| | Class 1 Device Recall Medline |  |
| Date Initiated by Firm | October 14, 2026 |
| Date Posted | November 25, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0475-2027 |
| Recall Event ID |
100129 |
| Product Classification |
Bed, ac-powered adjustable hospital - Product Code FNL
|
| Product | Medline Brand beds, including homecare beds, bariatric beds, and long-term-care beds, and their associated pendants, accessories, replacement parts, and bed-containing order packages
ALTERRA MAXX UNIVERSAL BED
FCEALTMAXX
BED ONLY, FCE1130, HI/LOW
FCE1130B
BED ONLY,4-MOTOR,HI/LOW,34.5W,LOCK
FCE1385B
DBD-ALTERRA QUICK RELEASE SIDE RAILS
FCEQRSIDE
DBD-BED ONLY, 3-MOTOR, FULL ELEC., 34.5"
FCE1100B
DBD-BED ONLY,4-MOTOR,HI/HOW,39"W
FCE1233B
DBD-BED ONLY,4-MOTOR,HI/LOW,34.5",LITE
FCE1200B
DBD-BED ONLY,4-MOTOR,HI/LOW,34.5"W,LOCK
FCE1232BL
DBD-BED ONLY,4-MOTOR,HI/LOW,34.5W
FCE1232B
DBD-BED PKG,SOFT RAILS,H-F,QUARTERED OAK
FCE1200SRQO
DBD-BED,4 MOTOR,HEAD/FOOT-QRTR OAK
FCE1200NRQO
DBDBED,FULL ELEC,29RAILS,H-F,QUARTERED
FCE1100QO
DBD-BED,FULL ELEC,HI/LOW,CENTRAL LK,5 FN
FCE1232BL-5
DBD-BED,HI/LOW,LOCK, H-F BOARD, QRTR OAK
FCE1232BLNRQO
DBD-BED,HI-LOW,ELECT,ALTERRA,CONTAINER O
FCE1232BX
DBD-BED,PKG,HI/LOW,LOCK,H-F BOARD,QRTR
FCE1385NRQO
DBD-BED,PKG,HI-LOW,29RLS,H-F BOARD,QTRD
FCE1232QO
DBD-BED,PKG,HI-LOW,H-F BOARD,QTRD OAK
FCE1232NRQO
DBD-BED,PKG,HI-LOW,SFTRLS,H-F BOARD,QTRD
FCE1232SRQO
DBD-BED,PKG,LOCK,SOFT RLS,H-F BOARD,AM C
FCE1130SRAC
DBD-BED,PKG,LOCK,SOFT RLS,H-F BOARD,QTRD
FCE1130SRQO
FCE1385SRQO
DBD-BED,PKG,MAXX,SOFT RLS, SOLAR OAK HFB
FCEALTMAXXQO
DBD-BED,PKG,SOFT RLS,36" HFB CHERRY,42"W
FCE1385QC2
DBD-BED,PKG,SOFT RLS,36" HFB OAK,42"W KI
FCE1385QC1
DBD-RAIL, XTRA TALL(SOLD AS PAIR),F/FCE
FCE1232RSRXT
DBD-RAIL,SIDE,HEAD SECTION,SOLD AS PAIR
FCE1232RSR
DBS-BED ONLY,4-MOTOR,HI/LOW,39"W,L
FCE1233BL
DBS-BEDPKG,LOCK,SOFT RLS,H-F BOARD,QTRD
FCE1232BLSRQO
RAIL,SIDE,HEAD SECTION,SOFT TOUCH
FCE1232RSRN |
| Code Information |
FCEALTMAXX
UDI-DI 10888277651838
FCE1130B
UDI-DI 10888277694521
FCE1385B
UDI-DI 10888277100022
FCEQRSIDE
UDI-DI 10884389104507
FCE1100B
UDI-DI 10080196313709
FCE1233B
UDI-DI 40884389106496
FCE1200B
UDI-DI 10080196564606
FCE1232BL
UDI-DI 40884389117270
FCE1232B
UDI-DI 40080196289067
FCE1200SRQO
UDI-DI 10888277195684
FCE1200NRQO
UDI-DI 10884389195680
FCE1100QO
UDI-DI 10888277145016
FCE1232BL-5
UDI-DI 10080196995363
FCE1232BLNRQO
UDI-DI 10884389195697
FCE1232BX
No UDI-DI (Discontinued)
FCE1385NRQO
UDI-DI 10888277136939
FCE1232QO
UDI-DI 10888277145030
FCE1232NRQO
UDI-DI 10884389195703
FCE1232SRQO
UDI-DI 10888277195707
FCE1130SRAC
No UDI-DI (Discontinued)
FCE1130SRQO
No UDI-DI (Discontinued)
FCE1385SRQO
UDI-DI 10080196151486
FCEALTMAXXQO
No UDI-DI (Discontinued)
FCE1385QC2
No U DI-DI (Discontinued)
FCE1385QC1
No UDI-DI (Discontinued)
FCE1232RSRXT
UDI-DI 40080196326106
FCE1232RSR
UDI-DI 40080196289081
FCE1233BL
UDI-DI 40884389141947
FCE1232BLSRQO
UDI-DI 10888277195691
FCE1232RSRN
