• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recallsee related information
Date Initiated by FirmAugust 08, 2003
Date PostedAugust 21, 2003
Recall Status1 Terminated 3 on February 19, 2004
Recall NumberZ-1145-03
Recall Event ID 26933
Product Classification Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms) - Product Code MHX
ProductIntegris Monitor Arm; model P997-04
Code Information All units manufactured prior to June 20, 2003.
FEI Number 1824206
Recalling Firm/
Manufacturer
Hill-Rom Co., Inc.
3 West Pearl St
Batesville IN 47006
For Additional Information ContactTechnical support
800-445-3720
Manufacturer Reason
for Recall
The monitor may separate from the monitor arm, allowing the monitor to fall from the arm.
FDA Determined
Cause 2
Other
ActionAn Urgent Device Correction letter dated 8/8/03 was sent to each hospital account receiving these products. The recall was extended to two additional products via a second letter issued on August 25, 2003. Hospitals were asked to take the units out of service, if possible, until Hill-Rom can inspect/correct them.
Quantity in Commerce2778
DistributionUnited States, Belgium, Canada, China, Egypt, Hong Kong, Japan, Mexico, Paraguay, Saudi Arabia, Switzerland, Taiwan, Turkey, United Arab Emirates and United Kingdom.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-