• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall Medline

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall Medlinesee related information
Date Initiated by FirmOctober 14, 2026
Date PostedNovember 25, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0478-2027
Recall Event ID 100129
Product Classification Bed, therapeutic, ac-powered, adjustable home-use - Product Code LLI
ProductMedline Brand beds, including homecare beds, bariatric beds, and long-term-care beds, and their associated pendants, accessories, replacement parts, and bed-containing order packages DBD-36X77X6 BEHAVIORAL FOCONVOLUTED MATT MSC36776BEHAV DBD-36X80X6 BEHAVIORAL FOCONVOLUTED MATT MSC36806BEHAV DBD-ENTENDER,MATTRESS,NYLEX, 48X5X10 MDTGAP48510 DBD-ENTENDER,MATTRESS,NYLEX, 54X5X10 MDTGAP54510 DBD-ENTENDER,MATTRESS,NYLEX, 54X5X6 MDTGAP5456 DBD-FOAM 42X80 FOAM/BOXSPRING MDT23FSET4280 DBD-FOAM MATTRESS 36X80 MSCADV80NC DBD-FOAM MATTRESS, 36X80X5 MSCADV805GC DBD-KIT, EXTENDS 39" MATTRESS INTO 48" M MDTEXT48KIT DBD-KIT, EXTENDS 39" MATTRESS INTO 54" M MDTEXT54KIT DBD-MAT, FOAM, NYLEX, NO VENTS, 35X80X3 MDT35803F DBD-MATT. ODYSSEY ELITE W/FIRE BAR. 36X8 MDT360084FBA DBD-MATTRESS ADVANTAGE CON 36X76X7 ARMGR MSCCON176A MSCCON176ARE DBD-MATTRESS ADVANTAGE CON 36X80X7 ARMGR MSCCON180ARE DBD-MATTRESS ADVANTAGE CON 36X84X7 ARMGR MSCCON184A MSCCON184ARE DBD-MATTRESS BARI ADV GRAPHITE 1000# 47X MSCADVG8B4280 DBD-MATTRESS BARIATRIC, 1000LB,RE,48X80 MDT23B148807R DBD-MATTRESS BARIATRIC, 1000LBS,F/B,54X8 MDT23B154807F DBD-MATTRESS CVR,ADVANTAGE BARI,800LB.,3 MSCADV8B3580C DBD-MATTRESS CVR,ADVANTAGE BARI,800LB.,4 MSCADV8B488C DBD-MATTRESS EXTENDER, 35X8X4 MDTEXTEND8 DBD-MATTRESS EXTENDER, FIRM, 3X80X6, NYL MDT239945 DBD-MATTRESS GAP PROTECTOR 36" 6"HIGH MDT23NOGAP2 DBD-MATTRESS GAP PROTECTOR 80"LONG 6"HIG MDT23NOGAP DBD-MATTRESS GEN,BULK SLCT BARI,42X80,RE MSCGENB4280RN DBD-MATTRESS GEN,BULK SLCT BARIMATT,39X8 MSCGENB3980X DBD-MATTRESS GEN,BULK SLCT BARIMATT,42X8 MSCGENB4280X MSCGENB4284X DBD-MATTRESS GEN,BULK SLCT BARIMATT,48X8 MSCGENB4880X DBD-MATTRESS GEN,BULK SLCT BARIMATT,54X8 MSCGENB5480X DBD-MATTRESS GEN,CTM SLCT BARIMATT,39X80 MSCGENB3980 DBD-MATTRESS GEN,CTM SLCT BARIMATT,42X80 MSCGENB4280 MSCGENB4280R MSCGENB4280RX MSCGENBAR542F DBD-MATTRESS GEN,CTM SLCT BARIMATT,42X84 MSCGENB4284 MSCGENB4284R MSCGENB4284RX DBD-MATTRESS GEN,CTM SLCT BARIMATT,42X86 MSCGENB4286X DBD-MATTRESS GEN,CTM SLCT BARIMATT,48X80 MSCGENB4880 MSCGENBAR148F DBD-MATTRESS GEN,CTM SLCT BARIMATT,48X86 MSCGENB4886X DBD-MATTRESS GEN,CTM SLCT BARIMATT,54X80 MSCGENB5480 DBD-MATTRESS NYLEX PREVENT F/B 36X80X6 MDT239980FA DBD-MATTRESS ODYSSEY ELITE 80" RE/FB MDT360080REF DBD-MATTRESS ODYSSEY ELITE 84" RE/FB MDT360084REF DBD-MATTRESS ODYSSEY ELITE, 36X76 FB MDT360076FBA DBD-MATTRESS