• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Photonova Spectra, Photonova Spectra Select

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Photonova Spectra, Photonova Spectra Selectsee related information
Date Initiated by FirmSeptember 21, 2026
Date PostedNovember 25, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0312-2027
Recall Event ID 100233
510(K)NumberK253520 
Product Classification System, x-ray, tomography, computed - Product Code JAK
ProductThe system is intended for head, whole body, cardiac and vascular X-ray Computed Tomography applications.
Code Information Serial: PCT8X2600003CN. GTIN: 00198953085950
FEI Number 2126677
Recalling Firm/
Manufacturer
GE Medical Systems, LLC
3000 N Grandview Blvd
Waukesha WI 53188-1615
For Additional Information Contact
262-544-3011
Manufacturer Reason
for Recall
During use of certain service tools (Detailed Calibration, Geometry Calibration, Phantom Centering, or Detector Performance), a sequence of events including scan aborts and pressing the Setup button, may allow initiation of X-ray scan without a physical press of the Start Scan button.
FDA Determined
Cause 2
Radiation Control for Health and Safety Act
ActionGE HealthCare (will) send a customer letter dated September 21, 2026, that provides instructions for immediate mitigation of the uncommanded X-ray occurrences during servicing/calibration events. This NOD (notification of defect) customer letter will be sent to the affected site in the US.
Quantity in Commerce2
DistributionWorldwide - US Nationwide distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = JAK
-
-