| Date Initiated by Firm | September 16, 2026 |
| Date Posted | November 25, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0551-2027 |
| Recall Event ID |
100227 |
| 510(K)Number | K173395 K213081 |
| Product Classification |
System, x-ray, fluoroscopic, image-intensified - Product Code JAA
|
| Product | Discovery RF180GMM equipped with Thales RF or FL4 detectors |
| Code Information |
Serial Number 01129 and Serial Number 01946 |
| FEI Number |
3002807092
|
Recalling Firm/ Manufacturer |
General Medical Merate S.p.A.
Via Partigiani 25
Seriate
Italy
|
Manufacturer Reason for Recall | GMM's Discovery RF180GMM equipped with Thales RF or FL4 detectors, system software does not display the air kerma rate (AKR) when the device is operated in a specific setting using detector's Logical Mode 11 at an image acquisition frequency greater than six images per second. |
FDA Determined Cause 2 | Radiation Control for Health and Safety Act |
| Action | GMM notified customers via letter dated September 16, 2026, providing Description of the issue, Risk associated with the issue, Affected products, Actions to prevent risk to users/patients, and Action planned for correction of the issue. |
| Quantity in Commerce | 470 |
| Distribution | U.S. Nationwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = JAA
|