UDI-DI 10080196313877
|
| FEI Number |
1417592
|
Recalling Firm/ Manufacturer |
Medline Industries, LP
3 Lakes Dr
Northfield
IL
60093-2753
|
| For Additional Information Contact | Karin Johnson 800-633-5463 |
Manufacturer Reason for Recall | All beds and their associated pendants, accessories, replacement parts, and bed-containing order packages are being removed from the market. This is related to previously identified issues: 1. electrical system safety risks that may result in fire, and 2. entrapment risk from the use of any bed accessories. |
FDA Determined Cause 2 | Device Design |
| Action | The firm began notifying customers via email and first-class mail on October 14, 2026, with an "URGENT MEDICAL DEVICE FIELD SAFETY NOTICE" dated October 14, 2026. An additional notification was posted on Medline's website on 10/24/26. The notification described the product, the problem and actions to be taken. Per firm's instructions:
1. Immediately unplug the bed and any associated battery packs or batteries from the
bed or accessories.
2. Do not use the electrical functionality of the bed. You may continue to use the
Medline Brand bed with a mattress in manual mode only until you replace the bed.
3. You may only make electrical adjustments to the bed in emergency situations (e.g.,
choking, CPR).
4. If the bed is plugged back in for emergency use, do not use extension cords.
5. If the bed is plugged back in for emergency use, ensure that the Medline Brand bed
pendants and power cords are routed and secured.
6. Do not use any bed accessories, including side rails, with Medline Brand beds.
Firms Follow Up Action:
1. Medline will provide our customers with a label containing safety instructions to
place on Medline Brand beds. If you do not receive a label from the provider or
distributor from whom you purchased or received the bed by November 30, 2026,
please contact Medline Recall Customer Service at 800-950-0128, option 7.
2. If you received a Medline Brand bed within the last 5 years, please contact the
provider or distributor from whom you purchased or received the bed to determine
whether you are eligible for a reimbursement and/or credit.
3. To facilitate the removal and replacement of the affected Medline Brand beds, you
may contact the provider or distributor from whom you purchased or received the
bed for further information.
4. If you purchased your bed directly from Medline, please contact Medline Recall
Customer Service at 800-950-0128, option 7. |
| Quantity in Commerce | 3,248,035 (3,217,297 US; 30,738 OUS) total |
| Distribution | US distribution nationwide, including Puerto Rico.
OUS distribution to the following countries/territories:
AMERICAN SAMOA
CANADA
CAYMAN ISLANDS
CHILE
CHINA
COSTA RICA
EL SALVADOR
GUAM
HONDURAS
INDIA
MEXICO
NICARAGUA
PANAMA
SAUDI ARABIA
TRINADAD AND TOBAGO
TURKS AND CAICOS
UAE
VIETNAM
VIRGIN ISLANDS
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|