 Class 2 Device Recall GLP Centrifuge Modulesee related information
Date Initiated by FirmOctober 21, 2026
Date PostedNovember 27, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0570-2027
Recall Event ID 100240
510(K)NumberK213486 
Product Classification Calculator/data processing module, for clinical use - Product Code JQP
ProductGLP Centrifuge Module For use with GLP systems Track, List Numbers 06Q03-01 and 06Q03-51
Code Information LN 06Q03-01: UDI-DI: 00380740186647, Serial Numbers: M05A010302, M05A010711; LN 06Q03-51: UDI-DI: 00380740208691, Serial Numbers: M20A010265, M20A010484, M20A010485, M20A010486, M20A010487, M20A010523, M20A010527, M20A010530, M20A010541, M20A010556, M20A010557, M20A010562, M20A010563, M20A010566, M20A010572, M20A010573, M20A010549, M20A010552
FEI Number 1415939
Recalling Firm/
Manufacturer
Abbott Laboratories
100 Abbott Park Rd Bldg Ap8b
Abbott Park IL 60064-3502
For Additional Information ContactSuchin Song
224-668-0343
Manufacturer Reason
for Recall
Certain GLP Centrifuge Modules containing a GLP Centrifuge updated to use R-290 (refrigerant-grade propane), were placed into service without the required CM Upgrade Kit for Coolant (ventilation system kit). Without the kit, if a propane leak occurs, propane could accumulate within the centrifuge enclosure near energized electrical components, creating a potential fire or explosion hazard. The issue is limited to modules with serial numbers below M20A011000 that were powered on without the upgrade kit installed.
FDA Determined
Cause 2
Under Investigation by firm
ActionAn Urgent Medical Device Correction dated 10/21/26 was sent to customers. An Abbott representative will contact you to discuss a service resolution. Necessary Actions to be Taken by Customer If there is an increase in temperature out of range errors from your Centrifuge Module potentially arising from a refrigerant leak, follow the Emergency shutdown procedure found in Section 5 of the GLP systems Track Operations Manual and contact your local Abbott Customer Service representative. Complete and return the Customer Reply Form. If you have forwarded the product listed above to other laboratories, please inform them of this product correction and provide them with a copy of this letter. Please retain this letter for your laboratory records. Contact Information If you or any of the health care providers you serve have questions regarding this information, please contact Customer Service at 1-877-4ABBOTT.
Quantity in Commerce20 units
DistributionWorldwide - US Nationwide distribution in the states of CA, CO, WI, and TX. The countries of Czech Republic, Indonesia, Poland, Portugal, Spain, UK.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.