|
Class 2 Device Recall |
|
Date Initiated by Firm |
February 20, 2002 |
Date Posted |
September 12, 2003 |
Recall Status1 |
Terminated 3 on April 11, 2005 |
Recall Number |
Z-1205-03 |
Recall Event ID |
26581 |
510(K)Number |
K912470
|
Product Classification |
Table, Radiographic, Non-Tilting, Powered - Product Code IZZ
|
Product |
Multidiagnost 97 |
Code Information |
The firm utilizes a site number specific to each unit: X1207, X1069, X0668, 76345, 17605, 17436, 6852, 6853, 13709, 13971, 6935, 13608, 13951, 6981, 6993, 35481, 86113, 38023, 41372, 62666, 62667, 38897, 37995, 59497, 62936, 62504, 62885, 62988, 62937, 74057, 38657, 38530, 38500, 74056, 59633, 45014, 45013, 47166, 40940 |
Recalling Firm/ Manufacturer |
Philips Medical Systems 22100 Bothell Everett Highway Bothell WA 98041
|
For Additional Information Contact |
425-487-7000
|
Manufacturer Reason for Recall |
Spontaneous movement of table without input from an operator
|
FDA Determined Cause 2 |
Other |
Action |
On 2/20/02 the firm issued Field Change Order to their field personnel, instructing them to visit each site and upgrade each unit. |
Quantity in Commerce |
39 |
Distribution |
The firm distributed units to hospitals and medical centers located throughout the US. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = IZZ and Original Applicant = PHILIPS MEDICAL SYSTEMS, INC.
|
|
|
|