 Class 1 Device Recall Medlinesee related information
Date Initiated by FirmOctober 14, 2026
Date PostedNovember 25, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0475-2027
Recall Event ID 100129
Product Classification Bed, ac-powered adjustable hospital - Product Code FNL
ProductMedline Brand beds, including homecare beds, bariatric beds, and long-term-care beds, and their associated pendants, accessories, replacement parts, and bed-containing order packages ALTERRA MAXX UNIVERSAL BED FCEALTMAXX BED ONLY, FCE1130, HI/LOW FCE1130B BED ONLY,4-MOTOR,HI/LOW,34.5W,LOCK FCE1385B DBD-ALTERRA QUICK RELEASE SIDE RAILS FCEQRSIDE DBD-BED ONLY, 3-MOTOR, FULL ELEC., 34.5" FCE1100B DBD-BED ONLY,4-MOTOR,HI/HOW,39"W FCE1233B DBD-BED ONLY,4-MOTOR,HI/LOW,34.5",LITE FCE1200B DBD-BED ONLY,4-MOTOR,HI/LOW,34.5"W,LOCK FCE1232BL DBD-BED ONLY,4-MOTOR,HI/LOW,34.5W FCE1232B DBD-BED PKG,SOFT RAILS,H-F,QUARTERED OAK FCE1200SRQO DBD-BED,4 MOTOR,HEAD/FOOT-QRTR OAK FCE1200NRQO DBDBED,FULL ELEC,29RAILS,H-F,QUARTERED FCE1100QO DBD-BED,FULL ELEC,HI/LOW,CENTRAL LK,5 FN FCE1232BL-5 DBD-BED,HI/LOW,LOCK, H-F BOARD, QRTR OAK FCE1232BLNRQO DBD-BED,HI-LOW,ELECT,ALTERRA,CONTAINER O FCE1232BX DBD-BED,PKG,HI/LOW,LOCK,H-F BOARD,QRTR FCE1385NRQO DBD-BED,PKG,HI-LOW,29RLS,H-F BOARD,QTRD FCE1232QO DBD-BED,PKG,HI-LOW,H-F BOARD,QTRD OAK FCE1232NRQO DBD-BED,PKG,HI-LOW,SFTRLS,H-F BOARD,QTRD FCE1232SRQO DBD-BED,PKG,LOCK,SOFT RLS,H-F BOARD,AM C FCE1130SRAC DBD-BED,PKG,LOCK,SOFT RLS,H-F BOARD,QTRD FCE1130SRQO FCE1385SRQO DBD-BED,PKG,MAXX,SOFT RLS, SOLAR OAK HFB FCEALTMAXXQO DBD-BED,PKG,SOFT RLS,36" HFB CHERRY,42"W FCE1385QC2 DBD-BED,PKG,SOFT RLS,36" HFB OAK,42"W KI FCE1385QC1 DBD-RAIL, XTRA TALL(SOLD AS PAIR),F/FCE FCE1232RSRXT DBD-RAIL,SIDE,HEAD SECTION,SOLD AS PAIR FCE1232RSR DBS-BED ONLY,4-MOTOR,HI/LOW,39"W,L FCE1233BL DBS-BEDPKG,LOCK,SOFT RLS,H-F BOARD,QTRD FCE1232BLSRQO RAIL,SIDE,HEAD SECTION,SOFT TOUCH FCE1232RSRN
Code Information FCEALTMAXX UDI-DI 10888277651838 FCE1130B UDI-DI 10888277694521 FCE1385B UDI-DI 10888277100022 FCEQRSIDE UDI-DI 10884389104507 FCE1100B UDI-DI 10080196313709 FCE1233B UDI-DI 40884389106496 FCE1200B UDI-DI 10080196564606 FCE1232BL UDI-DI 40884389117270 FCE1232B UDI-DI 40080196289067 FCE1200SRQO UDI-DI 10888277195684 FCE1200NRQO UDI-DI 10884389195680 FCE1100QO UDI-DI 10888277145016 FCE1232BL-5 UDI-DI 10080196995363 FCE1232BLNRQO UDI-DI 10884389195697 FCE1232BX No UDI-DI (Discontinued) FCE1385NRQO UDI-DI 10888277136939 FCE1232QO UDI-DI 10888277145030 FCE1232NRQO UDI-DI 10884389195703 FCE1232SRQO UDI-DI 10888277195707 FCE1130SRAC No UDI-DI (Discontinued) FCE1130SRQO No UDI-DI (Discontinued) FCE1385SRQO UDI-DI 10080196151486 FCEALTMAXXQO No UDI-DI (Discontinued) FCE1385QC2 No U DI-DI (Discontinued) FCE1385QC1 No UDI-DI (Discontinued) FCE1232RSRXT UDI-DI 40080196326106 FCE1232RSR UDI-DI 40080196289081 FCE1233BL UDI-DI 40884389141947 FCE1232BLSRQO UDI-DI 10888277195691 FCE1232RSRN UDI-DI 10080196313877
FEI Number 1417592
Recalling Firm/
Manufacturer
Medline Industries, LP
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information ContactKarin Johnson
800-633-5463
Manufacturer Reason
for Recall
All beds and their associated pendants, accessories, replacement parts, and bed-containing order packages are being removed from the market. This is related to previously identified issues: 1. electrical system safety risks that may result in fire, and 2. entrapment risk from the use of any bed accessories.
FDA Determined
Cause 2
Device Design
ActionThe firm began notifying customers via email and first-class mail on October 14, 2026, with an "URGENT MEDICAL DEVICE FIELD SAFETY NOTICE" dated October 14, 2026. An additional notification was posted on Medline's website on 10/24/26. The notification described the product, the problem and actions to be taken. Per firm's instructions: 1. Immediately unplug the bed and any associated battery packs or batteries from the bed or accessories. 2. Do not use the electrical functionality of the bed. You may continue to use the Medline Brand bed with a mattress in manual mode only until you replace the bed. 3. You may only make electrical adjustments to the bed in emergency situations (e.g., choking, CPR). 4. If the bed is plugged back in for emergency use, do not use extension cords. 5. If the bed is plugged back in for emergency use, ensure that the Medline Brand bed pendants and power cords are routed and secured. 6. Do not use any bed accessories, including side rails, with Medline Brand beds. Firms Follow Up Action: 1. Medline will provide our customers with a label containing safety instructions to place on Medline Brand beds. If you do not receive a label from the provider or distributor from whom you purchased or received the bed by November 30, 2026, please contact Medline Recall Customer Service at 800-950-0128, option 7. 2. If you received a Medline Brand bed within the last 5 years, please contact the provider or distributor from whom you purchased or received the bed to determine whether you are eligible for a reimbursement and/or credit. 3. To facilitate the removal and replacement of the affected Medline Brand beds, you may contact the provider or distributor from whom you purchased or received the bed for further information. 4. If you purchased your bed directly from Medline, please contact Medline Recall Customer Service at 800-950-0128, option 7.
Quantity in Commerce3,248,035 (3,217,297 US; 30,738 OUS) total
DistributionUS distribution nationwide, including Puerto Rico. OUS distribution to the following countries/territories: AMERICAN SAMOA CANADA CAYMAN ISLANDS CHILE CHINA COSTA RICA EL SALVADOR GUAM HONDURAS INDIA MEXICO NICARAGUA PANAMA SAUDI ARABIA TRINADAD AND TOBAGO TURKS AND CAICOS UAE VIETNAM VIRGIN ISLANDS
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-