• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall Arrow

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall Arrowsee related information
Date Initiated by FirmNovember 05, 2026
Date PostedNovember 25, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0502-2027
Recall Event ID 100262
510(K)NumberK122545 
Product Classification Catheter, intravascular, therapeutic, long-term greater than 30 days - Product Code LJS
ProductPressure Injectable Arrowg+ard Blue Advance Two-Lumen PICC Kit pre-loaded with Arrow VPS Precision Stylet, REF: ASK-45052-MC
Code Information REF/UDI-DI/Lot: ASK-45052-MC/10801902223347/33F25G0463, 33F26D0568
FEI Number 3013162291
Recalling Firm/
Manufacturer
ARROW INTERNATIONAL, LLC
3015 Carrington Mill Blvd
Morrisville NC 27560-5437
For Additional Information ContactCustomer Service
866-396-2111
Manufacturer Reason
for Recall
Intravascular, peripheral, central catheters and guidewires have hydrophilic coated spring wire guide components with coating that may swell inconsistently when hydrated with saline leading to coating clumping and potential for detachment, which may lead to particle release into the blood stream, which may result in inflammation, thrombosis, ischemia, tissue injury, stroke, further intervention.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionOn 11/5/2026, "URGENT MEDICAL DEVICE RECALL" notices were mailed and emailed to customers who were asked to do the following: 1) Cease use, distribution, and quarantine affected product. 2) Complete and return the acknowledgement form via email to E-mail: Recalls@teleflex.com In addition, distributors were asked to do the following: a) Provide a copy of this recall notice to all customers who have received impacted product. Each of your customers is then required to complete the Acknowledgement Form and return it to you as their distributor. b) If you have further distributed product outside of your country, please notify the recalling firm. Questions can be directed to the firm's Customer service, 1-866-396-2111, or Recalls@teleflex.com FSN Portal Link: https://fca.myteleflex.com/en/recall/000614
Quantity in Commerce180
DistributionWorldwide Distribution: (US nationwide) CA, SC, GA, AZ, CO, MO, LA, AR, TX, UT, MT, IA, MA, AL, WY, NJ, VA, PA, WV, DE, OH, NE, IN, FL, TN, NY, NH, MS, MD, MI, KS, NM, KY, OK, ME, NC, ID, WA, HI, NV, RI, OR, VT, IL, WI, SD, MN, DC, CT; (OUS countries of): Australia, Brunei, Canada, Chile, China, Colombia, India, Japan, South Korea, Mexico, Malaysia, Nepal, New Zealand, Philippines, Singapore, Thailand, Taiwan, and Vietnam.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = LJS
-
-