THERATECH 5500 36X80X6.5 MSCT5580 DBD-MATTRESS THERATECH 5500 36X80X6.5, R MSCT5580R DBD-MATTRESS THERATECH 5500,29X72X6.5",P MSCT5572PSYC DBD-MATTRESS THERATECH 5500,30X74X6.5" MSCT553074 DBD-MATTRESS THERATECH 5500,34X84X6" MSCT553484 DBD-MATTRESS THERATECH 5500,35X80X6" MSCT5535806 DBD-MATTRESS THERATECH 5500,35X84X6.5" MSCT553584 DBD-MATTRESS THERATECH 5500,36X75X6.5" MSCT5575 DBD-MATTRESS THERATECH 5500,36X78X6.5,AR MSCT5578A DBD-MATTRESS THERATECH 5500,36X79X6.5" MSCT5579 DBD-MATTRESS THERATECH 5500,37X79X6.5" MSCT553779 DBD-MATTRESS THERATECH 5500,38X82X6.5",P MSCT5582PSY DBD-MATTRESS THERATECH 5500,39X80X6.5 MSCT553980 DBD-MATTRESS TWIN NYLEX II CONV. FB 39X7 MDT23117539F DBD-MATTRESS W/RAISE EDGE,ADVTGE,36X80X5 MSCADV0380FSR DBD-MATTRESS, 31X88X6 PSYCH MSC31886PSY DBD-MATTRESS, ADV DIRECT, HOMECARE, 36X3 MSCADVDR3636 DBD-MATTRESS, ADVANTAGE, CON, 60X80X7, R MSCCON6080 DBD-MATTRESS, ADVANTAGE, GROUP I, 36X76X MSCADV0376F DBD-MATTRESS, ADVANTAGE, GROUP I, 36X80X MSCADV0380F MSCADV0380FS DBD-MATTRESS, ADVANTAGE, GROUP I, 36X84X MSCADV0384F DBD-MATTRESS, ADVANTAGE, HOMECARE, 36X80 MSCADVHC80 DBD-MATTRESS, ADVANTAGE300, 36X80X6, FB MSCADV0380FR DBD-MATTRESS, BARIATRIC, 1000LB, FIRE B, MDT23B139807F DBD-MATTRESS, BARIATRIC, 650LBS F/B, 48X MDT23B548846F DBD-MATTRESS, BARIATRIC,42X80, RE, 650LB MDT23B542806R DB
Code Information ALL LOTS. MDR107003L UDI-DI 40080196294344 MDR107002L UDI-DI 40080196294320 MDSRAILBAR UDI-DI 10884389957523 MDRBEDBSCFIF UDI-DI 10080196008704 MDRBEDBSCSBH UDI-DI 80196204444 MDR3EFRFM No UDI-DI (Discontinued) MDR3EFRIS No UDI-DI (Discontinued) MDR3EHRFM No UDI-DI (Discontinued) MDR3EHRIS No UDI-DI (Discontinued) MDR2EFR No UDI-DI (Discontinued) MDR2EFRIS No UDI-DI (Discontinued) MDR2EHRFM No UDI-DI (Discontinued) MDR2LHRTM No UDI-DI (Discontinued) MDR2EHRIS No UDI-DI (Discontinued) MDR2EFRFM No UDI-DI (Discontinued) MDR107003E UDI-DI 40080196321002 MDR107003E-4 UDI-DI 193489004031 MDR107003ELO UDI-DI 10080196546992 MDR107003 No UDI-DI (Discontinued) MDR107002P No UDI-DI (Discontinued) MDR107002 UDI-DI 40080196748090 MDR107002E-1 UDI-DI 193489003683 MDR107002E UDI-DI 40080196320999 MDR107002E-4 UDI-DI 193489003683 GIC-5410 No UDI-DI (Discontinued) MDR107001 UDI-DI 10888277106093 MDS89695 No UDI-DI (Discontinued) MDS89697 UDI-DI 80196746029 MDS89690 UDI-DI 10080196134656 MDS89694 No UDI-DI (Discontinued) MDREC107003 No UDI-DI (Discontinued) MDREC107103 No UDI-DI (Discontinued) MDRSW107103 No UDI-DI (Discontinued) MDSLOWBED UDI-DI 10080196361090 MDSLOWBASIC UDI-DI 10080196549641 MDR107003LO UDI-DI 40080196325901 MDS89698N UDI-DI 884389804615 MDS89695N UDI-DI 193489002709 MDS89694N UDI-DI 40884389118727
FEI Number 1417592
Recalling Firm/
Manufacturer
Medline Industries, LP
