| Date Initiated by Firm | September 22, 2026 |
| Date Posted | November 25, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0561-2027 |
| Recall Event ID |
100226 |
| Product Classification |
General surgery tray - Product Code LRO
|
| Product | EXTREMITY BURN ADD-ON;
Model/Catalog Number: JACK338-03;
Software Version: N/A;
Product Description: Convenience Kit;
Component: No; |
| Code Information |
Model/Catalog Number: JACK338-03;
UDI: 10809160386194;
Expiration Date: 8/5/2028;
Lot Number: 1692090; |
| FEI Number |
1047429
|
Recalling Firm/ Manufacturer |
AVID Medical, Inc.
9000 Westmont Dr
Toano
VA
23168-9351
|
| For Additional Information Contact | Marlene Jones +1-470-2804277 |
Manufacturer Reason for Recall | The honeycomb bandage within the kit was provided in its original packaging from the manufacturer instead of being removed prior to sterilization. The product may not be sterile as it was determined that the component is required to be stripped from manufacturer s packaging prior to sterilization.
|
FDA Determined Cause 2 | Process control |
| Action | On September 22, 2026 URGENT: MEDICAL DEVICE FIELD CORRECTION letters were sent to customers.
ACTIONS TO TAKE:
1. Identify, segregate, and quarantine all impacted kits identified above.
2. DO NOT USE and DISCARD the Bandage Honeycomb component that is located in the clear plastic wrap in the kit identified (refer to Figure 1). The unwrapped components within the kit may continue to be used.
3. Add warning labels to affected AVID kits to ensure all users are aware of the actions to remove and discard the Bandage Honeycomb that contains the original clear plastic packaging. A labeling template is enclosed for printing labels which can be affixed on all affected kits and cases in inventory. This label should be applied in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. Please contact your Sales Representative if pre-printed labels are needed.
- Customers (end-users): If you have received product from a distributor, please contact the distributor or your sales representative with any questions you may have.
- Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness.
4. Complete the attached Recall Response Form and return by October 9, 2026. |
| Quantity in Commerce | 48 |
| Distribution | US Nationwide distribution in the state of FL. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|