| | Class 2 Device Recall Diagnost 97 |  |
| Date Initiated by Firm | July 17, 2003 |
| Date Posted | July 20, 2004 |
| Recall Status1 |
Terminated 3 on January 12, 2005 |
| Recall Number | Z-1242-03 |
| Recall Event ID |
26837 |
| 510(K)Number | K912470 |
| Product | Diagnost 97 |
| Code Information |
Diagnost 97 system code 708023, 70852, 70853, 70855, 70859. The firm utilizes a ''site number'' specific to each unit: 13608, 17605, 37995, 38897, 40940, 47166, 59633, 62936, 62937, 6935, 6981, 74057, 13709, 17436, 35481, 59497, 62885, 74056, 76345, 13951, 13971, 38500, 38530, 38657, 41372, 45013, 45014, 62988, 6993, X0668, X1069, X1207, 6852, 62504, 38023, 62666, 62667, 6853 |
Recalling Firm/ Manufacturer |
Philips Medical Systems Sales & Service Region No. America 22100 Bothell Everett Highway Bothell WA 98041
|
| For Additional Information Contact | 425-487-7000 |
Manufacturer Reason for Recall | Labeling of push button caps located on the control panel may not clearly identify the applied Source to Image distance (SID). |
FDA Determined Cause 2 | Other |
| Action | On 7/17/03 the firm issued a letter dated June 12, 2003 to their customers advising that service personnel will visit the consignees and replace the buttons. |
| Quantity in Commerce | 38 |
| Distribution | The firm distributed devices to 81 hospitals and medical centers located throughout the U.S. |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code =
|
|
|
|