 Class 1 Device Recall Medlinesee related information
Date Initiated by FirmOctober 14, 2026
Date PostedNovember 25, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0479-2027
Recall Event ID 100129
Product Classification Bariatric bed - Product Code OSI
ProductMedline Brand beds, including homecare beds, bariatric beds, and long-term-care beds, and their associated pendants, accessories, replacement parts, and bed-containing order packages BED, BARIATRIC, LIGHTWEIGHT, 42 W MDR107004L DBD-BED PKG, BARIATRIC, HALF RAIL, FOAM MDR004HRFM DBD-BED PKG,BARI,LIGHTWEIGHT,HALF RAI MDR004LHRFM DBD-BED,BARIATRIC,FULL ELECTRIC MDR107004 DBD-FRAME,BED,BARIATRIC,EXPAND,1000LB,37 MDT3948 DBD-RAIL, BARIATRIC HALF BED F/ MDR10700 MDS89697BAR DBD-RAIL, F/ BAR BED MDR107005 MDS89697900 RAIL, HALF, BARIATRIC, F/MDR107004L MDS89697BARL
Code Information ALL LOTS. MDR107004L UDI-DI 10197344098385 MDR004HRFM No UDI-DI (Discontinued) MDR004LHRFM No UDI-DI (Discontinued) MDR107004 UDI-DI 40080196321019 MDT3948 UDI-DI 10080196373116 MDS89697BAR UDI-DI 10080196321155 MDS89697900 UDI-DI 10197344135448 MDS89697BARL UDI-DI 10197344098392
FEI Number 1417592
Recalling Firm/
Manufacturer
Medline Industries, LP
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information ContactKarin Johnson
800-633-5463
Manufacturer Reason
for Recall
All beds and their associated pendants, accessories, replacement parts, and bed-containing order packages are being removed from the market. This is related to previously identified issues: 1. electrical system safety risks that may result in fire, and 2. entrapment risk from the use of any bed accessories.
FDA Determined
Cause 2
Device Design
ActionThe firm began notifying customers via email and first-class mail on October 14, 2026, with an "URGENT MEDICAL DEVICE FIELD SAFETY NOTICE" dated October 14, 2026. An additional notification was posted on Medline's website on 10/24/26. The notification described the product, the problem and actions to be taken. Per firm's instructions: 1. Immediately unplug the bed and any associated battery packs or batteries from the bed or accessories. 2. Do not use the electrical functionality of the bed. You may continue to use the Medline Brand bed with a mattress in manual mode only until you replace the bed. 3. You may only make electrical adjustments to the bed in emergency situations (e.g., choking, CPR). 4. If the bed is plugged back in for emergency use, do not use extension cords. 5. If the bed is plugged back in for emergency use, ensure that the Medline Brand bed pendants and power cords are routed and secured. 6. Do not use any bed accessories, including side rails, with Medline Brand beds. Firms Follow Up Action: 1. Medline will provide our customers with a label containing safety instructions to place on Medline Brand beds. If you do not receive a label from the provider or distributor from whom you purchased or received the bed by November 30, 2026, please contact Medline Recall Customer Service at 800-950-0128, option 7. 2. If you received a Medline Brand bed within the last 5 years, please contact the provider or distributor from whom you purchased or received the bed to determine whether you are eligible for a reimbursement and/or credit. 3. To facilitate the removal and replacement of the affected Medline Brand beds, you may contact the provider or distributor from whom you purchased or received the bed for further information. 4. If you purchased your bed directly from Medline, please contact Medline Recall Customer Service at 800-950-0128, option 7.
Quantity in Commerce3,248,035 (3,217,297 US; 30,738 OUS) total
DistributionUS distribution nationwide, including Puerto Rico. OUS distribution to the following countries/territories: AMERICAN SAMOA CANADA CAYMAN ISLANDS CHILE CHINA COSTA RICA EL SALVADOR GUAM HONDURAS INDIA MEXICO NICARAGUA PANAMA SAUDI ARABIA TRINADAD AND TOBAGO TURKS AND CAICOS UAE VIETNAM VIRGIN ISLANDS
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-