587279, (01)00884838086760(21)DE69587324, (01)00884838086760(21)DE69587510, (01)00884838086760(21)DE69587632, (01)00884838086760(21)DE69587882, (01)00884838086760(21)DE69587890, (01)00884838086760(21)DE69588170, (01)00884838086760(21)DE69588311, (01)00884838086760(21)DE69588695, (01)00884838086760(21)DE69590209, (01)00884838086760(21)DE69591418, (01)00884838086760(21)DE69591829, (01)00884838086760(21)DE69592038, (01)00884838086760(21)DE69592083, (01)00884838086760(21)DE69592188, (01)00884838086760(21)DE69592641, (01)00884838086760(21)DE69592642, (01)00884838086760(21)DE69592907, (01)00884838086760(21)DE69593063, (01)00884838086760(21)DE69593517, (01)00884838086760(21)DE69594637, (01)00884838086760(21)DE69595818, (01)00884838086760(21)DE69596381, (01)00884838086760(21)DE69596473, (01)00884838086760(21)DE69596837, (01)00884838086760(21)DE69597181, (01)00884838086760(21)DE69597612, (01)00884838086760(21)DE69597647, (01)00884838086760(21)DE69597688, (01)00884838086760(21)DE69597806, (01)00884838086760(21)DE69599083, (01)00884838086760(21)DE69599304, (01)00884838086760(21)DE69599580, (01)00884838086760(21)DE69599717, (01)00884838086760(21)DE69599778, (01)00884838086760(21)DE695A0178, (01)00884838086760(21)DE695A0905, (01)00884838086760(21)DE695A1311, (01)00884838086760(21)DE695A1606, (01)00884838086760(21)DE695A1639, (01)00884838086760(21)DE695A1727, (01)00884838086760(21)DE695A1800, (01)00884838086760(21)DE695A2975, (01)00884838086760(21)DE695A3752, (01)00884838086760(21)DE695A3771, (01)00884838086760(21)DE695A4571, (01)00884838086760(21)DE695A4685, (01)00884838086760(21)DE695A5277, (01)00884838086760(21)DE695A5926, (01)00884838086760(21)DE695A6208, (01)00884838086760(21)DE695A6581, (01)00884838086760(21)DE695B0233, (01)00884838086760(21)DE695B0405, (01)00884838086760(21)DE695B0600, (01)00884838086760(21)DE695B0987, (01)00884838086760(21)DE695B2598, (01)00884838086760(21)DE695B4161, (01)00884838086760(21)DE695B4366, (01)00884838086760(21)DE695B4626, (01)00884838086760(21)DE695B5155, (01)00884838086760(21)DE695B6046, (01)00884838086760(21)DE695B6759, (01)00884838086760(21)DE695B7172, (01)00884838086760(21)DE695B8876, (01)00884838086760(21)DE695B8938, (01)00884838086760(21)DE695B8994, (01)00884838086760(21)DE695C0396, (01)00884838086760(21)DE695C1857, (01)00884838086760(21)DE695C4709, (01)00884838086760(21)DE695C5256, (01)00884838086760(21)DE695C6360, (01)00884838086760(21)DE774C7481, (01)00884838086760(21)DE774C7659, (01)00884838086760(21)DE774C8730, (01)00884838086760(21)DE774D1616, (01)00884838086760(21)DE785D3192, (01)00884838086760(21)DE785D3824, (01)00884838086760(21)DE785D4855, (01)00884838086760(21)DE785D4905, (01)00884838086760(21)DE785D8566, (01)00884838086760(21)DE785E0000, (01)00884838086760(21)DE785E0004, (01)00884838086760(21)DE785E0007, (01)00884838086760(21)DE785E0586, (01)00884838086760(21)DE785E1664, (01)00884838086760(21)DE785E1796, (01)00884838086760(21)DE785E6888, (01)00884838086760(21)DE785E8091, (01)00884838086760(21)DE785F0906, (01)00884838086760(21)DE785F4128, (01)00884838086760(21)DE819G8281, (01)00884838086760(21)DE819H4428, (01)00884838086760(21)DE819H5930, (01)00884838086760(21)DE819H6717, (01)00884838086760(21)DE819H9330, (01)00884838086760(21)DE819J0394, (01)00884838086760(21)DE819J0481, (01)00884838086760(21)DE819J1165, (01)00884838086760(21)DE819K1467, (01)00884838086760(21)DE819M4001.
