| Date Initiated by Firm | October 14, 2026 |
| Date Posted | November 12, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0431-2027 |
| Recall Event ID |
100163 |
| 510(K)Number | K241572 K254173 |
| Product Classification |
Interventional fluoroscopic x-ray system - Product Code OWB
|
| Product | ARTIS icono floor. Model Number: 11327700. |
| Code Information |
Model Number: 11327700. UDI-DI: 4056869149325. Serial Number: 170611. Software Versions VE40B and VE50B.
|
| FEI Number |
2240869
|
Recalling Firm/ Manufacturer |
Siemens Medical Solutions USA, Inc
40 Liberty Blvd
Malvern
PA
19355-1418
|
| For Additional Information Contact | Anastasia Sokolova 610-219-4834 |
Manufacturer Reason for Recall | Potential for loss of X-ray functionality after system startup or restart. |
FDA Determined Cause 2 | Software design |
| Action | Siemens notified consignees on about 10/14/2026 via letter titled "CUSTOMER ADVISORY NOTICE." Consignees were instructed to consider the potential for unavailability following startup or restart when planning the clinical use of affected systems. For critical, urgent, or time-sensitive procedures, healthcare professionals should determine in advance whether an affected system is appropriate for the planned procedure and consider established contingency measures, including the availability of an alternative system or treatment pathway. These precautions should remain in place until the corrective action has been completed on the affected system.
Siemens Healthineers is implementing Field Safety Corrective Action (FSCA) for affected systems. The corrective action includes a replacement of an affected component. The corrective action will be performed by Siemens Service Organization. This measure is intended to restore the affected systems to a configuration in which the described startup or restart issue is no longer expected to occur.
The service organization will contact consignees to schedule an appointment to perform corrective action. The corrective action is being distributed through Update Instruction AX047/26/S. If an earlier appointment is preferred, please contact your local Siemens Healthineers service organization.
Siemens is not recommending re-examination of patients.
Consignees were also requested to ensure all users of affected units are made aware of the issue and requested that consignees complete and return the response form provided. |
| Quantity in Commerce | 1 unit |
| Distribution | US Nationwide distribution in the states of CA, IA, KY, MN, NY, OH. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|