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information ContactKarin Johnson
800-633-5463
Manufacturer Reason
for Recall
All beds and their associated pendants, accessories, replacement parts, and bed-containing order packages are being removed from the market. This is related to previously identified issues: 1. electrical system safety risks that may result in fire, and 2. entrapment risk from the use of any bed accessories.
FDA Determined
Cause 2
Device Design
ActionThe firm began notifying customers via email and first-class mail on October 14, 2026, with an "URGENT MEDICAL DEVICE FIELD SAFETY NOTICE" dated October 14, 2026. An additional notification was posted on Medline's website on 10/24/26. The notification described the product, the problem and actions to be taken. Per firm's instructions: 1. Immediately unplug the bed and any associated battery packs or batteries from the bed or accessories. 2. Do not use the electrical functionality of the bed. You may continue to use the Medline Brand bed with a mattress in manual mode only until you replace the bed. 3. You may only make electrical adjustments to the bed in emergency situations (e.g., choking, CPR). 4. If the bed is plugged back in for emergency use, do not use extension cords. 5. If the bed is plugged back in for emergency use, ensure that the Medline Brand bed pendants and power cords are routed and secured. 6. Do not use any bed accessories, including side rails, with Medline Brand beds. Firms Follow Up Action: 1. Medline will provide our customers with a label containing safety instructions to place on Medline Brand beds. If you do not receive a label from the provider or distributor from whom you purchased or received the bed by November 30, 2026, please contact Medline Recall Customer Service at 800-950-0128, option 7. 2. If you received a Medline Brand bed within the last 5 years, please contact the provider or distributor from whom you purchased or received the bed to determine whether you are eligible for a reimbursement and/or credit. 3. To facilitate the removal and replacement of the affected Medline Brand beds, you may contact the provider or distributor from whom you purchased or received the bed for further information. 4. If you purchased your bed directly from Medline, please contact Medline Recall Customer Service at 800-950-0128, option 7.
Quantity in Commerce3,248,035 (3,217,297 US; 30,738 OUS) total
DistributionUS distribution nationwide, including Puerto Rico. OUS distribution to the following countries/territories: AMERICAN SAMOA CANADA CAYMAN ISLANDS CHILE CHINA COSTA RICA EL SALVADOR GUAM HONDURAS INDIA MEXICO NICARAGUA PANAMA SAUDI ARABIA TRINADAD AND TOBAGO TURKS AND CAICOS UAE VIETNAM VIRGIN ISLANDS
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-