4. Product/Model Number: 867039. UDI: (01)00884838083547(21)DE61405483, (01)00884838083547(21)DE61406638, (01)00884838083547(21)DE61408585, (01)00884838083547(21)DE61413541, (01)00884838083547(21)DE61414637, (01)00884838083547(21)DE61417356, (01)00884838083547(21)DE61421061, (01)00884838083547(21)DE69736352, (01)00884838083547(21)DE69737825, (01)00884838083547(21)DE69740576, (01)00884838083547(21)DE69750125, (01)00884838083547(21)DE69755726, (01)00884838083547(21)DE69756194, (01)00884838083547(21)DE69774584, (01)00884838083547(21)DE74280785, (01)00884838083547(21)DE74290220, (01)00884838083547(21)DE74293488, (01)00884838083547(21)DE74296275, (01)00884838083547(21)DE74299246, (01)00884838083547(21)DE742A1829, (01)00884838083547(21)DE742A1840, (01)00884838083547(21)DE742A5733, (01)00884838083547(21)DE742A5736, (01)00884838083547(21)DE742A6827, (01)00884838083547(21)DE742B0937, (01)00884838083547(21)DE742B5206, (01)00884838083547(21)DE742B5213, (01)00884838083547(21)DE742B8583, (01)00884838083547(21)DE742C0817, (01)00884838083547(21)DE742D5180, (01)00884838083547(21)DE742E9411, (01)00884838083547(21)DE804H0176, (01)00884838083547(21)DE804H1643, (01)00884838083547(21)DE804H2307, (01)00884838083547(21)DE804H2351, (01)00884838083547(21)DE804H5546.
5. Product/Model Number: 867040. UDI: (01)00884838083578(21)DE69769643, (01)00884838083578(21)DE69770245, (01)00884838083578(21)DE69771185, (01)00884838083578(21)DE69772114, (01)00884838083578(21)DE74287979, (01)00884838083578(21)DE74291729, (01)00884838083578(21)DE742A4301, (01)00884838083578(21)DE742A4771, (01)00884838083578(21)DE742A4784, (01)00884838083578(21)DE742A6931, (01)00884838083578(21)DE742A8977, (01)00884838083578(21)DE742A9399, (01)00884838083578(21)DE742B6393, (01)00884838083578(21)DE742B9545, (01)00884838083578(21)DE742C4266, (01)00884838083578(21)DE742D2094, (01)00884838083578(21)DE742D3540, (01)00884838083578(21)DE742D3558, (01)00884838083578(21)DE742D8882, (01)00884838083578(21)DE742E1597, (01)00884838083578(21)DE742E1655, (01)00884838083578(21)DE742E4883, (01)00884838083578(21)DE742E5767, (01)00884838083578(21)DE742E6821, (01)00884838083578(21)DE742F5435, (01)00884838083578(21)DE742F7886, (01)00884838083578(21)DE742G6987, (01)00884838083578(21)DE804G9999, (01)00884838083578(21)DE804H0319, (01)00884838083578(21)DE804H0552, (01)00884838083578(21)DE804H0589, (01)00884838083578(21)DE804H0778, (01)00884838083578(21)DE804H0844, (01)00884838083578(21)DE804H2025, (01)00884838083578(21)DE804H2213, (01)00884838083578(21)DE804H3111, (01)00884838083578(21)DE804H3333, (01)00884838083578(21)DE804H3511, (01)00884838083578(21)DE804H3954, (01)00884838083578(21)DE804H5026, (01)00884838083578(21)DE804H6482, (01)00884838083578(21)DE804H7295, (01)00884838083578(21)DE804H8258.
