| | Class 2 Device Recall Bayer |  |
| Date Initiated by Firm | September 26, 2003 |
| Date Posted | October 24, 2003 |
| Recall Status1 |
Terminated 3 on February 04, 2004 |
| Recall Number | Z-0066-04 |
| Recall Event ID |
27414 |
| 510(K)Number | K020616 |
| Product Classification |
Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph - Product Code CHL
|
| Product | Bayer Rapidpoint 405 Blood Gas Analyzer |
| Code Information |
Software Versions below 3.1 |
Recalling Firm/ Manufacturer |
Bayer Corp. 63 North Street Medfield MA 02052-1688
|
| For Additional Information Contact | Thomas Flynn 508-359-3877 |
Manufacturer Reason for Recall | Potential for bias shift in pO2 values due to air bubble in sample path |
FDA Determined Cause 2 | Other |
| Action | Bayer notified each customer by telephone and followed with a FedEx letter dated 9/26/03 providing upgraded software version 3.1 to install . |
| Quantity in Commerce | 494 units |
| Distribution | Nationwide
Foreign: Australia, Austria, Belgium. Canada, France, Germany, Italy, Japan, Korea, Mexico, Netherlands, Russia, S Ireland, Spain, Switzerland |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = CHL
|
|
|
|