 Class 1 Device Recall Medlinesee related information
Date Initiated by FirmOctober 14, 2026
Date PostedNovember 25, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0476-2027
Recall Event ID 100129
Product Classification Mattress, air flotation, alternating pressure - Product Code FNM
ProductMedline Brand beds, including homecare beds, bariatric beds, and long-term-care beds, and their associated pendants, accessories, replacement parts, and bed-containing order packages DBD-INFLATOR,QUICK,FOR DPS,W/ CONNECTOR MDT24QUICKD DBD-KIT LAL MAT 36X80X8, CUSH 18X18 MDT24SUPRDPSK DBD-MATTRESS ONLY, BARI, ALT PRESS, 42" MSCAPM42MAT DBD-MATTRESS, EQUALIZE AIRE 12000, MINI MDTEMINI DBD-MATTRESS, SERENE AIR "TRUE" LAL 48X8 MDT24SER48 DBD-MATTRESS,BARI,ALT PRESS,42X80X10 MSCBARIAPL42 MSCBARIAPM42 DBD-MATTRESS,BARI,ALT PRESS,42X80X10, RE MSCBARIAPL42R DBD-MATTRESS,BARI,ALT PRESS,48X80X10 MSCBARIAPM48 DBD-MATTRESS,EQAIRE PLUS 12000,35X80X7, MDTEP123580A MDTEP123580AR DBD-MATTRESS,EQAIRE PLUS 12000,35X84X7, MDTEP123584A DBD-MATTRESS,EQAIRE PLUS 12000,42X80X7, MDTEP124280A MDTEP124280AR DBD-MATTRESS,EQAIRE PLUS 12000,42X84X7, MDTEP124284A MDTEP124284AR DBD-MATTRESS,EQAIRE PLUS 9000,35X80X7, A MDTEP93580A DBD-MATTRESS,EQAIRE PLUS 9000,35X80X7, R MDTEP93580AR DBD-MATTRESS,EQAIRE PLUS 9000,35X84X7, A MDTEP93584A DBD-MATTRESS,EQAIRE PLUS 9000,42X80X7, A MDTEP94280A DBD-MATTRESS,EQAIRE PLUS 9000,42X80X7, R MDTEP94280AR DBD-MATTRESS,EQAIRE PLUS 9000,42X84X7, A MDTEP94284A DBD-MATTRESS,FUSION 2K, 36X80X6 MDTFUS2K80 DBD-MATTRESS,FUSION+, 36X80X6 MDTFUSION80 DBD-MATTRESS,FUSION+,36X84X6 MDTFUSION84 DBD-MATTRESS,LOW AIR,ALTERNATING PRESS MDT24SUPRACXC DBD-MATTRESS,REPLACEMENT,SUPRA APL MDT24APLMAT DBD-MATTRESS,REPLACEMENT,SUPRA CXC MDT24CXCMAT DBD-MATTRESS,SYSTEM,A10,LAL/ALTERNATING MDT24A10 DBD-MATTRESS,SYSTEM,MEDTECH MAX,36X80X10 MDT24MEDAIRMX DBD-MATTRESS,SYSTEM,MEDTECH MAX,TRUE,36" MDT24MEDMXRE DBD-MATTRESS,SYSTEM,MEDTECH,BARI,TRUE,42 MDT24B42 MDT24B42RN DBD-MATTRESS,SYSTEM,MEDTECH,TRUE LAL,36" MDT24MEDAIRRN DBD-MATTRESS,SYSTEM,MEDTECH,TRUE LAL,36, MDT24MEDRE DBD-MATTRESS,SYSTEM,MEDTECH,TRUE LAL,36X MDT24MEDAIR DBD-MATTRESS,SYSTEM,SERENE,TRUE LAL,36X8 MDT24SER36 DBD-MATTRESS,SYSTEM,SERENE,TRUE LAL,42X8 MDT24SER42 DBD-MATTRESS,SYSTEM,SUPRA APL,LAL/ALT. P MDT24SUPRAAPL DBD-MATTRESS,SYSTEM,SUPRA DPS,LAL/ALT. P MDT24SUPRADPS DBD-MATTRESS,SYSTEM,SUPRA DPS,LAL/ALT.,R MDT24DPSRE DBD-OVERLAY,SYSTEM,GERI CHAIR, ALTERNATI MDT24GERIAP DBD-OVERLAY,SYSTEM,SUPRA EXO,LAL/ALT. PR MDT24SUPRAEXO DBD-PUMP FOR BARI APL ALTERNATING PRES M MSCAPLPUMP DBD-PUMP FOR BARI APM ALTERNATING PRES M MSCAPMPUMP DBD-PUMP,AIR MATTRESS,F/MDT24SUPRACXC MDT24CXCPUMP DBD-PUMP,DYNABEST,SUPRA MATTRESS MDTAPXDPUMP DBD-PUMP,REPLACEMENT,A20,LAL/ALTERNATING MDT24A20PUMP DBD-PUMP,REPLACEMENT,SUPRA EXO,LAL/ALTER MDT24EXOPUMP
Code Information ALL LOTS. MDT24QUICKD No UDI-DI (Discontinued) MDT24SUPRDPSK No UDI-DI (Discontinued) MSCAPM42MAT UDI-DI 10080196404018 MDTEMINI UDI-DI 10080196380725 MDT24SER48 UDI-DI 4713025004361 MSCBARIAPL42 UDI-DI 10080196568789 MSCBARIAPM42 UDI-DI 10080196404292 MSCBARIAPL42R No UDI-DI (Discontinued) MSCBARIAPM48 UDI-DI 10080196404308 MDTEP123580A UDI-DI 10193489086072 MDTEP123580AR UDI-DI 10193489140293 MDTEP123584A UDI-DI 10193489086089 MDTEP124280A UDI-DI 10193489105964 MDTEP124280AR UDI-DI 10193489140347 MDTEP124284A UDI-DI 10193489105971 MDTEP124284AR UDI-DI 10193489140354 MDTEP93580A UDI-DI 10193489086096 MDTEP93580AR UDI-DI 10193489140378 MDTEP93584A UDI-DI 10193489086102 MDTEP94280A UDI-DI 10193489105988 MDTEP94280AR UDI-DI 10193489140422 MDTEP94284A UDI-DI 10193489105995 MDTFUS2K80 UDI-DI 10888277348622 MDTFUSION80 UDI-DI 10884389170793 MDTFUSION84 