6. Product/Model Number: 867041. UDI: (01)00884838083585(21)DE61402924, (01)00884838083585(21)DE61403144, (01)00884838083585(21)DE61403880, (01)00884838083585(21)DE61404556, (01)00884838083585(21)DE61404558, (01)00884838083585(21)DE61404559, (01)00884838083585(21)DE61404761, (01)00884838083585(21)DE61404775, (01)00884838083585(21)DE61405086, (01)00884838083585(21)DE61405116, (01)00884838083585(21)DE61405170, (01)00884838083585(21)DE61405556, (01)00884838083585(21)DE61405733, (01)00884838083585(21)DE61405837, (01)00884838083585(21)DE61406659, (01)00884838083585(21)DE61406688, (01)00884838083585(21)DE61407265, (01)00884838083585(21)DE61408523, (01)00884838083585(21)DE61409822, (01)00884838083585(21)DE61411272, (01)00884838083585(21)DE61412380, (01)00884838083585(21)DE61413596, (01)00884838083585(21)DE61415485, (01)00884838083585(21)DE61416226, (01)00884838083585(21)DE61416231, (01)00884838083585(21)DE61416236, (01)00884838083585(21)DE61416277, (01)00884838083585(21)DE61418085, (01)00884838083585(21)DE61418222, (01)00884838083585(21)DE61419364, (01)00884838083585(21)DE61420110, (01)00884838083585(21)DE61423547, (01)00884838083585(21)DE61423844, (01)00884838083585(21)DE61423974, (01)00884838083585(21)DE61425124, (01)00884838083585(21)DE61426201, (01)00884838083585(21)DE61426415, (01)00884838083585(21)DE61431781, (01)00884838083585(21)DE61431983, (01)00884838083585(21)DE61431999, (01)00884838083585(21)DE61432052, (01)00884838083585(21)DE61432428, (01)00884838083585(21)DE61432959, (01)00884838083585(21)DE61432965, (01)00884838083585(21)DE61433006, (01)00884838083585(21)DE69734400, (01)00884838083585(21)DE69736192, (01)00884838083585(21)DE69741310, (01)00884838083585(21)DE69744957, (01)00884838083585(21)DE69746346, (01)00884838083585(21)DE69746433, (01)00884838083585(21)DE69746703, (01)00884838083585(21)DE69747670, (01)00884838083585(21)DE69748004, (01)00884838083585(21)DE69748250, (01)00884838083585(21)DE69748662, (01)00884838083585(21)DE69749652, (01)00884838083585(21)DE69750430, (01)00884838083585(21)DE69753171, (01)00884838083585(21)DE69758677, (01)00884838083585(21)DE69758901, (01)00884838083585(21)DE69759066, (01)00884838083585(21)DE69759215, (01)00884838083585(21)DE69759268, (01)00884838083585(21)DE69759304, (01)00884838083585(21)DE69760294, (01)00884838083585(21)DE69760334, (01)00884838083585(21)DE69760723, (01)00884838083585(21)DE69761171, (01)00884838083585(21)DE69761597, (01)00884838083585(21)DE69764637, (01)00884838083585(21)DE69765058, (01)00884838083585(21)DE69765589, (01)00884838083585(21)DE69766117, (01)00884838083585(21)DE69766187, (01)00884838083585(21)DE69767119, (01)00884838083585(21)DE69767343, (01)00884838083585(21)DE69767512, (01)00884838083585(21)DE69767833, (01)00884838083585(21)DE69767853, (01)00884838083585(21)DE69768794, (01)00884838083585(21)DE69768821, (01)00884838083585(21)DE69775368, (01)00884838083585(21)DE69776125, (01)00884838083585(21)DE69776210, (01)00884838083585(21)DE69776440, (01)00884838083585(21)DE69777673, (01)00884838083585(21)DE69778003, (01)00884838083585(21)DE69778255, (01)00884838083585(21)DE69778833, (01)00884838083585(21)DE69779025, (01)00884838083585(21)DE69779375, (01)00884838083585(21)DE69779547, (01)00884838083585(21)DE69779639