| Class 2 Device Recall |  |
Date Initiated by Firm | April 01, 2003 |
Date Posted | April 22, 2003 |
Recall Status1 |
Terminated 3 on June 23, 2004 |
Recall Number | Z-0775-03 |
Recall Event ID |
25945 |
PMA Number | P020047 |
Product | Guidant Multi-Link Vision Coronary Stent System
Catalog Numbers: 30 different catalog numbers have been listed.
The firm indicates that total numbers of devices subject to this removal are being defined. |
Code Information |
All lots are included in the action. |
Recalling Firm/ Manufacturer |
Guidant Corporation ACS 26531 Ynez Road Temecula CA 92591-4628
|
For Additional Information Contact | James C. McMahon, PhD 909-914-2298 |
Manufacturer Reason for Recall | Complaints that the stents are being dislodged from the delivery system. |
FDA Determined Cause 2 | Other |
Action | Information as an Advisory Notice was sent by facsimile to the Competent Authorities and to distributors on April 1 2003. Recall letters will be delivered by hand in most instances. |
Quantity in Commerce | Estimated 7 to 9 thousand units |
Distribution | No USA distribution, only to the countries of Australia, Austria, Belgium, Denmark, Finland, France, Germany, Greece, Italy, Lebanon, Netherlands, Norway, Poland, Portugal, Spain, Sweden, Switzerland, Tunisia, United Kingdom. |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|