| | Class 2 Device Recall Photonova Spectra, Photonova Spectra Select |  |
| Date Initiated by Firm | September 21, 2026 |
| Date Posted | November 25, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0312-2027 |
| Recall Event ID |
100233 |
| 510(K)Number | K253520 |
| Product Classification |
System, x-ray, tomography, computed - Product Code JAK
|
| Product | The system is intended for head, whole body,
cardiac and vascular X-ray Computed Tomography applications. |
| Code Information |
Serial: PCT8X2600003CN. GTIN: 00198953085950 |
| FEI Number |
2126677
|
Recalling Firm/ Manufacturer |
GE Medical Systems, LLC
3000 N Grandview Blvd
Waukesha
WI
53188-1615
|
| For Additional Information Contact | 262-544-3011 |
Manufacturer Reason for Recall | During use of certain service tools (Detailed Calibration, Geometry Calibration, Phantom Centering, or Detector Performance), a sequence of events including scan aborts and pressing the Setup button, may allow initiation of X-ray scan without a physical press of the Start Scan button. |
FDA Determined Cause 2 | Radiation Control for Health and Safety Act |
| Action | GE HealthCare (will) send a customer letter dated September 21, 2026, that provides instructions for immediate mitigation
of the uncommanded X-ray occurrences during servicing/calibration events. This NOD
(notification of defect) customer letter will be sent to the affected site in the US. |
| Quantity in Commerce | 2 |
| Distribution | Worldwide - US Nationwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|