, (01)00884838083585(21)DE69779780, (01)00884838083585(21)DE74280218, (01)00884838083585(21)DE74280423, (01)00884838083585(21)DE74280597, (01)00884838083585(21)DE74280607, (01)00884838083585(21)DE74280632, (01)00884838083585(21)DE74281152, (01)00884838083585(21)DE74281850, (01)00884838083585(21)DE74282480, (01)00884838083585(21)DE74284368, (01)00884838083585(21)DE74285372, (01)00884838083585(21)DE74285545, (01)00884838083585(21)DE74285602, (01)00884838083585(21)DE74285604, (01)00884838083585(21)DE74285653, (01)00884838083585(21)DE74285696, (01)00884838083585(21)DE74285875, (01)00884838083585(21)DE74286415, (01)00884838083585(21)DE74286448, (01)00884838083585(21)DE74286482, (01)00884838083585(21)DE74287162, (01)00884838083585(21)DE74287170, (01)00884838083585(21)DE74287289, (01)00884838083585(21)DE74287325, (01)00884838083585(21)DE74288149, (01)00884838083585(21)DE74288275, (01)00884838083585(21)DE74288609, (01)00884838083585(21)DE74288634, (01)00884838083585(21)DE74289018, (01)00884838083585(21)DE74289024, (01)00884838083585(21)DE74289348, (01)00884838083585(21)DE74289674, (01)00884838083585(21)DE74290204, (01)00884838083585(21)DE74290356, (01)00884838083585(21)DE74290640, (01)00884838083585(21)DE74290651, (01)00884838083585(21)DE74291433, (01)00884838083585(21)DE74292138, (01)00884838083585(21)DE74292593, (01)00884838083585(21)DE74293402, (01)00884838083585(21)DE74293779, (01)00884838083585(21)DE74294321, (01)00884838083585(21)DE74294324, (01)00884838083585(21)DE74294361, (01)00884838083585(21)DE74294464, (01)00884838083585(21)DE74294635, (01)00884838083585(21)DE74295307, (01)00884838083585(21)DE74297519, (01)00884838083585(21)DE74297520, (01)00884838083585(21)DE74297530, (01)00884838083585(21)DE74298275, (01)00884838083585(21)DE74298317, (01)00884838083585(21)DE74298409, (01)00884838083585(21)DE74298751, (01)00884838083585(21)DE74298944, (01)00884838083585(21)DE74299036, (01)00884838083585(21)DE74299712, (01)00884838083585(21)DE74299781, (01)00884838083585(21)DE742A0244, (01)00884838083585(21)DE742A0523, (01)00884838083585(21)DE742A0916, (01)00884838083585(21)DE742A0920, (01)00884838083585(21)DE742A1150, (01)00884838083585(21)DE742A1373, (01)00884838083585(21)DE742A1692, (01)00884838083585(21)DE742A1852, (01)00884838083585(21)DE742A3171, (01)00884838083585(21)DE742A3688, (01)00884838083585(21)DE742A4050, (01)00884838083585(21)DE742A4432, (01)00884838083585(21)DE742A4496, (01)00884838083585(21)DE742A4576, (01)00884838083585(21)DE742A5351, (01)00884838083585(21)DE742A5466, (01)00884838083585(21)DE742A5540, (01)00884838083585(21)DE742A5773, (01)00884838083585(21)DE742A7589, (01)00884838083585(21)DE742A7787, (01)00884838083585(21)DE742A7952, (01)00884838083585(21)DE742A8361, (01)00884838083585(21)DE742A8668, (01)00884838083585(21)DE742A8749, (01)00884838083585(21)DE742A9577, (01)00884838083585(21)DE742A9691, (01)00884838083585(21)DE742B1364, (01)00884838083585(21)DE742B1506, (01)00884838083585(21)DE742B1513, (01)00884838083585(21)DE742B1582, (01)00884838083585(21)DE742B2124, (01)00884838083585(21)DE742B2950, (01)00884838083585(21)DE742B3376, (01)00884838083585(21)DE742B3443, (01)00884838083585(21)DE742B3593, (01)00884838083585(21)DE742B4349, (01)00884838083585(21)DE742B5114, (01)00884838083585(21)DE742B5130, (01)00884838083585(21)DE742B5885, (01)00884838083585(21)DE742B6021, (01)00884838083585(21)DE742B6022, (01)00884838083585(21)DE742B8445, (01)00884838083585(21)DE742B8817, (01)00884838083585(21)DE742B8985, (01)00884838083585(21)DE742B9030, (01)00884838083585(21)DE742B9069, (01)00884838083585(21)DE742B9339, (01)00884838083585(21)DE742B9903, (01)00884838083585(21)DE742B9933, (01)00884838083585(21)DE742C0098, (01)00884838083585(21)DE742C0239, (01)00884838083585(21)DE742C1172, (01)00884838083585(21)DE742C1222, (01)00884838083585(21)DE742C1400, (01)00884838083585(21)DE742C1512, (01)00884838083585(21)DE742C1517, (01)00884838083585(21)DE742C1560, (01)00884838083585(21)DE742C1862, (01)00884838083585(21)DE742C1903, (01)00884838083585(21)DE742C2178, (01)00884838083585(21)DE742C2904, (01)00884838083585(21)DE742C3042, (01)00884838083585(21)DE742C5741, (01)00884838083585(21)DE742C6194, (01)00884838083585(21)DE742C6583, (01)00884838083585(21)DE742C6623, (01)00884838083585(21)DE742C6881, (01)00884838083585(21)DE742C7956, (01)00884838083585(21)DE742C8076, (01)00884838083585(21)DE742C9991, (01)00884838083585(21)DE742D0271, (01)00884838083585(21)DE742D0554, (01)00884838083585(21)DE742D1316, (01)00884838083585(21)DE742D3731, (01)00884838083585(21)DE742D4874, (01)00884838083585(21)DE742D7453, (01)00884838083585(21)DE742D7484, (01)00884838083585(21)DE742D7560, (01)00884838083585(21)DE742D7926, (01)00884838083585(21)DE742E0678, (01)00884838083585(21)DE742E1015, (01)00884838083585(21)DE742E1222, (01)00884838083585(21)DE742E1294, (01)00884838083585(21)DE742E1707, (01)00884838083585(21)DE742E2970, (01)00884838083585(21)DE742E3247, (01)00884838083585(21)DE742E3914, (01)00884838083585(21)DE742E4005, (01)00884838083585(21)DE742E4120, (01)00884838083585(21)DE742E4124, (01)00884838083585(21)DE742E5347, (01)00884838083585(21)DE742E5998, (01)00884838083585(21)DE742E6320, (01)00884838083585(21)DE742E6533, (01)00884838083585(21)DE742E7876, (01)00884838083585(21)DE742E8817, (01)00884838083585(21)DE742E9729, (01)00884838083585(21)DE742E9957, (01)00884838083585(21)DE742F0741, (01)00884838083585(21)DE742F0800, (01)00884838083585(21)DE742F2101, (01)00884838083585(21)DE742F2467, (01)00884838083585(21)DE742F2873, (01)00884838083585(21)DE742F3166, (01)00884838083585(21)DE742F5536, (01)00884838083585(21)DE742F5954, (01)00884838083585(21)DE742F8116, (01)00884838083585(21)DE742F8441, (01)00884838083585(21)DE742F9262, (01)00884838083585(21)DE742F9743, (01)00884838083585(21)DE742F9768, (01)00884838083585(21)DE742F9790, (01)00884838083585(21)DE742G0061, (01)00884838083585(21)DE742G0090, (01)00884838083585(21)DE742G0481, (01)00884838083585(21)DE742G0645, (01)00884838083585(21)DE742G1653, (01)00884838083585(21)DE742G1709, (01)00884838083585(21)DE742G1778, (01)00884838083585(21)DE742G2287, (01)00884838083585(21)DE742G3594, (01)00884838083585(21)DE742G3689, (01)00884838083585(21)DE742G4295, (01)00884838083585(21)DE742G4505, (01)00884838083585(21)DE742G5982, (01)00884838083585(21)DE742G6103, (01)00884838083585(21)DE742G6452, (01)00884838083585(21)DE742G6573, (01)00884838083585(21)DE742G6712, (01)00884838083585(21)DE742G6811, (01)00884838083585(21)DE806G7939, (01)00884838083585(21)DE806G7998, (