UDI-DI 10888277170797 MDT24SUPRACXC UDI-DI 10080196372188 MDT24APLMAT UDI-DI 10080196371914 MDT24CXCMAT UDI-DI 10080196371990 MDT24A10 UDI-DI 10193489135671 MDT24MEDAIRMX UDI-DI 10080196372058 MDT24MEDMXRE UDI-DI 10080196372089 MDT24B42 UDI-DI 10080196371945 MDT24B42RN No UDI-DI (Discontinued) MDT24MEDAIRRN No UDI-DI (Discontinued) MDT24MEDRE UDI-DI 10080196372119 MDT24MEDAIR UDI-DI 10080196372034 MDT24SER36 UDI-DI 4713025002800 MDT24SER42 UDI-DI 4713025004361 MDT24SUPRAAPL UDI-DI 10080196372164 MDT24SUPRADPS UDI-DI 40080196372196 MDT24DPSRE UDI-DI 10080196372010 MDT24GERIAP UDI-DI 10888277144538 MDT24SUPRAEXO UDI-DI 10080196372201 MSCAPLPUMP UDI-DI 10888277667723 MSCAPMPUMP UDI-DI 80196404066 MDT24CXCPUMP UDI-DI 10080196372003 MDT24A20PUMP UDI-DI 10884389945896 MDT24EXOPUMP UDI-DI 10888277190696
FEI Number 1417592
Recalling Firm/
Manufacturer
Medline Industries, LP
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information ContactKarin Johnson
800-633-5463
Manufacturer Reason
for Recall
All beds and their associated pendants, accessories, replacement parts, and bed-containing order packages are being removed from the market. This is related to previously identified issues: 1. electrical system safety risks that may result in fire, and 2. entrapment risk from the use of any bed accessories.
FDA Determined
Cause 2
Device Design
ActionThe firm began notifying customers via email and first-class mail on October 14, 2026, with an "URGENT MEDICAL DEVICE FIELD SAFETY NOTICE" dated October 14, 2026. An additional notification was posted on Medline's website on 10/24/26. The notification described the product, the problem and actions to be taken. Per firm's instructions: 1. Immediately unplug the bed and any associated battery packs or batteries from the bed or accessories. 2. Do not use the electrical functionality of the bed. You may continue to use the Medline Brand bed with a mattress in manual mode only until you replace the bed. 3. You may only make electrical adjustments to the bed in emergency situations (e.g., choking, CPR). 4. If the bed is plugged back in for emergency use, do not use extension cords. 5. If the bed is plugged back in for emergency use, ensure that the Medline Brand bed pendants and power cords are routed and secured. 6. Do not use any bed accessories, including side rails, with Medline Brand beds. Firms Follow Up Action: 1. Medline will provide our customers with a label containing safety instructions to place on Medline Brand beds. If you do not receive a label from the provider or distributor from whom you purchased or received the bed by November 30, 2026, please contact Medline Recall Customer Service at 800-950-0128, option 7. 2. If you received a Medline Brand bed within the last 5 years, please contact the provider or distributor from whom you purchased or received the bed to determine whether you are eligible for a reimbursement and/or credit. 3. To facilitate the removal and replacement of the affected Medline Brand beds, you may contact the provider or distributor from whom you purchased or received the bed for further information. 4. If you purchased your bed directly from Medline, please contact Medline Recall Customer Service at 800-950-0128, option 7.
Quantity in Commerce3,248,035 (3,217,297 US; 30,738 OUS) total
DistributionUS distribution nationwide, including Puerto Rico. OUS distribution to the following countries/territories: AMERICAN SAMOA CANADA CAYMAN ISLANDS CHILE CHINA COSTA RICA EL SALVADOR GUAM HONDURAS INDIA MEXICO NICARAGUA PANAMA SAUDI ARABIA TRINADAD AND TOBAGO TURKS AND CAICOS UAE VIETNAM VIRGIN ISLANDS
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-