| | Class 2 Device Recall ROTAFLOW Centrifugal Pump System |  |
| Date Initiated by Firm | November 04, 2026 |
| Date Posted | November 24, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0544-2027 |
| Recall Event ID |
100287 |
| 510(K)Number | K991864 |
| Product Classification |
Pump, blood, cardiopulmonary bypass, non-roller type - Product Code KFM
|
| Product | ROTAFLOW Centrifugal Pump System;
RotaFlow English/US US-Plug ICU;
Model Number: 701051712;
Software Version: 1.4 (RotaFlow Power Supply Unit); |
| Code Information |
Model Number: 701051712;
UDI-DI: 4037691679013;
***All serial numbers affected*** |
| FEI Number |
3009507273
|
Recalling Firm/ Manufacturer |
Maquet Cardiopulmonary Gmbh
Kehler Str. 31
Rastatt
Germany
|
Manufacturer Reason for Recall | An issue can cause the device to stop charging
the battery after the battery operation mode has
been utilized. Further investigation identified a
software problem. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | On November 4, 2026 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers.
Actions to be taken:
1. Ensure all affected product in your inventory has been properly identified.
2. Please ensure this message is forwarded to any individuals that need notification within your organization or any organization where the affected devices have been transferred.
3. Upon availability of a software update, a Getinge service technician will contact you to schedule an update.
4. Contact Technical Support at techsupportACT.us@getinge.com and 888-943-8872, option 4. option 2 with any questions.
5. Please fill out the Customer Response Form using the link or QR code provided. Getinge has partnered with IQVIA MedTech to coordinate the recall response process. |
| Quantity in Commerce | 17 |
| Distribution | Worldwide distribution - US Nationwide and the countries of Angola, Australia, Austria, Azerbaijan, Bahamas, Bangladesh, Belgium, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Egypt, Estonia, France, Georgia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kenya, Korea, Republic of (South Korea), Luxembourg, Macao, Malaysia, Mauritania, Mexico, Moldova, Republic of, Netherlands, New Zealand, Nicaragua, Norway, Oman, Pakistan, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Republic Of China, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = KFM
|
|
|
|