01)00884838083585(21)DE806G8715, (01)00884838083585(21)DE806G9025, (01)00884838083585(21)DE806G9566, (01)00884838083585(21)DE806G9723, (01)00884838083585(21)DE806H1574, (01)00884838083585(21)DE806H2083, (01)00884838083585(21)DE806H3565, (01)00884838083585(21)DE806H3692, (01)00884838083585(21)DE806H5137, (01)00884838083585(21)DE806H5768, (01)00884838083585(21)DE806H5792, (01)00884838083585(21)DE806H5803, (01)00884838083585(21)DE806H6245, (01)00884838083585(21)DE806H6266, (01)00884838083585(21)DE806H6453, (01)00884838083585(21)DE806H6585, (01)00884838083585(21)DE806H6681, (01)00884838083585(21)DE806H7093, (01)00884838083585(21)DE806H7332, (01)00884838083585(21)DE806H7504, (01)00884838083585(21)DE806H7881, (01)00884838083585(21)DE806H7943, (01)00884838083585(21)DE806J0228, (01)00884838083585(21)DE806J0251, (01)00884838083585(21)DE806J0511, (01)00884838083585(21)DE806J0551, (01)00884838083585(21)DE806J1031, (01)00884838083585(21)DE806J1671, (01)00884838083585(21)DE806J2495, (01)00884838083585(21)DE806J3204, (01)00884838083585(21)DE806J3551, (01)00884838083585(21)DE806J3579, (01)00884838083585(21)DE806J3622, (01)00884838083585(21)DE806J4569, (01)00884838083585(21)DE806J5058, (01)00884838083585(21)DE806J5411, (01)00884838083585(21)DE806J5659, (01)00884838083585(21)DE806J5676, (01)00884838083585(21)DE806J5712, (01)00884838083585(21)DE806J6114, (01)00884838083585(21)DE806J6267, (01)00884838083585(21)DE806J6639, (01)00884838083585(21)DE806J6780, (01)00884838083585(21)DE806J6789, (01)00884838083585(21)DE806J6793, (01)00884838083585(21)DE806J6911, (01)00884838083585(21)DE806J6927, (01)00884838083585(21)DE806J7053, (01)00884838083585(21)DE806J7450, (01)00884838083585(21)DE806J7989, (01)00884838083585(21)DE806J8073, (01)00884838083585(21)DE806J9434, (01)00884838083585(21)DE806J9697, (01)00884838083585(21)DE806J9871, (01)00884838083585(21)DE806K0554, (01)00884838083585(21)DE806K2595, (01)00884838083585(21)DE806K2809, (01)00884838083585(21)DE806K3414, (01)00884838083585(21)DE806K4277, (01)00884838083585(21)DE806K5425, (01)00884838083585(21)DE806K5775, (01)00884838083585(21)DE806K5858, (01)00884838083585(21)DE806K6324, (01)00884838083585(21)DE806K7160, (01)00884838083585(21)DE806K7285, (01)00884838083585(21)DE806K7396, (01)00884838083585(21)DE806K7560, (01)00884838083585(21)DE806K7567, (01)00884838083585(21)DE806K7695, (01)00884838083585(21)DE806K7822, (01)00884838083585(21)DE806K8472, (01)00884838083585(21)DE806K8836, (01)00884838083585(21)DE806K9657, (01)00884838083585(21)DE806K9658, (01)00884838083585(21)DE806L0155, (01)00884838083585(21)DE806L0216, (01)00884838083585(21)DE806L0494, (01)00884838083585(21)DE806L0519, (01)00884838083585(21)DE806L0531, (01)00884838083585(21)DE806L0550, (01)00884838083585(21)DE806L0587, (01)00884838083585(21)DE806L0802, (01)00884838083585(21)DE806L0817, (01)00884838083585(21)DE806L2083, (01)00884838083585(21)DE806L2155, (01)00884838083585(21)DE806L2196, (01)00884838083585(21)DE806L2354, (01)00884838083585(21)DE806L2410, (01)00884838083585(21)DE806L2624, (01)00884838083585(21)DE806L2648, (01)00884838083585(21)DE806L2658, (01)00884838083585(21)DE806L2682, (01)00884838083585(21)DE806L2693, (01)00884838083585(21)DE806L2745, (01)00884838083585(21)DE806L3945, (01