| | Class 2 Device Recall GLP Centrifuge Module |  |
| Date Initiated by Firm | October 21, 2026 |
| Date Posted | November 27, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0570-2027 |
| Recall Event ID |
100240 |
| 510(K)Number | K213486 |
| Product Classification |
Calculator/data processing module, for clinical use - Product Code JQP
|
| Product | GLP Centrifuge Module For use with GLP systems Track, List Numbers 06Q03-01 and 06Q03-51 |
| Code Information |
LN 06Q03-01: UDI-DI: 00380740186647, Serial Numbers: M05A010302, M05A010711;
LN 06Q03-51: UDI-DI: 00380740208691, Serial Numbers: M20A010265, M20A010484, M20A010485, M20A010486, M20A010487, M20A010523, M20A010527, M20A010530, M20A010541, M20A010556, M20A010557, M20A010562, M20A010563, M20A010566, M20A010572, M20A010573, M20A010549, M20A010552 |
| FEI Number |
1415939
|
Recalling Firm/ Manufacturer |
Abbott Laboratories
100 Abbott Park Rd Bldg Ap8b
Abbott Park
IL
60064-3502
|
| For Additional Information Contact | Suchin Song 224-668-0343 |
Manufacturer Reason for Recall | Certain GLP Centrifuge Modules containing a GLP Centrifuge updated to use R-290 (refrigerant-grade propane), were placed into service without the required CM Upgrade Kit for Coolant (ventilation system kit). Without the kit, if a propane leak occurs, propane could accumulate within the centrifuge enclosure near energized electrical components, creating a potential fire or explosion hazard. The issue is limited to modules with serial numbers below M20A011000 that were powered on without the upgrade kit installed. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | An Urgent Medical Device Correction dated 10/21/26 was sent to customers.
An Abbott representative will contact you to discuss a service resolution.
Necessary Actions to be Taken by Customer
If there is an increase in temperature out of range errors from your Centrifuge Module potentially arising from a refrigerant leak, follow the Emergency shutdown procedure found in Section 5 of the GLP systems Track Operations Manual and contact your local Abbott Customer Service representative.
Complete and return the Customer Reply Form.
If you have forwarded the product listed above to other laboratories, please inform them of this product correction and provide them with a copy of this letter.
Please retain this letter for your laboratory records.
Contact Information
If you or any of the health care providers you serve have questions regarding this information, please contact Customer Service at 1-877-4ABBOTT. |
| Quantity in Commerce | 20 units |
| Distribution | Worldwide - US Nationwide distribution in the states of CA, CO, WI, and TX. The countries of Czech Republic, Indonesia, Poland, Portugal, Spain, UK. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|