| | Class 2 Device Recall |  |
| Date Initiated by Firm | February 20, 2002 |
| Date Posted | September 12, 2003 |
| Recall Status1 |
Terminated 3 on April 11, 2005 |
| Recall Number | Z-1205-03 |
| Recall Event ID |
26581 |
| 510(K)Number | K912470 |
| Product Classification |
Table, Radiographic, Non-Tilting, Powered - Product Code IZZ
|
| Product | Multidiagnost 97 |
| Code Information |
The firm utilizes a site number specific to each unit: X1207, X1069, X0668, 76345, 17605, 17436, 6852, 6853, 13709, 13971, 6935, 13608, 13951, 6981, 6993, 35481, 86113, 38023, 41372, 62666, 62667, 38897, 37995, 59497, 62936, 62504, 62885, 62988, 62937, 74057, 38657, 38530, 38500, 74056, 59633, 45014, 45013, 47166, 40940 |
Recalling Firm/ Manufacturer |
Philips Medical Systems 22100 Bothell Everett Highway Bothell WA 98041
|
| For Additional Information Contact | 425-487-7000 |
Manufacturer Reason for Recall | Spontaneous movement of table without input from an operator |
FDA Determined Cause 2 | Other |
| Action | On 2/20/02 the firm issued Field Change Order to their field personnel, instructing them to visit each site and upgrade each unit. |
| Quantity in Commerce | 39 |
| Distribution | The firm distributed units to hospitals and medical centers located throughout the US. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = IZZ
|
|
|
|