)00884838083585(21)DE806L3951, (01)00884838083585(21)DE806L3964, (01)00884838083585(21)DE806L3967, (01)00884838083585(21)DE806L3969, (01)00884838083585(21)DE806L4460, (01)00884838083585(21)DE806L4473, (01)00884838083585(21)DE806L4478, (01)00884838083585(21)DE806L4481, (01)00884838083585(21)DE806L4975, (01)00884838083585(21)DE806L5464, (01)00884838083585(21)DE806L5503, (01)00884838083585(21)DE806L7108, (01)00884838083585(21)DE806L7225, (01)00884838083585(21)DE806L7355, (01)00884838083585(21)DE806L7378, (01)00884838083585(21)DE806L7385, (01)00884838083585(21)DE806L7601, (01)00884838083585(21)DE806L8287, (01)00884838083585(21)DE806L8398, (01)00884838083585(21)DE806L8540, (01)00884838083585(21)DE806L8604, (01)00884838083585(21)DE806L8606, (01)00884838083585(21)DE806L9440, (01)00884838083585(21)DE806M1426, (01)00884838083585(21)DE806M4260, (01)00884838083585(21)DE806M4261, (01)00884838083585(21)DE806M4329, (01)00884838083585(21)DE806M5257, (01)00884838083585(21)DE806M5260, (01)00884838083585(21)DE806M5262, (01)00884838083585(21)DE806M5301, (01)00884838083585(21)DE806M5317, (01)00884838083585(21)DE806M5331, (01)00884838083585(21)DE806M5332, (01)00884838083585(21)DE806M6459, (01)00884838083585(21)DE806M7751, (01)00884838083585(21)DE806M9224.
|
| FEI Number |
3006648320
|
Recalling Firm/ Manufacturer |
Philips North America
222 Jacobs St
Cambridge
MA
02141-2296
|
| For Additional Information Contact | Philips Customer Services 800-722-9377 |
Manufacturer Reason for Recall | Potential for the Dual IBP Adapter to fail to maintain
defibrillation protection if two non-defibrillation proof transducers are connected simultaneously. |
FDA Determined Cause 2 | Device Design |
| Action | Philips North America notified consignees on about 10/16/2026 via letter titled "URGENT Medical Device Correction." Consignees were informed of the issue and hazard involved. They were instructed to ensure that compatible disposable pressure transducers and accessories (per IFU) are used in conjunction with Dual IBP Adapter, provide the notification letter to all those who need to be aware within each organization or to any organization where affected product(s) have been potentially transferred, place the notification with the documentation of the Philips Dual IBP Adapter and associated devices in a place where it is most likely to be seen and viewed, and complete and return the provided URGENT Medical Device Correction Response Form.
Distributors were instructed to send the provided URGENT Medical Device Correction to each customer to whom they have distributed any affected product(s) as soon as possible and no later than 5 days, include their contact information in the response form, perform a good faith effort to get the completed Response Form from each customer by following up a minimum of three (3) times, and if possible, using multiple contact methods, and inform Philips about the responses received.
A Philips representative will send a follow up communication to consignees to provide instructions on how to obtain updated Instructions for Use (IFU) and Addendum once available. |
| Quantity in Commerce | 290,291 units |
| Distribution | US Nationwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|