|
Class 3 Device Recall |
|
Date Initiated by Firm |
December 03, 2003 |
Date Posted |
January 06, 2004 |
Recall Status1 |
Terminated 3 on February 20, 2004 |
Recall Number |
Z-0299-04 |
Recall Event ID |
27915 |
Product Classification |
System, Test, Carcinoembryonic Antigen - Product Code DHX
|
Product |
Vitros Immunodiagnostic Products CEA REAGENT PACK, Lot 750, REF 192 0115, 100 coated wells per pack. Firm on the label: Ortho-Clinical Diagnostics, Amersham, UK. |
Code Information |
Lot #750, Exp. 14 January 2004 |
Recalling Firm/ Manufacturer |
Ortho-Clinical Diagnostics 100 Indigo Creek Drive Rochester NY 14650-0881
|
For Additional Information Contact |
Sherry L. Phillips 585-453-3728
|
Manufacturer Reason for Recall |
Reagent packs may contain wells that produce a low light signal that could result in calibration failures or negatively biased results.
|
FDA Determined Cause 2 |
Other |
Action |
U.S. customers notified via FED EX letter sent on 12/3/03. Letter includes modified QC procedures to use until replacement product is issued. Foreign affiliates notified via e-mail on 12/3/03 with instructions to notify customers. |
Quantity in Commerce |
524 reagent packs |
Distribution |
154 reagent packs shipped to consignees in the U.S. 370 reagent packs shipped to Ortho/J&J International Affiliates for further distribution to foreign users. Foreign consignees are located in Argentina, Australia, Belgium, Brazil, Chile, China, France, Germany, India, Italy, Japan, Mexico, Panama, Poland, Portugal, Puerto Rico, Singapore, Spain, United Kingdom, and Venezuela. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|