| | Class 1 Device Recall Arrow |  |
| Date Initiated by Firm | November 05, 2026 |
| Date Posted | November 25, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0498-2027 |
| Recall Event ID |
100262 |
| 510(K)Number | K112896 |
| Product Classification |
Catheter, intravascular, therapeutic, long-term greater than 30 days - Product Code LJS
|
| Product | Pressure Injectable Arrowg+ard Blue Advance One-Lumen PICC Kit pre-loaded with Arrow VPS Precision Stylet, REF: ASK-45541-CHS;
Pressure Injectable Arrowg+ard Blue Advance Two-Lumen PICC Kit pre-loaded with Arrow VPS Precision Stylet, REF: ASK-45552-CHS;
Pressure Injectable Arrowg+ard Blue Advance One-Lumen PICC with 130 cm Hydrophilic-Coated Nitinol Guidewire, REF: PR-45541-HPHNL |
| Code Information |
REF/UDI-DI/Lot:
ASK-45541-CHS/10801902223477/33F26A1422;
ASK-45552-CHS/10801902223484/33F26A1395, 33F26G0411;
PR-45541-HPHNL/10801902215786, 10801902188516/33F25B0399, 33F25G0030, 33F26A0137, 33F26B0738, 33F26C0475, 33F26D0315, 33F26E0189, 33F26F0344, 33F26G0135;
|
| FEI Number |
3013162291
|
Recalling Firm/ Manufacturer |
ARROW INTERNATIONAL, LLC
3015 Carrington Mill Blvd
Morrisville
NC
27560-5437
|
| For Additional Information Contact | Customer Service 866-396-2111 |
Manufacturer Reason for Recall | Intravascular, peripheral, central catheters and guidewires have hydrophilic coated spring wire guide components with coating that may swell inconsistently when hydrated with saline leading to coating clumping and potential for detachment, which may lead to particle release into the blood stream, which may result in inflammation, thrombosis, ischemia, tissue injury, stroke, further intervention. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | On 11/5/2026, "URGENT MEDICAL DEVICE RECALL" notices were mailed and emailed to customers who were asked to do the following:
1) Cease use, distribution, and quarantine affected product.
2) Complete and return the acknowledgement form via email to E-mail: Recalls@teleflex.com
In addition, distributors were asked to do the following:
a) Provide a copy of this recall notice to all customers who have received impacted product. Each of your customers is then required to complete the Acknowledgement Form and return it to you as their distributor.
b) If you have further distributed product outside of your country, please notify the recalling firm.
Questions can be directed to the firm's Customer service, 1-866-396-2111, or Recalls@teleflex.com
FSN Portal Link: https://fca.myteleflex.com/en/recall/000614 |
| Quantity in Commerce | 1,390 |
| Distribution | Worldwide Distribution: (US nationwide) CA, SC, GA, AZ, CO, MO, LA, AR, TX, UT, MT, IA, MA, AL, WY, NJ, VA, PA, WV, DE, OH, NE, IN, FL, TN, NY, NH, MS, MD, MI, KS, NM, KY, OK, ME, NC, ID, WA, HI, NV, RI, OR, VT, IL, WI, SD, MN, DC, CT; (OUS countries of): Australia, Brunei, Canada, Chile, China, Colombia, India, Japan, South Korea, Mexico, Malaysia, Nepal, New Zealand, Philippines, Singapore, Thailand, Taiwan, and Vietnam. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = LJS
|
|
|
|