| | Class 1 Device Recall Medline |  |
| Date Initiated by Firm | October 14, 2026 |
| Date Posted | November 25, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0480-2027 |
| Recall Event ID |
100129 |
| Product Classification |
Adult Portable Bed Rails for Medical Purposes - Product Code QTC
|
| Product | Medline Brand beds, including homecare beds, bariatric beds, and long-term-care beds, and their associated pendants, accessories, replacement parts, and bed-containing order packages
BAR,ASSIST,FLIP DOWN FOR FCE1232 BED
FCE1232ABN
DBD-ADJUSTABLE LENGTH HOME STYLE,3 BAR1
MDS89696N |
| Code Information |
ALL LOTS.
FCE1232ABN
UDI-DI 40884389106427
MDS89696N
UDI-DI 193489003133
|
| FEI Number |
1417592
|
Recalling Firm/ Manufacturer |
Medline Industries, LP
3 Lakes Dr
Northfield
IL
60093-2753
|
| For Additional Information Contact | Karin Johnson 800-633-5463 |
Manufacturer Reason for Recall | All beds and their associated pendants, accessories, replacement parts, and bed-containing order packages are being removed from the market. This is related to previously identified issues: 1. electrical system safety risks that may result in fire, and 2. entrapment risk from the use of any bed accessories. |
FDA Determined Cause 2 | Device Design |
| Action | The firm began notifying customers via email and first-class mail on October 14, 2026, with an "URGENT MEDICAL DEVICE FIELD SAFETY NOTICE" dated October 14, 2026. An additional notification was posted on Medline's website on 10/24/26. The notification described the product, the problem and actions to be taken. Per firm's instructions:
1. Immediately unplug the bed and any associated battery packs or batteries from the
bed or accessories.
2. Do not use the electrical functionality of the bed. You may continue to use the
Medline Brand bed with a mattress in manual mode only until you replace the bed.
3. You may only make electrical adjustments to the bed in emergency situations (e.g.,
choking, CPR).
4. If the bed is plugged back in for emergency use, do not use extension cords.
5. If the bed is plugged back in for emergency use, ensure that the Medline Brand bed
pendants and power cords are routed and secured.
6. Do not use any bed accessories, including side rails, with Medline Brand beds.
Firms Follow Up Action:
1. Medline will provide our customers with a label containing safety instructions to
place on Medline Brand beds. If you do not receive a label from the provider or
distributor from whom you purchased or received the bed by November 30, 2026,
please contact Medline Recall Customer Service at 800-950-0128, option 7.
2. If you received a Medline Brand bed within the last 5 years, please contact the
provider or distributor from whom you purchased or received the bed to determine
whether you are eligible for a reimbursement and/or credit.
3. To facilitate the removal and replacement of the affected Medline Brand beds, you
may contact the provider or distributor from whom you purchased or received the
bed for further information.
4. If you purchased your bed directly from Medline, please contact Medline Recall
Customer Service at 800-950-0128, option 7. |
| Quantity in Commerce | 3,248,035 (3,217,297 US; 30,738 OUS) total |
| Distribution | US distribution nationwide, including Puerto Rico.
OUS distribution to the following countries/territories:
AMERICAN SAMOA
CANADA
CAYMAN ISLANDS
CHILE
CHINA
COSTA RICA
EL SALVADOR
GUAM
HONDURAS
INDIA
MEXICO
NICARAGUA
PANAMA
SAUDI ARABIA
TRINADAD AND TOBAGO
TURKS AND CAICOS
UAE
VIETNAM
VIRGIN ISLANDS
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|