 Class 2 Device Recall Discovery RF180see related information
Date Initiated by FirmSeptember 16, 2026
Date PostedNovember 25, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0551-2027
Recall Event ID 100227
510(K)NumberK173395 K213081 
Product Classification System, x-ray, fluoroscopic, image-intensified - Product Code JAA
ProductDiscovery RF180GMM equipped with Thales RF or FL4 detectors
Code Information Serial Number 01129 and Serial Number 01946
FEI Number 3002807092
Recalling Firm/
Manufacturer
General Medical Merate S.p.A.
Via Partigiani 25
Seriate Italy
Manufacturer Reason
for Recall
GMM's Discovery RF180GMM equipped with Thales RF or FL4 detectors, system software does not display the air kerma rate (AKR) when the device is operated in a specific setting using detector's Logical Mode 11 at an image acquisition frequency greater than six images per second.
FDA Determined
Cause 2
Radiation Control for Health and Safety Act
ActionGMM notified customers via letter dated September 16, 2026, providing Description of the issue, Risk associated with the issue, Affected products, Actions to prevent risk to users/patients, and Action planned for correction of the issue.
Quantity in Commerce470
DistributionU.S. Nationwide distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.